Development of a designated comparison method for alkaline phosphatase measurements and its application to evaluating routine methods. (3rd May 2021)
- Record Type:
- Journal Article
- Title:
- Development of a designated comparison method for alkaline phosphatase measurements and its application to evaluating routine methods. (3rd May 2021)
- Main Title:
- Development of a designated comparison method for alkaline phosphatase measurements and its application to evaluating routine methods
- Authors:
- Liu, Qingxiang
Guo, Qi
Wang, Jing
Deng, Yuhang
Zeng, Jie
Zhou, Weiyan
Zhao, Haijian
Zhang, Chuanbao - Abstract:
- Abstract: The International Federation of Clinical Chemistry and Laboratory Medicine (IFCC) published the reference measurement procedure (RMP) for ALP measurement in 2011. However, the RMP is of high requirements for laboratories, complicated, time-consuming and high cost of reagents. Many manufacturers do not trace results to the higher procedure. And there is currently no designated comparison method (DCM) for ALP measurement. Thus, the standardization of ALP measurement is hindered. Automatic biochemical analyzers are easy to operate and widely used in clinical laboratories. Therefore, according to the RMP, establishing a DCM based on an automatic biochemical analyzer will be a practical way to establish traceability to the accuracy base and promote the standardization of ALP measurement. On the basis of conforming to the RMP recommended by IFCC as far as possible, the DCM was established based on a Thermo Indiko automatic biochemical analyzer. Performances of the method were validated. The DCM repeatability and within laboratory imprecision was <1% and <2.5%, respectively. For evaluation of trueness, the biases were within the equivalent limits. Measurement procedure comparisons and biases estimation were carried out between the DCM, the RMP, and the six routine methods using a panel of 40 individual human serum samples. The comparisons between the DCM and the RMP gave satisfying results. Compared with the DCM, the relative biases of some routine methods failed to meetAbstract: The International Federation of Clinical Chemistry and Laboratory Medicine (IFCC) published the reference measurement procedure (RMP) for ALP measurement in 2011. However, the RMP is of high requirements for laboratories, complicated, time-consuming and high cost of reagents. Many manufacturers do not trace results to the higher procedure. And there is currently no designated comparison method (DCM) for ALP measurement. Thus, the standardization of ALP measurement is hindered. Automatic biochemical analyzers are easy to operate and widely used in clinical laboratories. Therefore, according to the RMP, establishing a DCM based on an automatic biochemical analyzer will be a practical way to establish traceability to the accuracy base and promote the standardization of ALP measurement. On the basis of conforming to the RMP recommended by IFCC as far as possible, the DCM was established based on a Thermo Indiko automatic biochemical analyzer. Performances of the method were validated. The DCM repeatability and within laboratory imprecision was <1% and <2.5%, respectively. For evaluation of trueness, the biases were within the equivalent limits. Measurement procedure comparisons and biases estimation were carried out between the DCM, the RMP, and the six routine methods using a panel of 40 individual human serum samples. The comparisons between the DCM and the RMP gave satisfying results. Compared with the DCM, the relative biases of some routine methods failed to meet the bias limit derived from biological variation. … (more)
- Is Part Of:
- Scandinavian journal of clinical & laboratory investigation. Volume 81:Number 3(2021)
- Journal:
- Scandinavian journal of clinical & laboratory investigation
- Issue:
- Volume 81:Number 3(2021)
- Issue Display:
- Volume 81, Issue 3 (2021)
- Year:
- 2021
- Volume:
- 81
- Issue:
- 3
- Issue Sort Value:
- 2021-0081-0003-0000
- Page Start:
- 218
- Page End:
- 224
- Publication Date:
- 2021-05-03
- Subjects:
- Alkaline phosphatase -- reference measurement procedure -- designed comparison method -- method comparison -- standardization
Clinical biochemistry -- Periodicals
Physiology, Pathological -- Periodicals
Physiology, Experimental -- Periodicals
Medicine -- Research -- Periodicals
Clinical medicine -- Periodicals
616.0072 - Journal URLs:
- http://informahealthcare.com/loi/clb ↗
http://informahealthcare.com ↗ - DOI:
- 10.1080/00365513.2021.1901305 ↗
- Languages:
- English
- ISSNs:
- 0036-5513
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 8087.500000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 16713.xml