Management of Nanostim leadless pacemaker early-life battery failures. A single center experience. (24th May 2021)
- Record Type:
- Journal Article
- Title:
- Management of Nanostim leadless pacemaker early-life battery failures. A single center experience. (24th May 2021)
- Main Title:
- Management of Nanostim leadless pacemaker early-life battery failures. A single center experience
- Authors:
- Oosterwerff, F
Salavati, A
Adiyaman, A
Ghani, A
Smit, JJJ
Ramdat Misier, A
Delnoy, PPHM
Elvan, A - Abstract:
- Abstract: Funding Acknowledgements: Type of funding sources: None. Background/introduction Early studies of the Nanostim leadless pacemaker demonstrated technical feasibility effective pacing performance at the short-term. However, recently the Nanostim proved to be susceptible to early-life battery failures. Purpose In this study, we addressed the following research questions: 1) What percentage of Nanostim LPs experience early-life battery failure? 2) Is Nanostim extraction and subsequent pacemaker reimplantation safe and effective? 3) Is three monthly follow-up under the Nanostim advisory effective at preventing Nanostim LP related in-between hospitalizations? Methods In a retrospective study, we collected data of 49 Nanostim implanted patients with a mean follow up of 3 (±1.5) years at our tertiary hospital. Results Nanostim early-life battery failure in our population was 37% (18/49). Extraction of 14 Nanostim pacemakers (device age 1040 ± 467 days) in an older population (80 ± 7 years) was safe (0 complications) and effective (80% extraction success, 100% reimplantation success). All known cases of early-life battery failure were identified during the three monthly follow-up consultations. Conclusions Nanostim LP early-life battery failure is substantially higher than previously reported. In case of dysfunction Nanostim extraction in an older population is safe and effective. Three monthly follow-up is effective at preventing in-between Nanostim relatedAbstract: Funding Acknowledgements: Type of funding sources: None. Background/introduction Early studies of the Nanostim leadless pacemaker demonstrated technical feasibility effective pacing performance at the short-term. However, recently the Nanostim proved to be susceptible to early-life battery failures. Purpose In this study, we addressed the following research questions: 1) What percentage of Nanostim LPs experience early-life battery failure? 2) Is Nanostim extraction and subsequent pacemaker reimplantation safe and effective? 3) Is three monthly follow-up under the Nanostim advisory effective at preventing Nanostim LP related in-between hospitalizations? Methods In a retrospective study, we collected data of 49 Nanostim implanted patients with a mean follow up of 3 (±1.5) years at our tertiary hospital. Results Nanostim early-life battery failure in our population was 37% (18/49). Extraction of 14 Nanostim pacemakers (device age 1040 ± 467 days) in an older population (80 ± 7 years) was safe (0 complications) and effective (80% extraction success, 100% reimplantation success). All known cases of early-life battery failure were identified during the three monthly follow-up consultations. Conclusions Nanostim LP early-life battery failure is substantially higher than previously reported. In case of dysfunction Nanostim extraction in an older population is safe and effective. Three monthly follow-up is effective at preventing in-between Nanostim related hospitalization. … (more)
- Is Part Of:
- Europace. Volume 23:Supplement 3(2021)
- Journal:
- Europace
- Issue:
- Volume 23:Supplement 3(2021)
- Issue Display:
- Volume 23, Issue 3 (2021)
- Year:
- 2021
- Volume:
- 23
- Issue:
- 3
- Issue Sort Value:
- 2021-0023-0003-0000
- Page Start:
- Page End:
- Publication Date:
- 2021-05-24
- Subjects:
- Arrhythmia -- Treatment -- Periodicals
Cardiac pacing -- Periodicals
Catheter ablation -- Periodicals
Heart -- Physiology -- Periodicals
Electrophysiology -- Periodicals
617.4120645 - Journal URLs:
- http://europace.oxfordjournals.org/ ↗
http://ukcatalogue.oup.com/ ↗ - DOI:
- 10.1093/europace/euab116.378 ↗
- Languages:
- English
- ISSNs:
- 1099-5129
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 3829.340450
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File: