Validation of handheld fundus camera with mydriasis for retinal imaging of diabetic retinopathy screening in China: a prospective comparison study. Issue 10 (29th October 2020)
- Record Type:
- Journal Article
- Title:
- Validation of handheld fundus camera with mydriasis for retinal imaging of diabetic retinopathy screening in China: a prospective comparison study. Issue 10 (29th October 2020)
- Main Title:
- Validation of handheld fundus camera with mydriasis for retinal imaging of diabetic retinopathy screening in China: a prospective comparison study
- Authors:
- Xiao, Baixiang
Liao, Qinghua
Li, Yanping
Weng, Fan
Jin, Ling
Wang, Yanfang
Huang, Wenyong
Yi, Jinglin
Burton, Matthew J
Yip, Jennifer LY - Abstract:
- Abstract : Objectives: To investigate the clinical validity of using a handheld fundus camera to detect diabetic retinopathy (DR) in China. Design and settings: Prospective comparison study of the handheld fundus camera with a standard validated instrument in detection of DR in hospital and a community screening clinic in Guangdong Province, China. Participants: Participants aged 18 years and over with diabetes who were able to provide informed consent and agreed to attend the dilated eye examination with handheld tests and a standard desktop camera. Primary and secondary outcome measures: Primary outcome was the proportion of those with referable DR (R2 and above) identified by the handheld fundus camera (the index test) compared with the standard camera. Secondary outcome was the comparison of proportion of gradable images obtained from each test. Results: In this study, we examined 304 people (608 eyes) with each of the two cameras under mydriasis. The handheld camera detected 119 eyes (19.5%) with some level of DR, 81 (13.3%) of them were referable, while the standard camera detected 132 eyes (21.7%) with some level of DR and 83 (13.7%) were referable. It seems that the standard camera found more eyes with referable DR, although McNemar's test detected no significant difference between the two cameras. Of the 608 eyes with images obtained by desktop camera, 598 (98.4%) images were of sufficient quality for grading, 12 (1.9%) images were not gradable. By the handheldAbstract : Objectives: To investigate the clinical validity of using a handheld fundus camera to detect diabetic retinopathy (DR) in China. Design and settings: Prospective comparison study of the handheld fundus camera with a standard validated instrument in detection of DR in hospital and a community screening clinic in Guangdong Province, China. Participants: Participants aged 18 years and over with diabetes who were able to provide informed consent and agreed to attend the dilated eye examination with handheld tests and a standard desktop camera. Primary and secondary outcome measures: Primary outcome was the proportion of those with referable DR (R2 and above) identified by the handheld fundus camera (the index test) compared with the standard camera. Secondary outcome was the comparison of proportion of gradable images obtained from each test. Results: In this study, we examined 304 people (608 eyes) with each of the two cameras under mydriasis. The handheld camera detected 119 eyes (19.5%) with some level of DR, 81 (13.3%) of them were referable, while the standard camera detected 132 eyes (21.7%) with some level of DR and 83 (13.7%) were referable. It seems that the standard camera found more eyes with referable DR, although McNemar's test detected no significant difference between the two cameras. Of the 608 eyes with images obtained by desktop camera, 598 (98.4%) images were of sufficient quality for grading, 12 (1.9%) images were not gradable. By the handheld camera, 590 (97.0%) were gradable and 20 (3.2%) images were not gradable. The two cameras reached high agreement on diagnosis of retinopathy and maculopathy at all the levels of retinopathy. Conclusion: Although it could not take the place of standard desktop camera on clinic fundus examination, the handheld fundus camera showed promising role on preliminary DR screening at primary level in China. To ensure quality images, mydriasis is required. … (more)
- Is Part Of:
- BMJ open. Volume 10:Issue 10(2020)
- Journal:
- BMJ open
- Issue:
- Volume 10:Issue 10(2020)
- Issue Display:
- Volume 10, Issue 10 (2020)
- Year:
- 2020
- Volume:
- 10
- Issue:
- 10
- Issue Sort Value:
- 2020-0010-0010-0000
- Page Start:
- Page End:
- Publication Date:
- 2020-10-29
- Subjects:
- diabetic retinopathy -- diabetic nephropathy & vascular disease -- ophthalmology
Medicine -- Research -- Periodicals
610.72 - Journal URLs:
- http://www.bmj.com/archive ↗
http://bmjopen.bmj.com/ ↗ - DOI:
- 10.1136/bmjopen-2020-040196 ↗
- Languages:
- English
- ISSNs:
- 2044-6055
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 16954.xml