- Verifying the validity of creatinine measurement in low‐concentrated urine spot samples—Photospectrometry versus liquid chromatography–tandem mass spectrometry. Issue 6 (28th January 2021)
- Main Title:
- Verifying the validity of creatinine measurement in low‐concentrated urine spot samples—Photospectrometry versus liquid chromatography–tandem mass spectrometry
- Authors:
- Franz, Simon
Skopp, Gisela
Dame, Torsten
Musshoff, Frank - Abstract:
- Abstract: One of the major challenges of testing drugs of abuse is the detection of highly diluted urine samples. The ingestion of a large amount of fluid can considerably reduce the concentration of substances, possibly resulting in inaccurate drug testing. For detection, determination of urinary creatinine is a widely established procedure. In this study, results from the most popular methods, including photospectrometry (Jaffe) and liquid chromatography–tandem mass spectrometry (LC–MS/MS), have been compared regarding 327 urine abstinence control samples. Because samples with creatinine concentrations close to the cutoff of 20 mg/dL are of particular interest, only samples below 50 mg/dL were considered. Results revealed a close correlation of creatinine concentrations by both analytical methods with an R 2 value of 0.9005. A mean concentration difference of 3.30 ± 3.45 mg/dL was observed, indicating a moderate underestimation by the Jaffe reaction. Graphical analyses showed high accordance between both methods with only a few outliers. Due to easy handling and for economic reasons, the spectrometric method is often preferred over LC–MS/MS. For urine samples with creatinine concentrations close to the cutoff, confirmation through a second method should be performed to avoid a possible disadvantage or even severe consequences for the respective individual. It is recommended that each laboratory establishes a reliable verification method. Abstract : PhotospectrometryAbstract: One of the major challenges of testing drugs of abuse is the detection of highly diluted urine samples. The ingestion of a large amount of fluid can considerably reduce the concentration of substances, possibly resulting in inaccurate drug testing. For detection, determination of urinary creatinine is a widely established procedure. In this study, results from the most popular methods, including photospectrometry (Jaffe) and liquid chromatography–tandem mass spectrometry (LC–MS/MS), have been compared regarding 327 urine abstinence control samples. Because samples with creatinine concentrations close to the cutoff of 20 mg/dL are of particular interest, only samples below 50 mg/dL were considered. Results revealed a close correlation of creatinine concentrations by both analytical methods with an R 2 value of 0.9005. A mean concentration difference of 3.30 ± 3.45 mg/dL was observed, indicating a moderate underestimation by the Jaffe reaction. Graphical analyses showed high accordance between both methods with only a few outliers. Due to easy handling and for economic reasons, the spectrometric method is often preferred over LC–MS/MS. For urine samples with creatinine concentrations close to the cutoff, confirmation through a second method should be performed to avoid a possible disadvantage or even severe consequences for the respective individual. It is recommended that each laboratory establishes a reliable verification method. Abstract : Photospectrometry (Jaffe) and LC–MS/MS for determination of creatinine in low‐concentrated urine samples is compared. Confirmation through a second method for samples with creatinine concentrations below the cutoff is recommended to avoid a possible disadvantage for participants of drug‐testing programs. … (more)
- Is Part Of:
- Drug testing and analysis. Volume 13:Issue 6(2021)
- Journal:
- Drug testing and analysis
- Issue:
- Volume 13:Issue 6(2021)
- Issue Display:
- Volume 13, Issue 6 (2021)
- Year:
- 2021
- Volume:
- 13
- Issue:
- 6
- Issue Sort Value:
- 2021-0013-0006-0000
- Page Start:
- 1136
- Page End:
- 1144
- Publication Date:
- 2021-01-28
- Subjects:
- creatinine -- liquid chromatography–tandem mass spectrometry -- method comparison -- photospectrometry -- urine
Drugs -- Analysis -- Periodicals
Drug testing -- Periodicals
Chemistry, Forensic -- Periodicals
615.1901 - Journal URLs:
- http://onlinelibrary.wiley.com/journal/10.1002/(ISSN)1942-7611 ↗
http://rzblx1.uni-regensburg.de/ezeit/warpto.phtml?colors=7&jour_id=110501 ↗
http://www3.interscience.wiley.com/journal/121408477/home ↗
http://onlinelibrary.wiley.com/ ↗ - DOI:
- 10.1002/dta.3003 ↗
- Languages:
- English
- ISSNs:
- 1942-7603
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 3629.424000
British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 17564.xml