A new method for risk factor assessment of organs at risk including conformity index in radiotherapy treatment plan. Issue 2 (24th June 2021)
- Record Type:
- Journal Article
- Title:
- A new method for risk factor assessment of organs at risk including conformity index in radiotherapy treatment plan. Issue 2 (24th June 2021)
- Main Title:
- A new method for risk factor assessment of organs at risk including conformity index in radiotherapy treatment plan
- Authors:
- Ansari, Shahanawaj
Zope, Mukesh Kumar
Yadav, Neha - Abstract:
- Abstract: Introduction: A comprehensive analysing method has been required since long in the field of radiotherapy. The basic purpose of all techniques has been to deliver the prescribed dose safely to the target volume containing tumour and as well as to reduce dose to organs at risk (OARs). The detailed comparison between different treatment techniques is very difficult and inexplicit as well. The gradual improvement in imaging software has made easy to users to assess spatial arrangement of tumour, critical organs and isodose lines in the form of a single 3D representation that can be observed from all angles. The conformity index (CI) alone cannot provide practical information about treatment plans as it is a single isodose line quantity. Aim: The aim of this study was to develop a new method to assess the degree of damage numerically for OARs along with CI assessment for the target. Materials and Methods: The radiotherapy plans of 30 patients of different sites, diagnosed as cancer, were selected for this study irrespective of gender. Out of 30 cases, 8 plans were of head and neck, 2 were of glyoblastoma (GBM), 10 were of pelvis, 5 were of left breast and other 5 were of oesophagus cancer. The mean age was 42 years ranging from 31 to 72 years. Patient's consents were taken before starting the treatment and carried out this research. Risk factor (RF) for OARs depends on volume of irradiation (VVOI ), total volume of the organ (VTVO ) and tolerance dose (DTDO ). AllAbstract: Introduction: A comprehensive analysing method has been required since long in the field of radiotherapy. The basic purpose of all techniques has been to deliver the prescribed dose safely to the target volume containing tumour and as well as to reduce dose to organs at risk (OARs). The detailed comparison between different treatment techniques is very difficult and inexplicit as well. The gradual improvement in imaging software has made easy to users to assess spatial arrangement of tumour, critical organs and isodose lines in the form of a single 3D representation that can be observed from all angles. The conformity index (CI) alone cannot provide practical information about treatment plans as it is a single isodose line quantity. Aim: The aim of this study was to develop a new method to assess the degree of damage numerically for OARs along with CI assessment for the target. Materials and Methods: The radiotherapy plans of 30 patients of different sites, diagnosed as cancer, were selected for this study irrespective of gender. Out of 30 cases, 8 plans were of head and neck, 2 were of glyoblastoma (GBM), 10 were of pelvis, 5 were of left breast and other 5 were of oesophagus cancer. The mean age was 42 years ranging from 31 to 72 years. Patient's consents were taken before starting the treatment and carried out this research. Risk factor (RF) for OARs depends on volume of irradiation (VVOI ), total volume of the organ (VTVO ) and tolerance dose (DTDO ). All radiotherapy plans (Intensity Modulated Radiotherapy (IMRT) and Volumetric Modulated Arc Therapy (VMAT)) were generated using eclipse planning system, version 11.0 (Varian Medical System, Palo Alto, California, USA). Result: The formula developed to assess degree of damage of OARs including CI of the target is risk factor conformity index (RFC) = CI + RF. In head and neck cases, for right parotid, the maximum value of RF is 1·50 and minimum value is observed as 0·97. Optic nerve, brainstem and spinal cord are completely safe as their RF values are found to be 0 on RF scale. Conclusion: RFC is a comprehensive evaluation tool encompassing a wider range of clinically relevant parameters, isodose volumes and tolerance dose of OARs. It is an advance analysing method to check both the qualitative and quantitative nature of a conformal plan, and at the same time, it assesses the degree of damage of OARs. If RF ≥ 1, then OAR will be completely damaged as a result of irradiation. If RF = 0, then OAR will remain safe totally during the course of irradiation. … (more)
- Is Part Of:
- Journal of radiotherapy in practice. Volume 20:Issue 2(2021)
- Journal:
- Journal of radiotherapy in practice
- Issue:
- Volume 20:Issue 2(2021)
- Issue Display:
- Volume 20, Issue 2 (2021)
- Year:
- 2021
- Volume:
- 20
- Issue:
- 2
- Issue Sort Value:
- 2021-0020-0002-0000
- Page Start:
- 144
- Page End:
- 152
- Publication Date:
- 2021-06-24
- Subjects:
- cancer, -- IMRT, -- isodose line, -- organ at risk, -- risk factor
Radiotherapy -- Periodicals
615.842005 - Journal URLs:
- http://journals.cambridge.org/action/displayJournal?jid=JRP ↗
- DOI:
- 10.1017/S1460396920000205 ↗
- Languages:
- English
- ISSNs:
- 1460-3969
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library HMNTS - ELD Digital store
- Ingest File:
- 16707.xml