THU0671 Can an inner disposable glove be used under an electrogoniometric glove for measuring finger movement without loss of accuracy?. (15th June 2017)
- Record Type:
- Journal Article
- Title:
- THU0671 Can an inner disposable glove be used under an electrogoniometric glove for measuring finger movement without loss of accuracy?. (15th June 2017)
- Main Title:
- THU0671 Can an inner disposable glove be used under an electrogoniometric glove for measuring finger movement without loss of accuracy?
- Authors:
- Connolly, J
Gardiner, P
Condell, J
Curran, K
Small, D - Abstract:
- Abstract : Background: Improving joint mobility is an important outcome for patients with arthritis, but finger joint range of motion is rarely measured in clinic. Electronic gloves with movement sensors have been developed to measure joint movement accurately and it is now possible to assess dynamic mobility of the finger joints. However these gloves are expensive and it is likely that when carrying out measurements in the patient population they would be used with inner disposable gloves to avoid nonsocomical infection. Establishing accuracy and usability of electronic gloves whilst wearing disposable inner gloves is therefore an important pre-requisite for studies in patients with arthritis. Objectives: To establish the accuracy and repeatability of measurements of finger movement obtained using two different electrogoniometric gloves worn with and without an inner disposable glove. Methods: We used two different types of electrogoniometric glove for the purpose of this study. One is the commercially available 5DT dataglove 14 Ultra (5DT, 2011) and the other was produced to our specifications by Tyndall National Institute, University College Cork. We called this the "IMU glove". We developed a graphical interface for both devices to facilitate detailed evaluation of joint movement in each finger. Both gloves were tested using a protocol adapted from Dipietro, Sabatini, & Dario, (2003). Results: Table 1 displays comparison of Coefficient of Variation (CV) readings for bothAbstract : Background: Improving joint mobility is an important outcome for patients with arthritis, but finger joint range of motion is rarely measured in clinic. Electronic gloves with movement sensors have been developed to measure joint movement accurately and it is now possible to assess dynamic mobility of the finger joints. However these gloves are expensive and it is likely that when carrying out measurements in the patient population they would be used with inner disposable gloves to avoid nonsocomical infection. Establishing accuracy and usability of electronic gloves whilst wearing disposable inner gloves is therefore an important pre-requisite for studies in patients with arthritis. Objectives: To establish the accuracy and repeatability of measurements of finger movement obtained using two different electrogoniometric gloves worn with and without an inner disposable glove. Methods: We used two different types of electrogoniometric glove for the purpose of this study. One is the commercially available 5DT dataglove 14 Ultra (5DT, 2011) and the other was produced to our specifications by Tyndall National Institute, University College Cork. We called this the "IMU glove". We developed a graphical interface for both devices to facilitate detailed evaluation of joint movement in each finger. Both gloves were tested using a protocol adapted from Dipietro, Sabatini, & Dario, (2003). Results: Table 1 displays comparison of Coefficient of Variation (CV) readings for both data gloves. Figure shows this information graphically. Results show no significant change for 5DT angular readings with and without a surgical glove worn underneath the data glove. Results for PIP sensors show an improvement in repeatability with a surgical glove. CV variance was smaller for MCP sensors with a surgical glove worn underneath the data glove compared with no surgical glove. CV for the IMU data glove show negligible changes in MCP readings when a surgical glove is worn underneath. PIP readings show small changes when using a surgical glove. Conclusions: Inner disposable gloves can be worn when using electrogoniometric gloves for testing finger movement without loss of accuracy or any significant discomfort in patients with arthritis. References: 5DT, 2011. 5DT Data Glove 14 Ultra [WWW Document]. URL http://www.5dt.com/products/pdataglove14.html (accessed 1.10.12). Dipietro, L., Sabatini, A.M., Dario, P., 2003. Evaluation of an instrumented glove for hand-movement acquisition. J. Rehabil. Res. Dev. 40, 179–89. Disclosure of Interest: None declared … (more)
- Is Part Of:
- Annals of the rheumatic diseases. Volume 76(2017)Supplement 2
- Journal:
- Annals of the rheumatic diseases
- Issue:
- Volume 76(2017)Supplement 2
- Issue Display:
- Volume 76, Issue 2 (2017)
- Year:
- 2017
- Volume:
- 76
- Issue:
- 2
- Issue Sort Value:
- 2017-0076-0002-0000
- Page Start:
- 460
- Page End:
- 460
- Publication Date:
- 2017-06-15
- Subjects:
- Rheumatism -- Periodicals
616.723005 - Journal URLs:
- http://ard.bmjjournals.com/ ↗
http://www.pubmedcentral.nih.gov/tocrender.fcgi?journal=149&action=archive ↗
http://www.bmj.com/archive ↗
http://gateway.ovid.com/server3/ovidweb.cgi?T=JS&MODE=ovid&D=ovft&PAGE=titles&SEARCH=annals+of+the+rheumatic+diseases.tj&NEWS=N ↗ - DOI:
- 10.1136/annrheumdis-2017-eular.4429 ↗
- Languages:
- English
- ISSNs:
- 0003-4967
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 18004.xml