Demonstration of nonlinearity bias in the measurement of the apparent diffusion coefficient in multicenter trials. Issue 3 (2nd May 2015)
- Record Type:
- Journal Article
- Title:
- Demonstration of nonlinearity bias in the measurement of the apparent diffusion coefficient in multicenter trials. Issue 3 (2nd May 2015)
- Main Title:
- Demonstration of nonlinearity bias in the measurement of the apparent diffusion coefficient in multicenter trials
- Authors:
- Malyarenko, Dariya I.
Newitt, David
J. Wilmes, Lisa
Tudorica, Alina
Helmer, Karl G.
Arlinghaus, Lori R.
Jacobs, Michael A.
Jajamovich, Guido
Taouli, Bachir
Yankeelov, Thomas E.
Huang, Wei
Chenevert, Thomas L. - Abstract:
- Abstract : Purpose: Characterize system‐specific bias across common magnetic resonance imaging (MRI) platforms for quantitative diffusion measurements in multicenter trials. Methods: Diffusion weighted imaging (DWI) was performed on an ice‐water phantom along the superior–inferior (SI) and right–left (RL) orientations spanning ±150 mm. The same scanning protocol was implemented on 14 MRI systems at seven imaging centers. The bias was estimated as a deviation of measured from known apparent diffusion coefficient (ADC) along individual DWI directions. The relative contributions of gradient nonlinearity, shim errors, imaging gradients, and eddy currents were assessed independently. The observed bias errors were compared with numerical models. Results: The measured systematic ADC errors scaled quadratically with offset from isocenter, and ranged between −55% (SI) and 25% (RL). Nonlinearity bias was dependent on system design and diffusion gradient direction. Consistent with numerical models, minor ADC errors (±5%) due to shim, imaging and eddy currents were mitigated by double echo DWI and image coregistration of individual gradient directions. Conclusion: The analysis confirms gradient nonlinearity as a major source of spatial DW bias and variability in off‐center ADC measurements across MRI platforms, with minor contributions from shim, imaging gradients and eddy currents. The developed protocol enables empiric description of systematic bias in multicenter quantitative DWIAbstract : Purpose: Characterize system‐specific bias across common magnetic resonance imaging (MRI) platforms for quantitative diffusion measurements in multicenter trials. Methods: Diffusion weighted imaging (DWI) was performed on an ice‐water phantom along the superior–inferior (SI) and right–left (RL) orientations spanning ±150 mm. The same scanning protocol was implemented on 14 MRI systems at seven imaging centers. The bias was estimated as a deviation of measured from known apparent diffusion coefficient (ADC) along individual DWI directions. The relative contributions of gradient nonlinearity, shim errors, imaging gradients, and eddy currents were assessed independently. The observed bias errors were compared with numerical models. Results: The measured systematic ADC errors scaled quadratically with offset from isocenter, and ranged between −55% (SI) and 25% (RL). Nonlinearity bias was dependent on system design and diffusion gradient direction. Consistent with numerical models, minor ADC errors (±5%) due to shim, imaging and eddy currents were mitigated by double echo DWI and image coregistration of individual gradient directions. Conclusion: The analysis confirms gradient nonlinearity as a major source of spatial DW bias and variability in off‐center ADC measurements across MRI platforms, with minor contributions from shim, imaging gradients and eddy currents. The developed protocol enables empiric description of systematic bias in multicenter quantitative DWI studies. Magn Reson Med 75:1312–1323, 2016. © 2015 Wiley Periodicals, Inc. … (more)
- Is Part Of:
- Magnetic resonance in medicine. Volume 75:Issue 3(2016:Mar.)
- Journal:
- Magnetic resonance in medicine
- Issue:
- Volume 75:Issue 3(2016:Mar.)
- Issue Display:
- Volume 75, Issue 3 (2016)
- Year:
- 2016
- Volume:
- 75
- Issue:
- 3
- Issue Sort Value:
- 2016-0075-0003-0000
- Page Start:
- 1312
- Page End:
- 1323
- Publication Date:
- 2015-05-02
- Subjects:
- quantitative diffusion MRI -- multicenter trials -- gradient nonlinearity -- ADC mapping
Nuclear magnetic resonance -- Periodicals
Electron paramagnetic resonance -- Periodicals
616.07548 - Journal URLs:
- http://onlinelibrary.wiley.com/journal/10.1002/(ISSN)1522-2594 ↗
http://onlinelibrary.wiley.com/ ↗ - DOI:
- 10.1002/mrm.25754 ↗
- Languages:
- English
- ISSNs:
- 0740-3194
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 5337.798000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 17502.xml