A complex flow phantom for medical imaging: ring vortex phantom design and technical specification. (3rd April 2019)
- Record Type:
- Journal Article
- Title:
- A complex flow phantom for medical imaging: ring vortex phantom design and technical specification. (3rd April 2019)
- Main Title:
- A complex flow phantom for medical imaging: ring vortex phantom design and technical specification
- Authors:
- Ambrogio, Simone
Walker, Adrian
Narracott, Andrew
Ferrari, Simone
Verma, Prashant
Fenner, John - Abstract:
- Abstract: Cardiovascular fluid dynamics exhibit complex flow patterns, such as recirculation and vortices. Quantitative analysis of these complexities supports diagnosis, leading to early prediction of pathologies. Quality assurance of technologies that image such flows is challenging but essential, and to this end, a novel, cost-effective, portable, complex flow phantom is proposed and the design specifications are provided. The vortex ring is the flow of choice because it offers patterns comparable to physiological flows and is stable, predictable, reproducible and controllable. This design employs a piston/cylinder system for vortex ring generation, coupled to an imaging tank full of fluid, for vortex propagation. The phantom is motor-driven and by varying piston speed, piston displacement and orifice size, vortex rings with different characteristics can be produced. Two measurement methods, namely Laser-PIV and an optical/video technique, were used to test the phantom under a combination of configurations. Vortex rings with a range of travelling velocities (approximately 1–80 cm/s) and different output-orifice diameters (10–25 mm) were produced with reproducibility typically better than ±10%. Although ultrasound compatibility has been demonstrated, longer-term ambitions include adapting the design to support comparative studies with different modalities, such as MRA and X-ray-CTA.
- Is Part Of:
- Journal of medical engineering & technology. Volume 43:Number 3(2019)
- Journal:
- Journal of medical engineering & technology
- Issue:
- Volume 43:Number 3(2019)
- Issue Display:
- Volume 43, Issue 3 (2019)
- Year:
- 2019
- Volume:
- 43
- Issue:
- 3
- Issue Sort Value:
- 2019-0043-0003-0000
- Page Start:
- 190
- Page End:
- 201
- Publication Date:
- 2019-04-03
- Subjects:
- Medical imaging phantom -- Doppler ultrasound quality assurance -- vortex ring -- complex flow phantom -- laser PIV
Biomedical engineering -- Periodicals
Medical technology -- Periodicals
610.28 - Journal URLs:
- http://informahealthcare.com/journal/jmt ↗
http://www.tandfonline.com/toc/ijmt20/current ↗
http://informahealthcare.com ↗
http://www.tandf.co.uk/journals/titles/03091902.asp ↗ - DOI:
- 10.1080/03091902.2019.1640309 ↗
- Languages:
- English
- ISSNs:
- 0309-1902
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 5017.057000
British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 16652.xml