Extracellular vesicle measurements with nanoparticle tracking analysis – An accuracy and repeatability comparison between NanoSight NS300 and ZetaView. Issue 1 (1st December 2019)
- Record Type:
- Journal Article
- Title:
- Extracellular vesicle measurements with nanoparticle tracking analysis – An accuracy and repeatability comparison between NanoSight NS300 and ZetaView. Issue 1 (1st December 2019)
- Main Title:
- Extracellular vesicle measurements with nanoparticle tracking analysis – An accuracy and repeatability comparison between NanoSight NS300 and ZetaView
- Authors:
- Bachurski, Daniel
Schuldner, Maximiliane
Nguyen, Phuong-Hien
Malz, Alexandra
Reiners, Katrin S
Grenzi, Patricia C
Babatz, Felix
Schauss, Astrid C
Hansen, Hinrich P
Hallek, Michael
Pogge von Strandmann, Elke - Abstract:
- ABSTRACT: The expanding field of extracellular vesicle (EV) research needs reproducible and accurate methods to characterize single EVs. Nanoparticle Tracking Analysis (NTA) is commonly used to determine EV concentration and diameter. As the EV field is lacking methods to easily confirm and validate NTA data, questioning the reliability of measurements remains highly important. In this regard, a comparison addressing measurement quality between different NTA devices such as Malvern's NanoSight NS300 or Particle Metrix' ZetaView has not yet been conducted. To evaluate the accuracy and repeatability of size and concentration determinations of both devices, we employed comparative methods including transmission electron microscopy (TEM) and single particle interferometric reflectance imaging sensing (SP-IRIS) by ExoView. Multiple test measurements with nanospheres, liposomes and ultracentrifuged EVs from human serum and cell culture supernatant were performed. Additionally, serial dilutions and freeze-thaw cycle-dependent EV decrease were measured to determine the robustness of each system. Strikingly, NanoSight NS300 exhibited a 2.0–2.1-fold overestimation of polystyrene and silica nanosphere concentration. By measuring serial dilutions of EV samples, we demonstrated higher accuracy in concentration determination by ZetaView (% BIAS range: 2.7–8.5) in comparison with NanoSight NS300 (% BIAS range: 32.9–36.8). The concentration measurements by ZetaView were also more precise (%ABSTRACT: The expanding field of extracellular vesicle (EV) research needs reproducible and accurate methods to characterize single EVs. Nanoparticle Tracking Analysis (NTA) is commonly used to determine EV concentration and diameter. As the EV field is lacking methods to easily confirm and validate NTA data, questioning the reliability of measurements remains highly important. In this regard, a comparison addressing measurement quality between different NTA devices such as Malvern's NanoSight NS300 or Particle Metrix' ZetaView has not yet been conducted. To evaluate the accuracy and repeatability of size and concentration determinations of both devices, we employed comparative methods including transmission electron microscopy (TEM) and single particle interferometric reflectance imaging sensing (SP-IRIS) by ExoView. Multiple test measurements with nanospheres, liposomes and ultracentrifuged EVs from human serum and cell culture supernatant were performed. Additionally, serial dilutions and freeze-thaw cycle-dependent EV decrease were measured to determine the robustness of each system. Strikingly, NanoSight NS300 exhibited a 2.0–2.1-fold overestimation of polystyrene and silica nanosphere concentration. By measuring serial dilutions of EV samples, we demonstrated higher accuracy in concentration determination by ZetaView (% BIAS range: 2.7–8.5) in comparison with NanoSight NS300 (% BIAS range: 32.9–36.8). The concentration measurements by ZetaView were also more precise (% CV range: 0.0–4.7) than measurements by NanoSight NS300 (% CV range: 5.4–10.7). On the contrary, quantitative TEM imaging indicated more accurate EV sizing by NanoSight NS300 (% DTEM range: 79.5–134.3) compared to ZetaView (% DTEM range: 111.8–205.7), while being equally repeatable (NanoSight NS300% CV range: 0.8–6.7; ZetaView: 1.4–7.8). However, both devices failed to report a peak EV diameter below 60 nm compared to TEM and SP-IRIS. Taken together, NTA devices differ strongly in their hardware and software affecting measuring results. ZetaView provided a more accurate and repeatable depiction of EV concentration, whereas NanoSight NS300 supplied size measurements of higher resolution. … (more)
- Is Part Of:
- Journal of extracellular vesicles. Volume 8:Issue 1(2019)
- Journal:
- Journal of extracellular vesicles
- Issue:
- Volume 8:Issue 1(2019)
- Issue Display:
- Volume 8, Issue 1 (2019)
- Year:
- 2019
- Volume:
- 8
- Issue:
- 1
- Issue Sort Value:
- 2019-0008-0001-0000
- Page Start:
- Page End:
- Publication Date:
- 2019-12-01
- Subjects:
- Extracellular vesicles -- exosomes -- nanoparticle tracking analysis -- transmission electron microscopy -- single particle interferometric reflectance imaging sensing -- accuracy -- repeatability -- reproducibility
Cells -- Mechanical properties -- Periodicals
Transport Vesicles
Cells -- Mechanical properties
Periodicals
Fulltext
Internet Resources
Periodicals
Electronic journals
Periodicals
571.63 - Journal URLs:
- http://www.ncbi.nlm.nih.gov/pmc/journals/2180/ ↗
https://www.tandfonline.com/toc/zjev20/current ↗
https://onlinelibrary.wiley.com/journal/20013078 ↗
http://www.tandfonline.com/ ↗ - DOI:
- 10.1080/20013078.2019.1596016 ↗
- Languages:
- English
- ISSNs:
- 2001-3078
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 17298.xml