Combining C-arm CT with a new remote operated positioning and guidance system for guidance of minimally invasive spine interventions. (5th March 2014)
- Record Type:
- Journal Article
- Title:
- Combining C-arm CT with a new remote operated positioning and guidance system for guidance of minimally invasive spine interventions. (5th March 2014)
- Main Title:
- Combining C-arm CT with a new remote operated positioning and guidance system for guidance of minimally invasive spine interventions
- Authors:
- Czerny, Christoph
Eichler, Katrin
Croissant, Yann
Schulz, Boris
Kronreif, Gernot
Schmidt, Renate
von Roden, Martin
Schomerus, Christof
Vogl, Thomas J
Marzi, Ingo
Zangos, Stephan - Abstract:
- Abstract : Objective: To report our experience using C-arm cone beam CT (C-arm CBCT) combined with the new remote operated positioning and guidance system, iSYS1, for needle guidance during spinal interventions. Methods: A C-arm CBCT with a flat panel angiography system was acquired (Artis Zeego; Siemens Healthcare Sector, Forchheim, Germany). Reconstruction of CT-like images and planning of the needle path were performed using a common workstation. The needle holder of iSYS1 acted as a guide during insertion of Kirschner (K) wires. 20 percutaneous K wires were placed in the pedicles at T2–T3, T7–T12, and L1–L2 in a cadaver specimen. Postprocedure C-arm CBCT scans were obtained to confirm the accuracy of the K wire placement. Results: All K wire placements were successfully performed. Mean planning time with Syngo iGuide was 4:16 min, mean positioning time of iSYS1 was 3:35 min, and mean placement time of the K wires was 2:22 min. Mean total intervention time was 10:13 min per pedicle. A mean deviation of 0.35 mm between the planned path and the placed K wire with a mean path length of 6.73 cm was documented. Conclusions: Our results demonstrate the potential of combining C-arm CBCT with iSYS1 for safe and accurate percutaneous placement of pedicle K wires in spinal interventions.
- Is Part Of:
- Journal of neurointerventional surgery. Volume 7:Number 4(2015)
- Journal:
- Journal of neurointerventional surgery
- Issue:
- Volume 7:Number 4(2015)
- Issue Display:
- Volume 7, Issue 4 (2015)
- Year:
- 2015
- Volume:
- 7
- Issue:
- 4
- Issue Sort Value:
- 2015-0007-0004-0000
- Page Start:
- 303
- Page End:
- 308
- Publication Date:
- 2014-03-05
- Subjects:
- Intervention -- Spine -- guidewire -- Navigation -- CT
Nervous system -- Surgery -- Periodicals
Cerebrovascular disease -- Surgery -- Periodicals
617.48 - Journal URLs:
- http://www.bmj.com/archive ↗
http://jnis.bmj.com/ ↗ - DOI:
- 10.1136/neurintsurg-2013-011034 ↗
- Languages:
- English
- ISSNs:
- 1759-8478
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 17984.xml