Radioguided occult lesion localisation using iodine‐125 seeds ('ROLLIS') for removal of impalpable breast lesions: First Australian experience. Issue 4 (14th April 2015)
- Record Type:
- Journal Article
- Title:
- Radioguided occult lesion localisation using iodine‐125 seeds ('ROLLIS') for removal of impalpable breast lesions: First Australian experience. Issue 4 (14th April 2015)
- Main Title:
- Radioguided occult lesion localisation using iodine‐125 seeds ('ROLLIS') for removal of impalpable breast lesions: First Australian experience
- Authors:
- Taylor, Donna B.
Bourke, Anita G.
Westcott, Eliza
Burrage, John
Latham, Bruce
Riley, Paul
Ballal, Helen
Kamyab, Roshi
Frost, Felicity
Dissanayake, Deepthi
Landman, Joanne
Phillips, Michael
Saunders, Christobel - Abstract:
- Abstract: Introduction: Approximately one‐third of breast cancers are impalpable and require pre‐operative image‐guided localisation. Hook‐wire localisation (HWL) is commonly used but has several disadvantages. Use of a low‐activity radioactive iodine‐125 seed is a promising alternative technique used in the USA and the Netherlands. This pilot study describes the first use of this in Australia. Methods: In this prospective pilot study, 21 participants with biopsy‐proven breast cancer underwent radioguided occult lesion localisation using iodine‐125 seed(s) (ROLLIS) with insertion of a hook‐wire for back up. Sentinel node biopsy was performed where indicated. Ease of hook‐wire and seed insertion, duration of the procedure, dependence on the seed versus hook‐wire during surgery, lesion location within the specimen, histopathology including size of radial margins, the ease of seed retrieval in pathology, and safe return of seeds for disposal were documented. Radiation dosimetry of staff was performed. Results: All seeds were placed within 3.5 mm of the lesion. All lesions and seeds were removed. One participant needed re‐excision for involved margins. Radiologists and surgeons both preferred ROLLIS. Surgeons were able to depend on the seed for localisation in all but one case. Sentinel node biopsy was successfully performed when required. Pathologists found seed retrieval quick and easy, with no detrimental effect on tissue processing. No radiation doses measurably aboveAbstract: Introduction: Approximately one‐third of breast cancers are impalpable and require pre‐operative image‐guided localisation. Hook‐wire localisation (HWL) is commonly used but has several disadvantages. Use of a low‐activity radioactive iodine‐125 seed is a promising alternative technique used in the USA and the Netherlands. This pilot study describes the first use of this in Australia. Methods: In this prospective pilot study, 21 participants with biopsy‐proven breast cancer underwent radioguided occult lesion localisation using iodine‐125 seed(s) (ROLLIS) with insertion of a hook‐wire for back up. Sentinel node biopsy was performed where indicated. Ease of hook‐wire and seed insertion, duration of the procedure, dependence on the seed versus hook‐wire during surgery, lesion location within the specimen, histopathology including size of radial margins, the ease of seed retrieval in pathology, and safe return of seeds for disposal were documented. Radiation dosimetry of staff was performed. Results: All seeds were placed within 3.5 mm of the lesion. All lesions and seeds were removed. One participant needed re‐excision for involved margins. Radiologists and surgeons both preferred ROLLIS. Surgeons were able to depend on the seed for localisation in all but one case. Sentinel node biopsy was successfully performed when required. Pathologists found seed retrieval quick and easy, with no detrimental effect on tissue processing. No radiation doses measurably above background were received by staff. Conclusion: ROLLIS is an easily learnt, safe and effective alternative technique to standard HWL. … (more)
- Is Part Of:
- Journal of medical imaging and radiation oncology. Volume 59:Issue 4(2015:Aug.)
- Journal:
- Journal of medical imaging and radiation oncology
- Issue:
- Volume 59:Issue 4(2015:Aug.)
- Issue Display:
- Volume 59, Issue 4 (2015)
- Year:
- 2015
- Volume:
- 59
- Issue:
- 4
- Issue Sort Value:
- 2015-0059-0004-0000
- Page Start:
- 411
- Page End:
- 420
- Publication Date:
- 2015-04-14
- Subjects:
- breast conserving surgery -- image‐guided localisation -- iodine‐125 -- radioactive seed localisation
Radiology, Medical -- Periodicals
Radiology, Medical -- Australasia -- Periodicals
616.0757 - Journal URLs:
- http://onlinelibrary.wiley.com/journal/10.1111/(ISSN)1754-9485 ↗
http://onlinelibrary.wiley.com/ ↗ - DOI:
- 10.1111/1754-9485.12302 ↗
- Languages:
- English
- ISSNs:
- 1754-9477
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 5017.072080
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 17758.xml