Numerical Biomechanics Models of the Interaction Between a Novel Transcatheter Mitral Valve Device and the Subvalvular Apparatus. Issue 4 (July 2021)
- Record Type:
- Journal Article
- Title:
- Numerical Biomechanics Models of the Interaction Between a Novel Transcatheter Mitral Valve Device and the Subvalvular Apparatus. Issue 4 (July 2021)
- Main Title:
- Numerical Biomechanics Models of the Interaction Between a Novel Transcatheter Mitral Valve Device and the Subvalvular Apparatus
- Authors:
- Marom, Gil
Plitman Mayo, Romina
Again, Nadav
Raanani, Ehud - Abstract:
- Objective: Mitral valve regurgitation (MR) is a common valvular heart disease where improper closing causes leakage. Currently, no transcatheter mitral valve device is commercially available. Raanani (co-author) and colleagues have previously proposed a unique rotational implantation, ensuring anchoring by metallic arms that pull the chordae tendineae. This technique is now being implemented in a novel device design. The aim of this study is to quantify the rotational implantation effect on the mitral annulus kinematics and on the stresses in the chordae and papillary muscles. Methods: Finite element analysis of the rotational step of the implantation in a whole heart model is employed to compare 5 arm designs with varying diameters (25.9 mm to 32.4 mm) and rotation angles (up to 140°). The arm rotation that grabs the chordae was modeled when the valve was in systolic configuration. Results: An increase in the rotation angle results in reduced mitral annulus perimeters. Larger rotation angles led to higher chordae stresses with the 29.8 mm experiencing the maximum stresses. The calculated chordae stresses suggest that arm diameter should be <27.8 mm and the rotation angle <120°. Conclusions: The upper limit of this diameter range is preferred in order to reduce the stresses in the papillary muscles while grabbing more chords. The findings of this study can help improving the design and performance of this unique device and procedural technique.
- Is Part Of:
- Innovations. Volume 16:Issue 4(2021)
- Journal:
- Innovations
- Issue:
- Volume 16:Issue 4(2021)
- Issue Display:
- Volume 16, Issue 4 (2021)
- Year:
- 2021
- Volume:
- 16
- Issue:
- 4
- Issue Sort Value:
- 2021-0016-0004-0000
- Page Start:
- 327
- Page End:
- 333
- Publication Date:
- 2021-07
- Subjects:
- heart valves -- transcatheter mitral valve -- numerical models -- biomechanics -- finite element analysis -- cardiovascular devices
Cardiovascular system -- Surgery -- Periodicals
Heart -- Surgery -- Periodicals
Blood-vessels -- Surgery -- Periodicals
Chest -- Surgery -- Periodicals
Thoracic Surgical Procedures -- methods -- Periodicals
Vascular Surgical Procedures -- methods -- Periodicals
Thorax -- Chirurgie -- Méthodologie -- Périodiques
Vaisseaux sanguins -- Chirurgie -- Méthodologie -- Périodiques
Blood-vessels -- Surgery
Chest -- Surgery
Periodicals
617.41 - Journal URLs:
- http://gateway.ovid.com/ovidweb.cgi?T=JS&MODE=ovid&NEWS=n&PAGE=toc&D=ovft&AN=01243895-000000000-00000 ↗
http://journals.lww.com/innovjournal/pages/default.aspx ↗
http://www.lww.com/product/?1556-9845 ↗
http://journals.lww.com/pages/default.aspx ↗ - DOI:
- 10.1177/1556984521999362 ↗
- Languages:
- English
- ISSNs:
- 1556-9845
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 16936.xml