Evaluation of Biox IIA ear oximeter. Issue 11 (November 1985)
- Record Type:
- Journal Article
- Title:
- Evaluation of Biox IIA ear oximeter. Issue 11 (November 1985)
- Main Title:
- Evaluation of Biox IIA ear oximeter.
- Authors:
- Tweeddale, P M
Douglas, N J - Abstract:
- Abstract : Arterial blood gas tensions, pH, and carboxyhaemoglobin were measured on 322 occasions in 165 patients and the actual oxygen saturation of the haemoglobin (Sao2) was compared with the ear oxygen saturation (SEO2) recorded during the arterial sampling with a Biox IIA ear oximeter. The overall agreement between SEO2 and Sao2 was good, with a mean difference in saturation (SEO2-Sao2) of + 1.5% (SD 3.0). The difference in saturation remained similar at all levels of arterial saturation observed and was unaffected by carboxyhaemoglobin concentration. On four occasions (1% of readings), however, SEO2 and Sao2 differed by more than 10% and such occasional errors might be misleading in clinical practice.
- Is Part Of:
- Thorax. Volume 40:Issue 11(1985)
- Journal:
- Thorax
- Issue:
- Volume 40:Issue 11(1985)
- Issue Display:
- Volume 40, Issue 11 (1985)
- Year:
- 1985
- Volume:
- 40
- Issue:
- 11
- Issue Sort Value:
- 1985-0040-0011-0000
- Page Start:
- 825
- Page End:
- 827
- Publication Date:
- 1985-11
- Subjects:
- Chest -- Diseases -- Periodicals
Thorax
Chest -- Diseases
Periodicals
Periodicals
617.54 - Journal URLs:
- http://thorax.bmjjournals.com/contents-by-date.0.shtml ↗
http://www.bmj.com/archive ↗ - DOI:
- 10.1136/thx.40.11.825 ↗
- Languages:
- English
- ISSNs:
- 0040-6376
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 17653.xml