A comparative analysis of four activated clotting time measurement devices in cardiac surgery with cardiopulmonary bypass. (September 2021)
- Record Type:
- Journal Article
- Title:
- A comparative analysis of four activated clotting time measurement devices in cardiac surgery with cardiopulmonary bypass. (September 2021)
- Main Title:
- A comparative analysis of four activated clotting time measurement devices in cardiac surgery with cardiopulmonary bypass
- Authors:
- Li, Han
Serrick, Cyril
Rao, Vivek
Yip, Paul M - Abstract:
- Background: In cardiac surgery on cardiopulmonary bypass (CPB), heparin anticoagulation is monitored by point-of-care measurement of activated clotting time (ACT). The objective of this study was to compare four ACT systems in cardiac surgery in terms of their reproducibility, agreement and potential clinical impact at relevant medical decision points. Methods: The study included 40 cardiac surgery patients. Samples were taken at five time points before (T1), after heparinization for CPB (T2, T3, T4), and after heparin reversal (T5). The reproducibility, correlation, and differences in ACT values were assessed with two devices from each of the four ACT systems: Instrumentation Laboratory Hemochron Elite (Hmch), Medtronic HMS Plus (HMS), Abbott i-STAT, and Helena Abrazo. Subrange analyses were performed for low ACT values (results from T1, T5) and high ACT values (results from T2, T3, T4). Results: Within-system analysis showed strong linear correlation between paired measurements (R = 0.968-0.993). However, Hmch showed poorer reproducibility with highest proportion of values that exceed a difference of 10% and highest overall standard error of 74 seconds across the measurement range compared to that of the others (range 39-47 seconds, respectively). For inter-system comparison, using Hmch as reference, ACTs were strongly correlated as follows: HMS (R = 0.938), i-STAT (R = 0.911), and Abrazo (R = 0.911). Agreement analysis in the high ACT range showed HMS tended to haveBackground: In cardiac surgery on cardiopulmonary bypass (CPB), heparin anticoagulation is monitored by point-of-care measurement of activated clotting time (ACT). The objective of this study was to compare four ACT systems in cardiac surgery in terms of their reproducibility, agreement and potential clinical impact at relevant medical decision points. Methods: The study included 40 cardiac surgery patients. Samples were taken at five time points before (T1), after heparinization for CPB (T2, T3, T4), and after heparin reversal (T5). The reproducibility, correlation, and differences in ACT values were assessed with two devices from each of the four ACT systems: Instrumentation Laboratory Hemochron Elite (Hmch), Medtronic HMS Plus (HMS), Abbott i-STAT, and Helena Abrazo. Subrange analyses were performed for low ACT values (results from T1, T5) and high ACT values (results from T2, T3, T4). Results: Within-system analysis showed strong linear correlation between paired measurements (R = 0.968-0.993). However, Hmch showed poorer reproducibility with highest proportion of values that exceed a difference of 10% and highest overall standard error of 74 seconds across the measurement range compared to that of the others (range 39-47 seconds, respectively). For inter-system comparison, using Hmch as reference, ACTs were strongly correlated as follows: HMS (R = 0.938), i-STAT (R = 0.911), and Abrazo (R = 0.911). Agreement analysis in the high ACT range showed HMS tended to have higher ACT values with +11% bias over Hmch, whereas i-STAT (–8% bias) and Abrazo (–13% bias) tended to underestimate. Post-protamine ACT results were dependent on device type where Hmch yielded highest post-protamine ACT (+13% higher than baseline) compared to –16% for HMS, –10% for iSTAT and 0% for Abrazo. Conclusions: Each device had individual reproducibility and biases, which may impact peri-operative heparin management. Careful validation must be undertaken when adopting a different method as decision limits would be affected. Clinicians should also be cautious using ACT as the only indicator for full heparin reversal. … (more)
- Is Part Of:
- Perfusion. Volume 36:Number 6(2021)
- Journal:
- Perfusion
- Issue:
- Volume 36:Number 6(2021)
- Issue Display:
- Volume 36, Issue 6 (2021)
- Year:
- 2021
- Volume:
- 36
- Issue:
- 6
- Issue Sort Value:
- 2021-0036-0006-0000
- Page Start:
- 610
- Page End:
- 619
- Publication Date:
- 2021-09
- Subjects:
- activated clotting time -- cardiac surgery -- heparin -- point-of-care testing device -- protamine
Perfusion (Physiology) -- Periodicals
Blood -- Circulation, Artificial -- Periodicals
Heart -- Surgery -- Periodicals
Extracorporeal Circulation -- Periodicals
Perfusion -- Periodicals
Circulation extracorporelle -- Périodiques
Perfusion -- Périodiques
617.41 - Journal URLs:
- http://prf.sagepub.com/ ↗
http://www.uk.sagepub.com/home.nav ↗ - DOI:
- 10.1177/0267659120949351 ↗
- Languages:
- English
- ISSNs:
- 0267-6591
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 16693.xml