Accuracy of five systems for self-monitoring of blood glucose in the hands of adult lay-users and professionals applying ISO 15197:2013 accuracy criteria and potential insulin dosing errors. (1st February 2019)
- Record Type:
- Journal Article
- Title:
- Accuracy of five systems for self-monitoring of blood glucose in the hands of adult lay-users and professionals applying ISO 15197:2013 accuracy criteria and potential insulin dosing errors. (1st February 2019)
- Main Title:
- Accuracy of five systems for self-monitoring of blood glucose in the hands of adult lay-users and professionals applying ISO 15197:2013 accuracy criteria and potential insulin dosing errors
- Authors:
- Jendrike, Nina
Baumstark, Annette
Pleus, Stefan
Liebing, Christina
Kamecke, Ulrike
Haug, Cornelia
Freckmann, Guido - Abstract:
- Abstract: Objective: In this study, accuracy in the hands of intended users was evaluated for five self-monitoring of blood glucose (SMBG) systems based on ISO 15197:2013, and possibly related insulin dosing errors were calculated. In addition, accuracy was assessed in the hands of study personnel. Methods: For each system (Accu-Chek 1 Aviva Connect [A], Contour 2 Next One [B], FreeStyle Freedom Lite 3 [C], GlucoMen 4 areo [D] and OneTouch Verio 5 [E]) one test strip lot was evaluated as required by ISO 15197:2013, clause 8. Number and percentage of SMBG measurements within ±15 mg/dl and ±15% of the comparison measurements at glucose concentrations <100 mg/dl and ≥100 mg/dl, respectively, were calculated. In addition, data is presented in surveillance error grids, and insulin dosing errors were modeled. The study was registered at ClinicalTrials.gov (NCT03033849). Results: Four systems (A, B, C, D) fulfilled the tested reagent system lot ISO 15197:2013 accuracy criteria with the tested reagent system lot with at least 95% (lay-users) and 99.5% (study personnel) of results within the defined limits. Measurements with all five systems were within the clinically acceptable zones of the consensus error grid and the surveillance error grid. Median modeled insulin dosing errors were between -0.8 and +0.6 units for measurements performed by lay-users and between -0.7 and +0.8 units for study personnel. Frequent lay-user errors were not checking the test strips' expiry date,Abstract: Objective: In this study, accuracy in the hands of intended users was evaluated for five self-monitoring of blood glucose (SMBG) systems based on ISO 15197:2013, and possibly related insulin dosing errors were calculated. In addition, accuracy was assessed in the hands of study personnel. Methods: For each system (Accu-Chek 1 Aviva Connect [A], Contour 2 Next One [B], FreeStyle Freedom Lite 3 [C], GlucoMen 4 areo [D] and OneTouch Verio 5 [E]) one test strip lot was evaluated as required by ISO 15197:2013, clause 8. Number and percentage of SMBG measurements within ±15 mg/dl and ±15% of the comparison measurements at glucose concentrations <100 mg/dl and ≥100 mg/dl, respectively, were calculated. In addition, data is presented in surveillance error grids, and insulin dosing errors were modeled. The study was registered at ClinicalTrials.gov (NCT03033849). Results: Four systems (A, B, C, D) fulfilled the tested reagent system lot ISO 15197:2013 accuracy criteria with the tested reagent system lot with at least 95% (lay-users) and 99.5% (study personnel) of results within the defined limits. Measurements with all five systems were within the clinically acceptable zones of the consensus error grid and the surveillance error grid. Median modeled insulin dosing errors were between -0.8 and +0.6 units for measurements performed by lay-users and between -0.7 and +0.8 units for study personnel. Frequent lay-user errors were not checking the test strips' expiry date, applying blood incorrectly and handling the device incorrectly. Conclusion: In this study, the systems showed slight differences in the number of results within ISO 15197:2013 accuracy limits. Inaccurate SMBG measurements can result in insulin dosing errors and adversely affect glycemic control. Trial registration: ClinicalTrials.gov identifier: NCT03033849. … (more)
- Is Part Of:
- Current medical research and opinion. Volume 35:Number 2(2019)
- Journal:
- Current medical research and opinion
- Issue:
- Volume 35:Number 2(2019)
- Issue Display:
- Volume 35, Issue 2 (2019)
- Year:
- 2019
- Volume:
- 35
- Issue:
- 2
- Issue Sort Value:
- 2019-0035-0002-0000
- Page Start:
- 301
- Page End:
- 311
- Publication Date:
- 2019-02-01
- Subjects:
- Self-monitoring of blood glucose -- system accuracy -- ISO 15197 -- user performance -- lay-user -- insulin dosing error -- surveillance error grid
Clinical medicine -- Periodicals
Therapeutics -- Periodicals
615.5 - Journal URLs:
- http://informahealthcare.com ↗
- DOI:
- 10.1080/03007995.2018.1491832 ↗
- Languages:
- English
- ISSNs:
- 0300-7995
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 3500.301000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 17195.xml