A Mechanical Comparison of the Compressive Force Generated by Various Headless Compression Screws and the Impact of Fracture Gap Size. (September 2021)
- Record Type:
- Journal Article
- Title:
- A Mechanical Comparison of the Compressive Force Generated by Various Headless Compression Screws and the Impact of Fracture Gap Size. (September 2021)
- Main Title:
- A Mechanical Comparison of the Compressive Force Generated by Various Headless Compression Screws and the Impact of Fracture Gap Size
- Authors:
- Ilyas, Asif M.
Mahoney, Jonathan M.
Bucklen, Brandon S. - Abstract:
- Background: There is evidence that interfragmentary fracture gap size may affect the compression achievable with a modern headless compression screw (HCS). This mechanical study compared the compression achieved by 3 commercial HCS systems through various fracture gaps: CAPTIVATE Headless (Globus Medical, Inc, Audubon, Pennsylvania), Synthes (DePuy Synthes, Westchester, Pennsylvania), and Acumed Acutrak 2 (Acumed LLC, Hillsboro, Oregon).Methods: Screws were inserted into a custom test fixture composed of polyurethane synthetic bone foam fragments, separated by a layer of easily compressible polyurethane foam simulating a fracture gap. Compression was measured after final insertion and countersinking. The effect of the interfragmentary fracture gap size on the compression generated was also investigated.Results: The CAPTIVATE Headless 3.0 mm screw (70.1 ± 5.7 N) and the Synthes 3.0 mm screw (64.9 ± 7.3 N) achieved similar compressive forces after final countersink. Similar comparisons were found for the CAPTIVATE Headless 2.5 mm and Synthes 2.4 mm screws, and the CAPTIVATE Headless 4.0 mm and Acutrak 2 Standard screws. The final compression of the CAPTIVATE Headless 2.5 mm and Synthes 2.4 mm screws was not significantly affected when the fracture gap was doubled from 2 to 4 mm, but was reduced significantly by 95.9% with the Acutrak 2 Micro screw.Conclusion: When comparing like-sized screws, the CAPTIVATE, Synthes, and Acutrak 2 HCS systems demonstrated similar potentialBackground: There is evidence that interfragmentary fracture gap size may affect the compression achievable with a modern headless compression screw (HCS). This mechanical study compared the compression achieved by 3 commercial HCS systems through various fracture gaps: CAPTIVATE Headless (Globus Medical, Inc, Audubon, Pennsylvania), Synthes (DePuy Synthes, Westchester, Pennsylvania), and Acumed Acutrak 2 (Acumed LLC, Hillsboro, Oregon).Methods: Screws were inserted into a custom test fixture composed of polyurethane synthetic bone foam fragments, separated by a layer of easily compressible polyurethane foam simulating a fracture gap. Compression was measured after final insertion and countersinking. The effect of the interfragmentary fracture gap size on the compression generated was also investigated.Results: The CAPTIVATE Headless 3.0 mm screw (70.1 ± 5.7 N) and the Synthes 3.0 mm screw (64.9 ± 7.3 N) achieved similar compressive forces after final countersink. Similar comparisons were found for the CAPTIVATE Headless 2.5 mm and Synthes 2.4 mm screws, and the CAPTIVATE Headless 4.0 mm and Acutrak 2 Standard screws. The final compression of the CAPTIVATE Headless 2.5 mm and Synthes 2.4 mm screws was not significantly affected when the fracture gap was doubled from 2 to 4 mm, but was reduced significantly by 95.9% with the Acutrak 2 Micro screw.Conclusion: When comparing like-sized screws, the CAPTIVATE, Synthes, and Acutrak 2 HCS systems demonstrated similar potential compressive forces. However, compared with the CAPTIVATE Headless and Synthes HCS systems, which are inserted with a compression sleeve that is not gap distance–dependent, the Acutrak 2 HCS system demonstrated less compression when the simulated fracture gap size was increased to 4 mm. … (more)
- Is Part Of:
- Hand. Volume 16:Number 5(2021)
- Journal:
- Hand
- Issue:
- Volume 16:Number 5(2021)
- Issue Display:
- Volume 16, Issue 5 (2021)
- Year:
- 2021
- Volume:
- 16
- Issue:
- 5
- Issue Sort Value:
- 2021-0016-0005-0000
- Page Start:
- 604
- Page End:
- 611
- Publication Date:
- 2021-09
- Subjects:
- headless compression screw -- scaphoid fracture -- internal fixation -- fracture gap -- interfragmentary compression -- bone fragments -- compression sleeve
Hand -- Surgery -- Periodicals
Hand -- Surgery
Periodicals
617.57005 - Journal URLs:
- http://www.springerlink.com/content/119980/ ↗
http://journals.sagepub.com/toc/HAN/current ↗
http://www.springer.com/gb/ ↗ - DOI:
- 10.1177/1558944719877890 ↗
- Languages:
- English
- ISSNs:
- 1558-9447
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 4241.550050
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 17054.xml