An empirical artifact correction for proton computed tomography. (June 2021)
- Record Type:
- Journal Article
- Title:
- An empirical artifact correction for proton computed tomography. (June 2021)
- Main Title:
- An empirical artifact correction for proton computed tomography
- Authors:
- Dickmann, Jannis
Sarosiek, Christina
Götz, Stefanie
Pankuch, Mark
Coutrakon, George
Johnson, Robert P.
Schulte, Reinhard W.
Parodi, Katia
Landry, Guillaume
Dedes, George - Abstract:
- Highlights: Single particle tracking proton CT accuracy can be degraded by image artifacts. An empirical method was used to reduced artifacts in scans with a preclinical scanner. The mean average percentage error was reduced by 47% and was below 0.48%. Abstract: Purpose: To reduce image artifacts of proton computed tomography (pCT) from a preclinical scanner, for imaging of the relative stopping power (RSP) needed for particle therapy treatment planning using a simple empirical artifact correction method. Methods: We adapted and employed a correction method previously used for beam-hardening correction in x-ray CT which makes use of a single scan of a custom-built homogeneous phantom with known RSP. Exploiting the linearity of the filtered backprojection operation, a function was found which corrects water-equivalent path lengths (RSP line integrals) in experimental scans using a prototype pCT scanner. The correction function was applied to projection values of subsequent scans of a homogeneous water phantom, a sensitometric phantom with various inserts and an anthropomorphic head phantom. Data were acquired at two different incident proton energies to test the robustness of the method. Results: Inaccuracies in the detection process caused an offset and known ring artifacts in the water phantom which were considerably reduced using the proposed method. The mean absolute percentage error (MAPE) of mean RSP values of all inserts of the sensitometric phantom and the waterHighlights: Single particle tracking proton CT accuracy can be degraded by image artifacts. An empirical method was used to reduced artifacts in scans with a preclinical scanner. The mean average percentage error was reduced by 47% and was below 0.48%. Abstract: Purpose: To reduce image artifacts of proton computed tomography (pCT) from a preclinical scanner, for imaging of the relative stopping power (RSP) needed for particle therapy treatment planning using a simple empirical artifact correction method. Methods: We adapted and employed a correction method previously used for beam-hardening correction in x-ray CT which makes use of a single scan of a custom-built homogeneous phantom with known RSP. Exploiting the linearity of the filtered backprojection operation, a function was found which corrects water-equivalent path lengths (RSP line integrals) in experimental scans using a prototype pCT scanner. The correction function was applied to projection values of subsequent scans of a homogeneous water phantom, a sensitometric phantom with various inserts and an anthropomorphic head phantom. Data were acquired at two different incident proton energies to test the robustness of the method. Results: Inaccuracies in the detection process caused an offset and known ring artifacts in the water phantom which were considerably reduced using the proposed method. The mean absolute percentage error (MAPE) of mean RSP values of all inserts of the sensitometric phantom and the water phantom was reduced from 0.87% to 0.44% and from 0.86% to 0.48% for the two incident energies respectively. In the head phantom a clear reduction of artifacts was observed. Conclusions: Image artifacts of experimental pCT scans with a prototype scanner could substantially be reduced both in homogeneous, heterogeneous and anthropomorphic phantoms. RSP accuracy was also improved. … (more)
- Is Part Of:
- Physica medica. Volume 86(2021)
- Journal:
- Physica medica
- Issue:
- Volume 86(2021)
- Issue Display:
- Volume 86, Issue 2021 (2021)
- Year:
- 2021
- Volume:
- 86
- Issue:
- 2021
- Issue Sort Value:
- 2021-0086-2021-0000
- Page Start:
- 57
- Page End:
- 65
- Publication Date:
- 2021-06
- Subjects:
- Proton computed tomography -- Particle therapy -- Artifact reduction -- Image quality
Medical physics -- Periodicals
Biophysics -- Periodicals
Biophysics -- Periodicals
Imagerie médicale -- Périodiques
Radiothérapie -- Périodiques
Rayons X -- Sécurité -- Mesures -- Périodiques
Physique -- Périodiques
Médecine -- Périodiques
610.153 - Journal URLs:
- http://www.sciencedirect.com/science/journal/11201797 ↗
http://www.clinicalkey.com/dura/browse/journalIssue/11201797 ↗
http://www.clinicalkey.com.au/dura/browse/journalIssue/11201797 ↗
http://www.elsevier.com/journals ↗
http://www.physicamedica.com ↗ - DOI:
- 10.1016/j.ejmp.2021.05.018 ↗
- Languages:
- English
- ISSNs:
- 1120-1797
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 6475.070000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 17259.xml