A 3D-transient elastohydrodynamic lubrication hip implant model to compare ultra high molecular weight polyethylene with more compliant polycarbonate polyurethane acetabular cups. (July 2021)
- Record Type:
- Journal Article
- Title:
- A 3D-transient elastohydrodynamic lubrication hip implant model to compare ultra high molecular weight polyethylene with more compliant polycarbonate polyurethane acetabular cups. (July 2021)
- Main Title:
- A 3D-transient elastohydrodynamic lubrication hip implant model to compare ultra high molecular weight polyethylene with more compliant polycarbonate polyurethane acetabular cups
- Authors:
- Ford, Audrey
Hua, Zikai
Ferguson, Stephen J.
Pruitt, Lisa A.
Gao, Leiming - Abstract:
- Abstract: Wear remains a significant challenge in the design of orthopedic implants such as total hip replacements. Early elastohydrodynamic lubrication modeling has predicted thicker lubrication films in hip replacement designs with compliant polycarbonate polyurethane (PCU) bearing materials compared to stiffer materials like ultra-high molecular weight polyethylene (UHMWPE). The predicted thicker lubrication films suggest improved friction and wear performance. However, when compared to the model predictions, experimental wear studies showed mixed results. The mismatch between the model and experimental results may lie in the simplifying assumptions of the early models such as: steady state conditions, one dimensional rotation and loading, and high viscosities. This study applies a 3D-transient elastohydrodynamic model based on an ISO standard gait cycle to better understand the interaction between material stiffness and film thickness in total hip arthroplasty material couples. Similar to previous, simplified models, we show that the average and central film thickness of PCU ( ∼ 0 . 4 μ m ) is higher than that of UHMWPE ( ∼ 0 . 2 μ m ). However, in the 3D-transient model, the film thickness distribution was largely asymmetric and the minimum film thickness occurred outside of the central axis. Although the overall film thickness of PCU was higher than UHMWPE, the minimum film thickness of PCU was lower than UHMPWE for the majority of the gait cycle. The minimum filmAbstract: Wear remains a significant challenge in the design of orthopedic implants such as total hip replacements. Early elastohydrodynamic lubrication modeling has predicted thicker lubrication films in hip replacement designs with compliant polycarbonate polyurethane (PCU) bearing materials compared to stiffer materials like ultra-high molecular weight polyethylene (UHMWPE). The predicted thicker lubrication films suggest improved friction and wear performance. However, when compared to the model predictions, experimental wear studies showed mixed results. The mismatch between the model and experimental results may lie in the simplifying assumptions of the early models such as: steady state conditions, one dimensional rotation and loading, and high viscosities. This study applies a 3D-transient elastohydrodynamic model based on an ISO standard gait cycle to better understand the interaction between material stiffness and film thickness in total hip arthroplasty material couples. Similar to previous, simplified models, we show that the average and central film thickness of PCU ( ∼ 0 . 4 μ m ) is higher than that of UHMWPE ( ∼ 0 . 2 μ m ). However, in the 3D-transient model, the film thickness distribution was largely asymmetric and the minimum film thickness occurred outside of the central axis. Although the overall film thickness of PCU was higher than UHMWPE, the minimum film thickness of PCU was lower than UHMPWE for the majority of the gait cycle. The minimum film thickness of PCU also had a larger range throughout the gait cycle. Both materials were found to be operating between boundary and mixed lubrication regimes. This 3D-transient model reveals a more nuanced interaction between bearing material stiffness and film thickness that supports the mixed results found in experimental wear studies of PCU hip implant designs. … (more)
- Is Part Of:
- Journal of the mechanical behavior of biomedical materials. Volume 119(2021)
- Journal:
- Journal of the mechanical behavior of biomedical materials
- Issue:
- Volume 119(2021)
- Issue Display:
- Volume 119, Issue 2021 (2021)
- Year:
- 2021
- Volume:
- 119
- Issue:
- 2021
- Issue Sort Value:
- 2021-0119-2021-0000
- Page Start:
- Page End:
- Publication Date:
- 2021-07
- Subjects:
- Elastohydrodynamic lubrication -- Orthopedic biomaterials -- Hip arthroplasty -- Polycarbonate polyurethane -- Ultra high molecular weight polyethylene
Biomedical materials -- Periodicals
Biomedical materials -- Mechanical properties -- Periodicals
Biomedical materials
Biomedical materials -- Mechanical properties
Periodicals
Electronic journals
610.28 - Journal URLs:
- http://www.sciencedirect.com/science/journal/17516161 ↗
http://www.elsevier.com/journals ↗ - DOI:
- 10.1016/j.jmbbm.2021.104472 ↗
- Languages:
- English
- ISSNs:
- 1751-6161
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 5015.809000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 16809.xml