Development of a movable 4πβ(LS)−γ coincidence counting system for activity standardization of β−γ emitters. (August 2021)
- Record Type:
- Journal Article
- Title:
- Development of a movable 4πβ(LS)−γ coincidence counting system for activity standardization of β−γ emitters. (August 2021)
- Main Title:
- Development of a movable 4πβ(LS)−γ coincidence counting system for activity standardization of β−γ emitters
- Authors:
- Lee, J.M.
Agung, A.
Hwang, S.H.
Lee, K.B.
Hwang, H.Y. - Abstract:
- Abstract: A new movable 3PM-γ coincidence system, based on 4πβ(LS)−γ coincidence counting, for activity measurement of β-γ emitters has been designed at the Korea Research Institute of Standards and Science (KRISS). The system incorporates 3 PM tubes on the plane and two detectors placed above and below the center of the plane. The 3 PM tubes for β–counters in the plane are movable up to 100 mm from a liquid scintillation vial, thus enabling the variation of β−detection efficiencies by a geometrical technique. A NaI(Tl) γ-counter was set above for the present work. The β-event is determined by counting the logical sum of three double coincidences. All the necessary electronics, i.e., logical sum, adjusting the duration of dead-time of each counting channel and coincidence resolving times, and analyzing coincidence relation, were specially designed to be fabricated in an integrated circuit. Details of the detectors, the electronics, the overall movable 3PM-γ coincidence system are presented, as well as the results of investigations to assess its operating characteristics. Validation measurements have been performed with 60 Co and 57 Co sources. The highest β−detection efficiency achieved with 60 Co and 57 Co was 97% and 95%, respectively. The activity concentration determined with a new system agreed with calibrated values within the uncertainty range. Further results from validation measurements and the corresponding uncertainty budgets are presented. Highlights: Design ofAbstract: A new movable 3PM-γ coincidence system, based on 4πβ(LS)−γ coincidence counting, for activity measurement of β-γ emitters has been designed at the Korea Research Institute of Standards and Science (KRISS). The system incorporates 3 PM tubes on the plane and two detectors placed above and below the center of the plane. The 3 PM tubes for β–counters in the plane are movable up to 100 mm from a liquid scintillation vial, thus enabling the variation of β−detection efficiencies by a geometrical technique. A NaI(Tl) γ-counter was set above for the present work. The β-event is determined by counting the logical sum of three double coincidences. All the necessary electronics, i.e., logical sum, adjusting the duration of dead-time of each counting channel and coincidence resolving times, and analyzing coincidence relation, were specially designed to be fabricated in an integrated circuit. Details of the detectors, the electronics, the overall movable 3PM-γ coincidence system are presented, as well as the results of investigations to assess its operating characteristics. Validation measurements have been performed with 60 Co and 57 Co sources. The highest β−detection efficiency achieved with 60 Co and 57 Co was 97% and 95%, respectively. The activity concentration determined with a new system agreed with calibrated values within the uncertainty range. Further results from validation measurements and the corresponding uncertainty budgets are presented. Highlights: Design of movable 3PM-γ coincidence counter based on 4πβ (LS) −γ (NaI(Tl)) system. Logical sum of three double coincidences for β event. Validation of geometrical efficiency variation & linear efficiency-extrapolation. Apply on activity measurement of 60 Co and 57 Co. … (more)
- Is Part Of:
- Applied radiation and isotopes. Volume 174(2021)
- Journal:
- Applied radiation and isotopes
- Issue:
- Volume 174(2021)
- Issue Display:
- Volume 174, Issue 2021 (2021)
- Year:
- 2021
- Volume:
- 174
- Issue:
- 2021
- Issue Sort Value:
- 2021-0174-2021-0000
- Page Start:
- Page End:
- Publication Date:
- 2021-08
- Subjects:
- A movable 3PM-γ coincidence counter -- 4πβ (LS)−γ(NaI(Tl)) system -- Geometrical efficiency variation -- Logical sum of double coincidences -- Standardization of 60Co and 57Co
Radiology -- Periodicals
Radiation -- Industrial applications -- Periodicals
Nuclear chemistry -- Periodicals
Internet resource
Periodical
660.298 - Journal URLs:
- http://www.sciencedirect.com/science/journal/09698043 ↗
http://catalog.hathitrust.org/api/volumes/oclc/27456684.html ↗
http://www.elsevier.com/journals ↗ - DOI:
- 10.1016/j.apradiso.2021.109743 ↗
- Languages:
- English
- ISSNs:
- 0969-8043
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 1576.565000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 17220.xml