ECMO implantation training: Needle penetration in 3D printable materials and porcine aorta. (November 2021)
- Record Type:
- Journal Article
- Title:
- ECMO implantation training: Needle penetration in 3D printable materials and porcine aorta. (November 2021)
- Main Title:
- ECMO implantation training: Needle penetration in 3D printable materials and porcine aorta
- Authors:
- Salewski, Christoph
Spintzyk, Sebastian
von Steuben, Thore
Sandoval Boburg, Rodrigo
Nemeth, Attila
Schille, Christine
Acharya, Metesh
Geis-Gerstorfer, Jürgen
Wendel, Hans-Peter
Popov, Aron-Frederik
Schlensak, Christian - Abstract:
- Aim: Patients with cardiogenic shock or ARDS, for example, in COVID-19/SARS-CoV-2, may require extracorporeal membrane oxygenation (ECMO). An ECLS/ECMO model simulating challenging vascular anatomy is desirable for cannula insertion training purposes. We assessed the ability of various 3D-printable materials to mimic the penetration properties of human tissue by using porcine aortae. Methods: A test bench for needle penetration and piercing in sampled porcine aorta and preselected 3D-printable polymers was assembled. The 3D-printable materials had Shore A hardness of 10, 20, and 50. 17G Vygon 1.0 × 1.4 mm × 70 mm needles were used for penetration tests. Results: For the porcine tissue and Shore A 10, Shore A 20, and Shore A 50 polymers, penetration forces of 0.9036 N, 0.9725 N, 1.0386 N, and 1.254 N were needed, respectively. For piercing through the porcine tissue and Shore A 10, Shore A 20, and Shore A 50 polymers, forces of 0.8399 N, 1.244 N, 1.475 N, and 1.482 N were needed, respectively. ANOVA showed different variances among the groups, and pairwise two-tailed t -tests showed significantly different needle penetration and piercing forces, except for penetration of Shore A 10 and 20 polymers (p = 0.234 and p = 0.0857). Significantly higher forces were required for all other materials. Conclusion: Shore A 10 and 20 polymers have similar needle penetration properties compared to the porcine tissue. Significantly more force is needed to pierce through the material fully.Aim: Patients with cardiogenic shock or ARDS, for example, in COVID-19/SARS-CoV-2, may require extracorporeal membrane oxygenation (ECMO). An ECLS/ECMO model simulating challenging vascular anatomy is desirable for cannula insertion training purposes. We assessed the ability of various 3D-printable materials to mimic the penetration properties of human tissue by using porcine aortae. Methods: A test bench for needle penetration and piercing in sampled porcine aorta and preselected 3D-printable polymers was assembled. The 3D-printable materials had Shore A hardness of 10, 20, and 50. 17G Vygon 1.0 × 1.4 mm × 70 mm needles were used for penetration tests. Results: For the porcine tissue and Shore A 10, Shore A 20, and Shore A 50 polymers, penetration forces of 0.9036 N, 0.9725 N, 1.0386 N, and 1.254 N were needed, respectively. For piercing through the porcine tissue and Shore A 10, Shore A 20, and Shore A 50 polymers, forces of 0.8399 N, 1.244 N, 1.475 N, and 1.482 N were needed, respectively. ANOVA showed different variances among the groups, and pairwise two-tailed t -tests showed significantly different needle penetration and piercing forces, except for penetration of Shore A 10 and 20 polymers (p = 0.234 and p = 0.0857). Significantly higher forces were required for all other materials. Conclusion: Shore A 10 and 20 polymers have similar needle penetration properties compared to the porcine tissue. Significantly more force is needed to pierce through the material fully. The most similar tested material to porcine aorta for needle penetration and piercing in ECMO-implantation is the silicon Shore A 10 polymer. This silicon could be a 3D-printable material in surgical training for ECMO-implantation. … (more)
- Is Part Of:
- Perfusion. Volume 36:Number 8(2021)
- Journal:
- Perfusion
- Issue:
- Volume 36:Number 8(2021)
- Issue Display:
- Volume 36, Issue 8 (2021)
- Year:
- 2021
- Volume:
- 36
- Issue:
- 8
- Issue Sort Value:
- 2021-0036-0008-0000
- Page Start:
- 798
- Page End:
- 802
- Publication Date:
- 2021-11
- Subjects:
- extracorporeal membrane oxygenation -- ECMO -- extracorporeal life support -- ECLS -- COVID-19 -- SARS-CoV-2 -- 3D printing -- medical education -- ZwickRoell -- needle penetration
Perfusion (Physiology) -- Periodicals
Blood -- Circulation, Artificial -- Periodicals
Heart -- Surgery -- Periodicals
Extracorporeal Circulation -- Periodicals
Perfusion -- Periodicals
Circulation extracorporelle -- Périodiques
Perfusion -- Périodiques
617.41 - Journal URLs:
- http://prf.sagepub.com/ ↗
http://www.uk.sagepub.com/home.nav ↗ - DOI:
- 10.1177/0267659120967194 ↗
- Languages:
- English
- ISSNs:
- 0267-6591
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 17632.xml