Agreement between the Takeda UA-731 automatic blood pressure measuring device and the manual mercury sphygmomanometer: an assessment under field conditions in Newcastle upon Tyne, UK. Issue 2 (April 1996)
- Record Type:
- Journal Article
- Title:
- Agreement between the Takeda UA-731 automatic blood pressure measuring device and the manual mercury sphygmomanometer: an assessment under field conditions in Newcastle upon Tyne, UK. Issue 2 (April 1996)
- Main Title:
- Agreement between the Takeda UA-731 automatic blood pressure measuring device and the manual mercury sphygmomanometer: an assessment under field conditions in Newcastle upon Tyne, UK.
- Authors:
- Cartwright, C
Unwin, N
Stephenson, P - Abstract:
- Abstract : STUDY OBJECTIVE: To assess agreement between two Takeda UA-731 automatic blood pressure measuring devices (referred to as machines A and B) and two manual mercury sphygmomanometers. DESIGN: A 'Y' connector attached each Takeda UA-731 to a manual mercury sphygmomanometer. Simultaneous measurements were made on adult subjects. SETTING: A population based cardiovascular disease survey in Newcastle upon Tyne, UK. PARTICIPANTS: Measurements on machine A were compared in 71 individuals (all women), and on machine B in 75 individuals (9 men, 66 women). The age range of subjects was 28 to 76 years and median ages were 59 years for machine A and 50 years for machine B. MAIN RESULTS: Blood pressure (mmHg) ranged from 72 to 212 systolic and 44 to 102 diastolic. Both Takedas gave significantly lower readings than the manual devices for systolic and diastolic pressures: differences were mean (SD: 95% CI) 3.7 mmHg (6.5: 2.2, 5.2) for machine A systolic, 2.3 mmHg (4.5: 1.3, 3.4) machine A diastolic; 1.8 mmHg (6.2: 0.4, 3.3) machine B systolic, and 1.8 (4.4: 0.8, 2.8) machine B diastolic. On the British Hypertension Society criteria, machine A was graded C on systolic measurements and B on diastolic; machine B was graded B on both systolic and diastolic measurements. CONCLUSIONS: The performance of these machines compares favourably with the Dinamap 8100, recently adopted for survey work by the Department of Health. The Takeda UA-731 looks promising for epidemiological surveyAbstract : STUDY OBJECTIVE: To assess agreement between two Takeda UA-731 automatic blood pressure measuring devices (referred to as machines A and B) and two manual mercury sphygmomanometers. DESIGN: A 'Y' connector attached each Takeda UA-731 to a manual mercury sphygmomanometer. Simultaneous measurements were made on adult subjects. SETTING: A population based cardiovascular disease survey in Newcastle upon Tyne, UK. PARTICIPANTS: Measurements on machine A were compared in 71 individuals (all women), and on machine B in 75 individuals (9 men, 66 women). The age range of subjects was 28 to 76 years and median ages were 59 years for machine A and 50 years for machine B. MAIN RESULTS: Blood pressure (mmHg) ranged from 72 to 212 systolic and 44 to 102 diastolic. Both Takedas gave significantly lower readings than the manual devices for systolic and diastolic pressures: differences were mean (SD: 95% CI) 3.7 mmHg (6.5: 2.2, 5.2) for machine A systolic, 2.3 mmHg (4.5: 1.3, 3.4) machine A diastolic; 1.8 mmHg (6.2: 0.4, 3.3) machine B systolic, and 1.8 (4.4: 0.8, 2.8) machine B diastolic. On the British Hypertension Society criteria, machine A was graded C on systolic measurements and B on diastolic; machine B was graded B on both systolic and diastolic measurements. CONCLUSIONS: The performance of these machines compares favourably with the Dinamap 8100, recently adopted for survey work by the Department of Health. The Takeda UA-731 looks promising for epidemiological survey work but before it can be fully recommended further evaluations are needed. … (more)
- Is Part Of:
- Journal of epidemiology and community health. Volume 50:Issue 2(1996)
- Journal:
- Journal of epidemiology and community health
- Issue:
- Volume 50:Issue 2(1996)
- Issue Display:
- Volume 50, Issue 2 (1996)
- Year:
- 1996
- Volume:
- 50
- Issue:
- 2
- Issue Sort Value:
- 1996-0050-0002-0000
- Page Start:
- 218
- Page End:
- 222
- Publication Date:
- 1996-04
- Subjects:
- Public health -- Periodicals
Epidemiology -- Periodicals
614.4 - Journal URLs:
- http://jech.bmj.com/ ↗
http://www.jstor.org/journals/0143005X.html ↗
http://www.pubmedcentral.nih.gov/tocrender.fcgi?journal=165&action=archive ↗
http://www.bmj.com/archive ↗ - DOI:
- 10.1136/jech.50.2.218 ↗
- Languages:
- English
- ISSNs:
- 0143-005X
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 17773.xml