Technical Note: Experimental characterization of the dose deposition in parallel MRI‐linacs at various magnetic field strengths. Issue 11 (9th September 2019)
- Record Type:
- Journal Article
- Title:
- Technical Note: Experimental characterization of the dose deposition in parallel MRI‐linacs at various magnetic field strengths. Issue 11 (9th September 2019)
- Main Title:
- Technical Note: Experimental characterization of the dose deposition in parallel MRI‐linacs at various magnetic field strengths
- Authors:
- Begg, Jarrad
Alnaghy, Sarah J.
Causer, Trent
Alharthi, Thahabah
George, Armia
Glaubes, Laura
Dong, Bin
Goozee, Gary
Keall, Paul
Jelen, Urszula
Liney, Gary
Holloway, Lois - Abstract:
- Abstract : Purpose: Dose deposition measurements for parallel MRI‐linacs have previously only shown comparisons between 0 T and a single available magnetic field. The Australian MRI‐Linac consists of a magnet coupled with a dual energy linear accelerator and a 120 leaf Multi‐Leaf Collimator with the radiation beam parallel to the magnetic field. Two different magnets, with field strengths of 1 and 1.5 T, were used during prototyping. This work aims to characterize the impact of the magnetic field at 1 and 1.5 T on dose deposition, possible by comparing dosimetry measured at both magnetic field strengths to measurements without the magnetic field. Methods: Dose deposition measurements focused on a comparison of beam quality (TPR20/10 ), PDD, profiles at various depths, surface doses, and field size output factors. Measurements were acquired at 0, 1, and 1.5 T. Beam quality was measured using an ion chamber in solid water at isocenter with appropriate TPR20/10 buildup. PDDs and profiles were acquired via EBT3 film placed in solid water either parallel or perpendicular to the radiation beam. Films at surface were used to determine surface dose. Output factors were measured in solid water using an ion chamber at isocenter with 10 cm solid water buildup. Results: Beam quality was within ±0.5% of the 0 T value for the 1 and 1.5 T magnetic field strengths. PDDs and profiles showed agreement for the three magnetic field strengths at depths beyond 20 mm. Deposited dose increased atAbstract : Purpose: Dose deposition measurements for parallel MRI‐linacs have previously only shown comparisons between 0 T and a single available magnetic field. The Australian MRI‐Linac consists of a magnet coupled with a dual energy linear accelerator and a 120 leaf Multi‐Leaf Collimator with the radiation beam parallel to the magnetic field. Two different magnets, with field strengths of 1 and 1.5 T, were used during prototyping. This work aims to characterize the impact of the magnetic field at 1 and 1.5 T on dose deposition, possible by comparing dosimetry measured at both magnetic field strengths to measurements without the magnetic field. Methods: Dose deposition measurements focused on a comparison of beam quality (TPR20/10 ), PDD, profiles at various depths, surface doses, and field size output factors. Measurements were acquired at 0, 1, and 1.5 T. Beam quality was measured using an ion chamber in solid water at isocenter with appropriate TPR20/10 buildup. PDDs and profiles were acquired via EBT3 film placed in solid water either parallel or perpendicular to the radiation beam. Films at surface were used to determine surface dose. Output factors were measured in solid water using an ion chamber at isocenter with 10 cm solid water buildup. Results: Beam quality was within ±0.5% of the 0 T value for the 1 and 1.5 T magnetic field strengths. PDDs and profiles showed agreement for the three magnetic field strengths at depths beyond 20 mm. Deposited dose increased at shallower depths due to electron focusing. Output factors showed agreement within 1%. Conclusion: Dose deposition at depth for a parallel MRI‐linac was not significantly impacted by either a 1 or 1.5 T magnetic field. PDDs and profiles at shallow depths and surface dose measurements showed significant differences between 0, 1, and 1.5 T due to electron focusing. … (more)
- Is Part Of:
- Medical physics. Volume 46:Issue 11(2019)
- Journal:
- Medical physics
- Issue:
- Volume 46:Issue 11(2019)
- Issue Display:
- Volume 46, Issue 11 (2019)
- Year:
- 2019
- Volume:
- 46
- Issue:
- 11
- Issue Sort Value:
- 2019-0046-0011-0000
- Page Start:
- 5152
- Page End:
- 5158
- Publication Date:
- 2019-09-09
- Subjects:
- dose deposition -- magnetic field -- MRI‐Linac -- parallel orientation
Medical physics -- Periodicals
Medical physics
Geneeskunde
Natuurkunde
Toepassingen
Biophysics
Periodicals
Periodicals
Electronic journals
610.153 - Journal URLs:
- http://scitation.aip.org/content/aapm/journal/medphys ↗
https://aapm.onlinelibrary.wiley.com/journal/24734209 ↗
http://www.aip.org/ ↗ - DOI:
- 10.1002/mp.13767 ↗
- Languages:
- English
- ISSNs:
- 0094-2405
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 5531.130000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 18010.xml