Comparison of central and peripheral corneal thickness measurements with scanning-slit, Scheimpflug and Fourier-domain ocular coherence tomography. Issue 9 (30th March 2015)
- Record Type:
- Journal Article
- Title:
- Comparison of central and peripheral corneal thickness measurements with scanning-slit, Scheimpflug and Fourier-domain ocular coherence tomography. Issue 9 (30th March 2015)
- Main Title:
- Comparison of central and peripheral corneal thickness measurements with scanning-slit, Scheimpflug and Fourier-domain ocular coherence tomography
- Authors:
- Randleman, J Bradley
Lynn, Michael J
Perez-Straziota, Claudia E
Weissman, Heather M
Kim, Sang Woo - Abstract:
- Abstract : Purpose: To compare central, regional and relational corneal thickness values obtained with multiple technologies in normal patients and to determine their equivalence and interchangeability. Methods: Retrospective analysis of 100 eyes from 50 patients evaluated by ultrasound pachymetry (Pachette II), scanning-slit (Orbscan II), Scheimpflug (Pentacam HR) and spectral-domain ocular coherence tomography (OCT) (RTVue-100) obtained as average values (OCT-A) and point measurements (OCT-P). Measurements included central corneal thickness (CCT) for all technologies and thinnest corneal thickness for scanning-slit, Scheimpflug and OCT. Peripheral thickness measurements were obtained at the 3 mm radius in the superior (S), nasal (N), inferior (I) and temporal (T) regions. Results: CCT values were: 563.9±36.1μ ultrasound, 570.9±36.1μ scanning-slit, 552.8±33.8μ Scheimpflug, 550.5±32.7μ (OCT-A), 549.4±32.7μ (OCT-P). Ultrasound and scanning-slit were significantly different from each other (p<0.0001), and both were significantly different from all other devices (p<0.0001), while Scheimpflug was similar to OCT-A and OCT-P (p=0.4). Differences between CCT and thinnest corneal thickness were significantly different from all technologies except scanning-slit and OCT-A. For peripheral values, almost all locations' measurements were significantly different from one another (p<0.0001). Superior–inferior values and ratios were also significantly different from one another for almostAbstract : Purpose: To compare central, regional and relational corneal thickness values obtained with multiple technologies in normal patients and to determine their equivalence and interchangeability. Methods: Retrospective analysis of 100 eyes from 50 patients evaluated by ultrasound pachymetry (Pachette II), scanning-slit (Orbscan II), Scheimpflug (Pentacam HR) and spectral-domain ocular coherence tomography (OCT) (RTVue-100) obtained as average values (OCT-A) and point measurements (OCT-P). Measurements included central corneal thickness (CCT) for all technologies and thinnest corneal thickness for scanning-slit, Scheimpflug and OCT. Peripheral thickness measurements were obtained at the 3 mm radius in the superior (S), nasal (N), inferior (I) and temporal (T) regions. Results: CCT values were: 563.9±36.1μ ultrasound, 570.9±36.1μ scanning-slit, 552.8±33.8μ Scheimpflug, 550.5±32.7μ (OCT-A), 549.4±32.7μ (OCT-P). Ultrasound and scanning-slit were significantly different from each other (p<0.0001), and both were significantly different from all other devices (p<0.0001), while Scheimpflug was similar to OCT-A and OCT-P (p=0.4). Differences between CCT and thinnest corneal thickness were significantly different from all technologies except scanning-slit and OCT-A. For peripheral values, almost all locations' measurements were significantly different from one another (p<0.0001). Superior–inferior values and ratios were also significantly different from one another for almost all devices with no consistent patterns detectible. Conclusions: There are significant clinically relevant differences between regional and relational thickness measurements obtained with ultrasound, scanning-slit, Scheimpflug and OCT devices. Screening metrics devised for one system do not appear directly applicable to other measurement systems. … (more)
- Is Part Of:
- British journal of ophthalmology. Volume 99:Issue 9(2015)
- Journal:
- British journal of ophthalmology
- Issue:
- Volume 99:Issue 9(2015)
- Issue Display:
- Volume 99, Issue 9 (2015)
- Year:
- 2015
- Volume:
- 99
- Issue:
- 9
- Issue Sort Value:
- 2015-0099-0009-0000
- Page Start:
- 1176
- Page End:
- 1181
- Publication Date:
- 2015-03-30
- Subjects:
- Cornea -- Diagnostic tests/Investigation -- Imaging -- Treatment Lasers
Ophthalmology -- Periodicals
617.7 - Journal URLs:
- http://bjo.bmj.com/ ↗
http://bjo.bmjjournals.com/ ↗
http://www.bmj.com/archive ↗ - DOI:
- 10.1136/bjophthalmol-2014-306340 ↗
- Languages:
- English
- ISSNs:
- 0007-1161
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 17715.xml