A Case of a C-Stem Fracture at the Head-Neck Junction and a Review of the Literature. (9th December 2012)
- Record Type:
- Journal Article
- Title:
- A Case of a C-Stem Fracture at the Head-Neck Junction and a Review of the Literature. (9th December 2012)
- Main Title:
- A Case of a C-Stem Fracture at the Head-Neck Junction and a Review of the Literature
- Authors:
- Morley, David
Starks, Ian
Lim, Justin - Other Names:
- Fisher D. A. Academic Editor.
Huber M. Academic Editor. - Abstract:
- Abstract : We report the first case of a fracture of the standard C-stem in combination with a large metal-on-metal articulation. This occurred at the head-neck junction. Analysis of the fractured stem showed evidence of fatigue failure with possible corrosion. The use of large femoral heads with neck adaptors and narrow tapers should be used with caution, especially in heavy, active patients.
- Is Part Of:
- Case reports in orthopedics. Volume 2012(2012)
- Journal:
- Case reports in orthopedics
- Issue:
- Volume 2012(2012)
- Issue Display:
- Volume 2012, Issue 2012 (2012)
- Year:
- 2012
- Volume:
- 2012
- Issue:
- 2012
- Issue Sort Value:
- 2012-2012-2012-0000
- Page Start:
- Page End:
- Publication Date:
- 2012-12-09
- Subjects:
- Orthopedics -- Periodicals
Orthopedics
Orthopedics
Periodicals
Electronic journals
Periodicals
616.7 - Journal URLs:
- https://www.hindawi.com/journals/crior/ ↗
http://www.ncbi.nlm.nih.gov/pmc/journals/1948/ ↗
http://bibpurl.oclc.org/web/48966 ↗
http://search.ebscohost.com/direct.asp?db=a9h&jid=%22EGTH%22&scope=site ↗ - DOI:
- 10.1155/2012/158604 ↗
- Languages:
- English
- ISSNs:
- 2090-6749
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library HMNTS - ELD Digital store
- Ingest File:
- 17512.xml