Energy calibration and performance testing of LEGe detector. Issue 23 (24th October 2019)
- Record Type:
- Journal Article
- Title:
- Energy calibration and performance testing of LEGe detector. Issue 23 (24th October 2019)
- Main Title:
- Energy calibration and performance testing of LEGe detector
- Authors:
- Hou, Dongjie
Wu, Chong
Li, Chengze
Guo, Siming
Wu, Jinjie - Abstract:
- Abstract : The low energy germanium detector (LEGe) is in all aspects optimised for performance at low and moderate energies. It has specific advantages over conventional planar or coaxial detectors. LEGe is a kind of semiconductor detector commonly used in X‐ray detection. It can display the spectrum of X‐ray and record photon counts. Photon fluence is used to describe the radiation field by recording the number of incident photons. In order to take full advantage of the low energy response of this intrinsically thin window detector, LEGe cryostats are usually equipped with a beryllium window, thin front and side contact allowing spectroscopy from 3 keV and above. The rear contact is of less than full area which gives a lower detector capacitance compared to a planar device of similar size. It is widely used in low energy gamma spectroscopy, X‐ray absorption spectroscopy, nuclear safeguards or X‐ray fluorescence spectrometer and X‐ray diffraction. Spectrum is an intuitional and accurate description of X‐ray radiation. In this work, experiment was carried out to calibrate the detector and test the spectrum, and then to calculate the energy resolution. The experiment results about energy resolution (full width at half maximum) are 222 eV (@5.9 keV) and 519 eV (@122 keV).
- Is Part Of:
- Journal of engineering. Volume 2019:Issue 23(2019)
- Journal:
- Journal of engineering
- Issue:
- Volume 2019:Issue 23(2019)
- Issue Display:
- Volume 2019, Issue 23 (2019)
- Year:
- 2019
- Volume:
- 2019
- Issue:
- 23
- Issue Sort Value:
- 2019-2019-0023-0000
- Page Start:
- 9031
- Page End:
- 9034
- Publication Date:
- 2019-10-24
- Subjects:
- photon counting -- X‐ray detection -- X‐ray diffraction -- germanium radiation detectors -- X‐ray fluorescence analysis -- fluorescence -- calibration -- gamma‐ray detection -- gamma‐ray spectroscopy
electron volt energy 5.9 keV -- electron volt energy 122.0 keV -- electron volt energy 519.0 eV -- electron volt energy 222.0 eV -- electron volt energy 3.0 keV -- photon counts -- energy resolution -- X‐ray radiation -- X‐ray diffraction -- X‐ray fluorescence spectrometer -- X‐ray absorption spectroscopy -- low energy gamma spectroscopy -- planar device -- lower detector capacitance -- rear contact -- beryllium window -- LEGe cryostats -- intrinsically thin window detector -- low energy response -- incident photons -- radiation field -- photon fluence -- X‐ray detection -- semiconductor detector -- low energy germanium detector -- LEGe detector -- energy calibration
Engineering -- Periodicals
Engineering
Electronic journals
Periodicals
620.005 - Journal URLs:
- http://digital-library.theiet.org/content/journals/joe ↗
https://ietresearch.onlinelibrary.wiley.com/journal/20513305 ↗
http://biburl.oclc.org/web/74111 ↗
http://ieeexplore.ieee.org/Xplore/home.jsp ↗ - DOI:
- 10.1049/joe.2018.9175 ↗
- Languages:
- English
- ISSNs:
- 2051-3305
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 4978.368000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 17047.xml