A comparison study between the results of the Siemens ADVIA 120 analyzer and the manual method for differential leukocyte counts in sheep. (4th April 2021)
- Record Type:
- Journal Article
- Title:
- A comparison study between the results of the Siemens ADVIA 120 analyzer and the manual method for differential leukocyte counts in sheep. (4th April 2021)
- Main Title:
- A comparison study between the results of the Siemens ADVIA 120 analyzer and the manual method for differential leukocyte counts in sheep
- Authors:
- Oikonomidis, Ioannis L.
Brozos, Christos
Kiossis, Evangelos
Tsouloufi, Theodora K.
Kritsepi‐Konstantinou, Maria - Abstract:
- Abstract: Background: The ADVIA 120 is a widely used hematology analyzer, which has not been previously validated for determining differential leukocyte counts in sheep. Objectives: We aimed to compare differential leukocyte counts on the ADVIA 120 (A‐Diff) with counts obtained using the manual method (M‐Diff) in sheep. Methods: Ethylenediaminetetraacetic acid‐anticoagulated blood samples analyzed within 4 hours of collection were used. Samples with inappropriately filled tubes, detectable clots, overtly erroneous ADVIA peroxidase cytograms, and poor‐quality blood smears were excluded from the study. Two independent observers compared the results of the A‐Diff with those of the M‐Diff. The M‐diff was performed by counting 200 leukocytes on a blood smear. Results: Overall, 88 samples (44 rams and 44 ewes) were included. The correlation between the A‐Diff and M‐Diff was high for neutrophils ( r = .873, P < .001), lymphocytes ( r = .863, P < .001), and eosinophils ( r = .750, P < .001), and low for monocytes ( r = .212, P = .048). The Passing‐Bablok regression analyses revealed constant error for eosinophils [1.17%; 95% confidence interval (CI): 0.67%, 1.55%] and proportional error for lymphocytes (0.84; 95% CI: 0.74, 0.95) and eosinophils (0.85; 95% CI: 0.74, 0.96). The Bland‐Altman analyses revealed negative biases of 2.4% and 3.0% for neutrophils and lymphocytes, respectively, and positive biases of 3.2% and 0.8% for monocytes and eosinophils, respectively. TheAbstract: Background: The ADVIA 120 is a widely used hematology analyzer, which has not been previously validated for determining differential leukocyte counts in sheep. Objectives: We aimed to compare differential leukocyte counts on the ADVIA 120 (A‐Diff) with counts obtained using the manual method (M‐Diff) in sheep. Methods: Ethylenediaminetetraacetic acid‐anticoagulated blood samples analyzed within 4 hours of collection were used. Samples with inappropriately filled tubes, detectable clots, overtly erroneous ADVIA peroxidase cytograms, and poor‐quality blood smears were excluded from the study. Two independent observers compared the results of the A‐Diff with those of the M‐Diff. The M‐diff was performed by counting 200 leukocytes on a blood smear. Results: Overall, 88 samples (44 rams and 44 ewes) were included. The correlation between the A‐Diff and M‐Diff was high for neutrophils ( r = .873, P < .001), lymphocytes ( r = .863, P < .001), and eosinophils ( r = .750, P < .001), and low for monocytes ( r = .212, P = .048). The Passing‐Bablok regression analyses revealed constant error for eosinophils [1.17%; 95% confidence interval (CI): 0.67%, 1.55%] and proportional error for lymphocytes (0.84; 95% CI: 0.74, 0.95) and eosinophils (0.85; 95% CI: 0.74, 0.96). The Bland‐Altman analyses revealed negative biases of 2.4% and 3.0% for neutrophils and lymphocytes, respectively, and positive biases of 3.2% and 0.8% for monocytes and eosinophils, respectively. The extremely low basophil percentages precluded a meaningful method comparison for this leukocyte type. Conclusions: The ADVIA 120 appears to perform well for determining neutrophil, lymphocyte, and eosinophil percentages in sheep as compared with the manual method. However, blood smear examinations are recommended for the confirmation of abnormal differential leukocyte count results. … (more)
- Is Part Of:
- Veterinary clinical pathology. Volume 50:Number 2(2021)
- Journal:
- Veterinary clinical pathology
- Issue:
- Volume 50:Number 2(2021)
- Issue Display:
- Volume 50, Issue 2 (2021)
- Year:
- 2021
- Volume:
- 50
- Issue:
- 2
- Issue Sort Value:
- 2021-0050-0002-0000
- Page Start:
- 203
- Page End:
- 208
- Publication Date:
- 2021-04-04
- Subjects:
- blood smear -- complete blood count -- hematology -- leukogram -- ovine -- white blood cells
Veterinary pathology -- Periodicals
636.089607 - Journal URLs:
- http://onlinelibrary.wiley.com/ ↗
- DOI:
- 10.1111/vcp.12935 ↗
- Languages:
- English
- ISSNs:
- 0275-6382
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 9227.015500
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 17354.xml