Safe procedures despite ultra low radiation doses during catheter ablations of atrial and ventricular arrhythmias—A multicenter experience. Issue 5 (15th March 2021)
- Record Type:
- Journal Article
- Title:
- Safe procedures despite ultra low radiation doses during catheter ablations of atrial and ventricular arrhythmias—A multicenter experience. Issue 5 (15th March 2021)
- Main Title:
- Safe procedures despite ultra low radiation doses during catheter ablations of atrial and ventricular arrhythmias—A multicenter experience
- Authors:
- Attanasio, Philipp
Huemer, Martin
Kaehler, Nora
Keller, Theresa
Schreiber, Tobias
Niehues, Reinhard
Katsani‐Potempa, Dimitra
Klein, Rolf Michael
Landmesser, Ulf
Deisenhofer, Isabel
Tutdibi, Osman
Bourier, Felix - Abstract:
- Abstract: Introduction: Despite the development of non‐fluoroscopic catheter visualization options, fluoroscopy is still used in most ablation procedures. The aim of this multicenter study was to evaluate the safety and efficacy of a new ultra‐low dose radiation protocol for EP procedures in a large number of patients. Methods and results: A total of 3462 consecutive patients (male 1926 (55.6%), age 64.4 ± 14.0 years, BMI 26.65 ± 4.70) undergoing radiofrequency ablation (left atrial (n = 2316 [66.9%], right atrial (n = 675 [19.5%], or ventricular (n = 471 [13.6%]) in three German centers were included in the analysis. Procedures were performed using a new ultra‐low dose protocol operating at 8nGy for fluoroscopy and 36nGy for cine‐loops. Additionally a very low framerate (2‐3FPS) was used. Using the new protocol very low Air kerma‐area product (KAP) values were achieved for left atrial ablations (104.25 ± 84.22 μGym 2 ), right atrial ablations (70.98 ± 94.79 μGym 2 ) and ablations for ventricular tachycardias or PVCs (78.62 ± 66.59 μGym 2 ). Acute procedural success was achieved in 3289/3388 (97.1%) while the rate of major complications was very low compared to previously published studies not using low dose settings (n = 20, 0.6%). Conclusion: The ultra‐low dose, low framerate protocol leads to very low radiation doses for all EP procedures while neither procedural time, fluoroscopy time nor success or complication rates were compromised. When compared to current real‐worldAbstract: Introduction: Despite the development of non‐fluoroscopic catheter visualization options, fluoroscopy is still used in most ablation procedures. The aim of this multicenter study was to evaluate the safety and efficacy of a new ultra‐low dose radiation protocol for EP procedures in a large number of patients. Methods and results: A total of 3462 consecutive patients (male 1926 (55.6%), age 64.4 ± 14.0 years, BMI 26.65 ± 4.70) undergoing radiofrequency ablation (left atrial (n = 2316 [66.9%], right atrial (n = 675 [19.5%], or ventricular (n = 471 [13.6%]) in three German centers were included in the analysis. Procedures were performed using a new ultra‐low dose protocol operating at 8nGy for fluoroscopy and 36nGy for cine‐loops. Additionally a very low framerate (2‐3FPS) was used. Using the new protocol very low Air kerma‐area product (KAP) values were achieved for left atrial ablations (104.25 ± 84.22 μGym 2 ), right atrial ablations (70.98 ± 94.79 μGym 2 ) and ablations for ventricular tachycardias or PVCs (78.62 ± 66.59 μGym 2 ). Acute procedural success was achieved in 3289/3388 (97.1%) while the rate of major complications was very low compared to previously published studies not using low dose settings (n = 20, 0.6%). Conclusion: The ultra‐low dose, low framerate protocol leads to very low radiation doses for all EP procedures while neither procedural time, fluoroscopy time nor success or complication rates were compromised. When compared to current real‐world Air KAP data the new ultra‐low dose fluoroscopy protocol reduces radiation exposure by more than 90%. … (more)
- Is Part Of:
- Pacing and clinical electrophysiology. Volume 44:Issue 5(2021)
- Journal:
- Pacing and clinical electrophysiology
- Issue:
- Volume 44:Issue 5(2021)
- Issue Display:
- Volume 44, Issue 5 (2021)
- Year:
- 2021
- Volume:
- 44
- Issue:
- 5
- Issue Sort Value:
- 2021-0044-0005-0000
- Page Start:
- 807
- Page End:
- 813
- Publication Date:
- 2021-03-15
- Subjects:
- ablation therapy -- antiscatter grid -- framerate reduction -- radiation dose
Cardiac pacing -- Periodicals
Electrophysiology -- Periodicals
617.4120645 - Journal URLs:
- http://onlinelibrary.wiley.com/journal/10.1111/(ISSN)1540-8159 ↗
http://www.blackwell-synergy.com/rd.asp?goto=journal&code=pace ↗
http://www.futuraco.com/journalsf.htm ↗
http://onlinelibrary.wiley.com/ ↗
http://firstsearch.oclc.org ↗
http://firstsearch.oclc.org/journal=0147-8389;screen=info;ECOIP ↗ - DOI:
- 10.1111/pace.14205 ↗
- Languages:
- English
- ISSNs:
- 0147-8389
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 6328.210000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 16749.xml