A practical method for the reuse of nanoDot OSLDs and predicting sensitivities up to at least 7000 cGy. Issue 4 (21st February 2020)
- Record Type:
- Journal Article
- Title:
- A practical method for the reuse of nanoDot OSLDs and predicting sensitivities up to at least 7000 cGy. Issue 4 (21st February 2020)
- Main Title:
- A practical method for the reuse of nanoDot OSLDs and predicting sensitivities up to at least 7000 cGy
- Authors:
- Zhuang, Audrey H.
Olch, Arthur J. - Abstract:
- Abstract : Purpose: Optically stimulated luminescence dosimeter (OSLDs) are often used to make in vivo dose measurements. Most users calibrate the OSLDs using the software provided by the vendor which typically is intended for doses up to about 300 cGy with an uncertainty of ±5.5%. OSLDs that reach that dose are then discarded, and new ones are purchased and calibrated. A method has been developed for reusing OSLDs and predicting dose sensitivity up to at least 7000 cGy. Materials and methods: The nanoDot OSLDs used in this study were routinely used to do in vivo measurements for TBI patients. Instead of using the calibration program provided by the vendor, each nanoDot was bleached to about 100 counts (~0.1 cGy), then calibrated with 50 cGy to produce a sensitivity specific to each nanoDot prior to the patient measurement. NanoDots were read in the hardware mode and the sensitivity factor was applied manually to subsequent patient in‐vivo TBI measurements. This was followed by bleaching prior to the next use. The changes of nanoDot sensitivity relative to accumulated dose were analyzed among nine nanoDots. In addition, a method to predict a nanoDot's sensitivity was investigated which aims to reduce the number of sensitivity calibrations while retaining dosimetric accuracy. Results: Individual per‐use nanodot calibrations were performed up to 7000 cGy for 37 clinical TBI patients. Among the nine nanoDots analyzed in this paper, the sensitivity vs accumulated dose decreasedAbstract : Purpose: Optically stimulated luminescence dosimeter (OSLDs) are often used to make in vivo dose measurements. Most users calibrate the OSLDs using the software provided by the vendor which typically is intended for doses up to about 300 cGy with an uncertainty of ±5.5%. OSLDs that reach that dose are then discarded, and new ones are purchased and calibrated. A method has been developed for reusing OSLDs and predicting dose sensitivity up to at least 7000 cGy. Materials and methods: The nanoDot OSLDs used in this study were routinely used to do in vivo measurements for TBI patients. Instead of using the calibration program provided by the vendor, each nanoDot was bleached to about 100 counts (~0.1 cGy), then calibrated with 50 cGy to produce a sensitivity specific to each nanoDot prior to the patient measurement. NanoDots were read in the hardware mode and the sensitivity factor was applied manually to subsequent patient in‐vivo TBI measurements. This was followed by bleaching prior to the next use. The changes of nanoDot sensitivity relative to accumulated dose were analyzed among nine nanoDots. In addition, a method to predict a nanoDot's sensitivity was investigated which aims to reduce the number of sensitivity calibrations while retaining dosimetric accuracy. Results: Individual per‐use nanodot calibrations were performed up to 7000 cGy for 37 clinical TBI patients. Among the nine nanoDots analyzed in this paper, the sensitivity vs accumulated dose decreased linearly up to about 3000 cGy, with linear fitting curve R 2 values above 0.93. After 3000 cGy of accumulated dose, the sensitivity started to plateau and tended to increase by 6000 cGy, with 2nd order polynomial curve R 2 values above 0.94. With this finding, an efficient and accurate method to predict nanoDots' sensitivities was developed. With the method applied to the nine OSLDs, a total of 127 sensitivities were predicted and retrospectively compared with measured sensitivities. The predicted sensitivities agreed with measured sensitivities within ±4.0% with an average of −0.8%. Conclusions: This study is the first to demonstrate the reuse of nanoDot OSLDs on numerous patients with accumulated dose up to 7000 cGy. Our nanoDot re‐usage methodology is accurate, cost‐saving and feasible. A time‐saving method is also provided that allows a user to reuse a nanoDot with sensitivities predicted with better accuracy than the 5.5% value provided by the conventional batch calibration method. … (more)
- Is Part Of:
- Medical physics. Volume 47:Issue 4(2020)
- Journal:
- Medical physics
- Issue:
- Volume 47:Issue 4(2020)
- Issue Display:
- Volume 47, Issue 4 (2020)
- Year:
- 2020
- Volume:
- 47
- Issue:
- 4
- Issue Sort Value:
- 2020-0047-0004-0000
- Page Start:
- 1481
- Page End:
- 1488
- Publication Date:
- 2020-02-21
- Subjects:
- in vivo dosimetry -- nanoDot -- optically stimulated luminescent dosimeters -- OSLD
Medical physics -- Periodicals
Medical physics
Geneeskunde
Natuurkunde
Toepassingen
Biophysics
Periodicals
Periodicals
Electronic journals
610.153 - Journal URLs:
- http://scitation.aip.org/content/aapm/journal/medphys ↗
https://aapm.onlinelibrary.wiley.com/journal/24734209 ↗
http://www.aip.org/ ↗ - DOI:
- 10.1002/mp.14059 ↗
- Languages:
- English
- ISSNs:
- 0094-2405
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 5531.130000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 18072.xml