Defibrillation success after subcutaneous implantable cardioverter-defibrillator generator replacement is not affected by impedance increase. (24th May 2021)
- Record Type:
- Journal Article
- Title:
- Defibrillation success after subcutaneous implantable cardioverter-defibrillator generator replacement is not affected by impedance increase. (24th May 2021)
- Main Title:
- Defibrillation success after subcutaneous implantable cardioverter-defibrillator generator replacement is not affected by impedance increase
- Authors:
- Van Der Stuijt, W
Pepplinkhuizen, S
Quast, ABE
Smeding, L
Olde Nordkamp, LRA
Wilde, AAM
Knops, RE - Abstract:
- Abstract: Funding Acknowledgements: Type of funding sources: None. Background: Routine defibrillation testing during implant and replacement of the subcutaneous implantable cardioverter-defibrillator (S-ICD) is recommended per current guidelines. Recently, concerns have been raised about an increase in shock impedance and consequent shock failure during defibrillation testing in S-ICD patients undergoing a generator replacement. Purpose: We aim to describe the defibrillation success rate in relation to the shock impedance in patients undergoing S-ICD generator replacement in our large tertiary center. Methods: In this retrospective analysis, data from replacement procedures were collected from all patients who underwent an S-ICD generator replacement in our center from June 2014 to December 2020. Defibrillation testing was performed with at least one shock of ≤65J, and a successful shock was defined as terminating the ventricular arrhythmia within 5 seconds after the shock. Results: A total of 133 patients underwent an S-ICD generator replacement, 5.8 ± 0.9 years after initial implant. Reasons for replacement were: reaching of elective replacement indicator (n = 119), early battery depletion (n = 9), complaints of generator pocket (n = 3) and device malfunction (n = 2). Defibrillation testing was performed in 111 patients (86.5%) undergoing a replacement procedure. Shock impedance data from both the implant and replacement procedure were available in 101 patients. The medianAbstract: Funding Acknowledgements: Type of funding sources: None. Background: Routine defibrillation testing during implant and replacement of the subcutaneous implantable cardioverter-defibrillator (S-ICD) is recommended per current guidelines. Recently, concerns have been raised about an increase in shock impedance and consequent shock failure during defibrillation testing in S-ICD patients undergoing a generator replacement. Purpose: We aim to describe the defibrillation success rate in relation to the shock impedance in patients undergoing S-ICD generator replacement in our large tertiary center. Methods: In this retrospective analysis, data from replacement procedures were collected from all patients who underwent an S-ICD generator replacement in our center from June 2014 to December 2020. Defibrillation testing was performed with at least one shock of ≤65J, and a successful shock was defined as terminating the ventricular arrhythmia within 5 seconds after the shock. Results: A total of 133 patients underwent an S-ICD generator replacement, 5.8 ± 0.9 years after initial implant. Reasons for replacement were: reaching of elective replacement indicator (n = 119), early battery depletion (n = 9), complaints of generator pocket (n = 3) and device malfunction (n = 2). Defibrillation testing was performed in 111 patients (86.5%) undergoing a replacement procedure. Shock impedance data from both the implant and replacement procedure were available in 101 patients. The median shock impedance of these patients during their replacement procedure was significantly higher than during their implant, 79Ω (IQR 66-94) and 66Ω (IQR 57.5-81) respectively (Z = -5.552, p < 0.001). Despite the higher shock impedance, first shock during defibrillation testing was successful in 105/111 patients (94.6%), with a success rate of 97.3% after two attempts. In the remaining three patients, the ventricular arrhythmia could only be terminated with a 80J shock. This was the case during both their initial implant and their replacement procedure. Shock impedance increase between implant and replacement was not significantly higher in patients with a successful first shock compared to patients with an unsuccessful first shock (Δ+11.1 ± 20.0Ω versus Δ+12.7 ± 27.6Ω, p = 0.86). Conclusion: In this large retrospective analysis, we have shown a first shock success rate during S-ICD generator replacement of 94.6%, which is similar to the success rate of defibrillation testing after initial implant. After multiple attempts, defibrillation testing success rate was 100%. Even though the median shock impedance during replacement was significantly higher than during the initial implant, there was no difference in impedance increase in patients with a successful shock compared to patients with an unsuccessful shock. … (more)
- Is Part Of:
- Europace. Volume 23:Supplement 3(2021)
- Journal:
- Europace
- Issue:
- Volume 23:Supplement 3(2021)
- Issue Display:
- Volume 23, Issue 3 (2021)
- Year:
- 2021
- Volume:
- 23
- Issue:
- 3
- Issue Sort Value:
- 2021-0023-0003-0000
- Page Start:
- Page End:
- Publication Date:
- 2021-05-24
- Subjects:
- Arrhythmia -- Treatment -- Periodicals
Cardiac pacing -- Periodicals
Catheter ablation -- Periodicals
Heart -- Physiology -- Periodicals
Electrophysiology -- Periodicals
617.4120645 - Journal URLs:
- http://europace.oxfordjournals.org/ ↗
http://ukcatalogue.oup.com/ ↗ - DOI:
- 10.1093/europace/euab116.416 ↗
- Languages:
- English
- ISSNs:
- 1099-5129
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 3829.340450
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 17093.xml