Investigation of combined kV/MV CBCT imaging with a high‐DQE MV detector. Issue 2 (6th December 2018)
- Record Type:
- Journal Article
- Title:
- Investigation of combined kV/MV CBCT imaging with a high‐DQE MV detector. Issue 2 (6th December 2018)
- Main Title:
- Investigation of combined kV/MV CBCT imaging with a high‐DQE MV detector
- Authors:
- Lindsay, C.
Bazalova‐Carter, M.
Wang, A.
Shedlock, D.
Wu, M.
Newson, M.
Xing, L.
Ansbacher, W.
Fahrig, R.
Star‐Lack, J. - Abstract:
- Abstract : Purpose: Combined kV–MV cone‐beam tomography (CBCT) imaging has been proposed for two potentially important image‐guided radiotherapy applications: (a) scan time reduction (STR) and (b) metal artifact reduction (MAR). However, the feasibility of these techniques has been in question due to the low detective quantum efficiencies (DQEs) of commercially available electronic portal imagers (EPIDs). The goal of the work was to test whether a prototype high DQE MV detector can be used to generate acceptable quality pretreatment CBCT images at acceptable dose levels. Methods: 6MV and 100 kVp projection data were acquired on a Truebeam system (Varian, Palo Alto, CA). The MV data were acquired using a prototype EPID containing two scintillators (a) a standard copper‐gadolinium oxysulfide (Cu‐GOS) screen having a zero‐frequency DQE (DQE(0)) value of 1.4%, and (b) a prototype‐focused cadmium tungstate (CWO) pixelated "strip" with a DQE(0) = 22%. The kV data were acquired using the standard onboard imager (DQE(0) = 70%). The angular spacing of the MV projections was 0.81° and the source output was 0.03 MU/projection while the kV projections were acquired with an angular spacing of 0.4° at 0.3 mAs/projection. Image quality was evaluated using (a) an 18‐cm diameter electron density phantom (CIRS, Norfolk, VA) with nine contrast inserts and (b) the resolution section of the 20‐cm diameter Catphan phantom (The Phantom Laboratory, Greenwich, NY). For the MAR studies, two opposingAbstract : Purpose: Combined kV–MV cone‐beam tomography (CBCT) imaging has been proposed for two potentially important image‐guided radiotherapy applications: (a) scan time reduction (STR) and (b) metal artifact reduction (MAR). However, the feasibility of these techniques has been in question due to the low detective quantum efficiencies (DQEs) of commercially available electronic portal imagers (EPIDs). The goal of the work was to test whether a prototype high DQE MV detector can be used to generate acceptable quality pretreatment CBCT images at acceptable dose levels. Methods: 6MV and 100 kVp projection data were acquired on a Truebeam system (Varian, Palo Alto, CA). The MV data were acquired using a prototype EPID containing two scintillators (a) a standard copper‐gadolinium oxysulfide (Cu‐GOS) screen having a zero‐frequency DQE (DQE(0)) value of 1.4%, and (b) a prototype‐focused cadmium tungstate (CWO) pixelated "strip" with a DQE(0) = 22%. The kV data were acquired using the standard onboard imager (DQE(0) = 70%). The angular spacing of the MV projections was 0.81° and the source output was 0.03 MU/projection while the kV projections were acquired with an angular spacing of 0.4° at 0.3 mAs/projection. Image quality was evaluated using (a) an 18‐cm diameter electron density phantom (CIRS, Norfolk, VA) with nine contrast inserts and (b) the resolution section of the 20‐cm diameter Catphan phantom (The Phantom Laboratory, Greenwich, NY). For the MAR studies, two opposing CIRS phantom inserts were replaced by steel rods. The reconstruction methods were based on combining MV and kV data into one sinogram. The MAR reconstruction utilized mostly kV raw data with only those rays corrupted by metal requiring replacement with MV data (total absorbed dose = 0.7 cGy). For the STR study, projections from partially overlapping 105°kV and MV acquisitions were combined to create a complete dataset that could have been acquired in 18 sec (absorbed dose = 2.5 cGy). MV‐only (4.3 cGy) and kV‐only (0.3 cGy) images were also reconstructed. Results: The average signal‐to‐noise ratio (SNR) of the inserts in the MV‐only CWO and GOS CIRS phantom images were 0.62× and 0.12× the SNR of the inserts in kV‐only image, respectively. The limiting spatial resolutions in the MV‐only GOS, MV‐only CWO, and kV‐only Catphan images were 3, 6, and 8 lp/cm, respectively. In the combined kV/CWO STR reconstruction, all contrast inserts were visible while only two were detectable in the kV/Cu‐GOS image due to high levels of noise (average SNRs of kV/CWO and kV/GOS inserts were 0.97× and 0.18× the SNR of the kV‐only inserts, respectively). In the kV–MV MAR reconstructions, streaking artifacts were substantially reduced with all inserts becoming clearly visible in the kV/CWO image while only two were visible in the kV/Cu‐GOS image (average SNRs of the kV/CWO and kV/Cu‐GOS CIRS with metal inserts were 0.94× and 0.35× the SNRs of the kV‐only CIRS without metal inserts). Conclusions: We have demonstrated that a high‐DQE MV detector can be applied to generating high‐quality combined kV–MV images for SRT and MAR. Clinically acceptable doses were utilized. … (more)
- Is Part Of:
- Medical physics. Volume 46:Issue 2(2019)
- Journal:
- Medical physics
- Issue:
- Volume 46:Issue 2(2019)
- Issue Display:
- Volume 46, Issue 2 (2019)
- Year:
- 2019
- Volume:
- 46
- Issue:
- 2
- Issue Sort Value:
- 2019-0046-0002-0000
- Page Start:
- 563
- Page End:
- 575
- Publication Date:
- 2018-12-06
- Subjects:
- CBCT imaging -- high DQE -- metal artifact reduction -- MV detector -- scan time reduction
Medical physics -- Periodicals
Medical physics
Geneeskunde
Natuurkunde
Toepassingen
Biophysics
Periodicals
Periodicals
Electronic journals
610.153 - Journal URLs:
- http://scitation.aip.org/content/aapm/journal/medphys ↗
https://aapm.onlinelibrary.wiley.com/journal/24734209 ↗
http://www.aip.org/ ↗ - DOI:
- 10.1002/mp.13291 ↗
- Languages:
- English
- ISSNs:
- 0094-2405
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 5531.130000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 18049.xml