Piloting the recording of electrode voltages (REVS) using surface electrodes as a test to identify cochlear implant electrode migration, extra-cochlear electrodes and basal electrodes causing discomfort. (4th May 2021)
- Record Type:
- Journal Article
- Title:
- Piloting the recording of electrode voltages (REVS) using surface electrodes as a test to identify cochlear implant electrode migration, extra-cochlear electrodes and basal electrodes causing discomfort. (4th May 2021)
- Main Title:
- Piloting the recording of electrode voltages (REVS) using surface electrodes as a test to identify cochlear implant electrode migration, extra-cochlear electrodes and basal electrodes causing discomfort
- Authors:
- Grasmeder, Mary
Verschuur, Carl
Ferris, Robyn
Basodan, Sundus
Newman, Tracey
Sanderson, Alan - Abstract:
- Abstract : Objectives: To determine if Electrode Voltage (EV) measurements are potentially suitable as a test for detecting extra-cochlear electrodes in cochlear implants (CIs). Methods: EV measurements were made using surface electrodes in live mode in 17 adult cochlear implant (CI) users. Repeatability, the effects of stimulation level, CI active electrode position, (active) recording electrode position and stimulation mode (for Nucleus devices) were investigated. Results/discussion: Recordings made in monopolar mode showed good repeatability when the active recording electrode was placed on the ipsilateral earlobe; voltages increased linearly with stimulation level as expected. EVs for basal electrodes differed greatly between partially inserted/migrated devices, fully inserted devices with all electrodes activated, and those with deactivated basal electrodes [ χ 2 (2) = 10.2, p < 0.05 for the most basal electrode]. EVs for Nucleus devices were small for electrodes on the array when compared to those for monopolar return electrodes, except for the participant with extra-cochlear electrodes. We argue that fibrosis around the electrode array facilitated current flow across the round window in this case. Conclusion: The test appears to be a viable approach to detect electrode migration and extra-cochlear electrodes in adult CI users and may also be sensitive to discomfort caused by current leakage from the basal end of the cochlea.
- Is Part Of:
- Cochlear implants international. Volume 22:Number 3(2021)
- Journal:
- Cochlear implants international
- Issue:
- Volume 22:Number 3(2021)
- Issue Display:
- Volume 22, Issue 3 (2021)
- Year:
- 2021
- Volume:
- 22
- Issue:
- 3
- Issue Sort Value:
- 2021-0022-0003-0000
- Page Start:
- 157
- Page End:
- 169
- Publication Date:
- 2021-05-04
- Subjects:
- Cochlear implant -- Electrode -- Voltage -- Integrity test -- Partial insertion -- Migration -- Extra-cochlear -- Fibrosis
Cochlear implants -- Periodicals
617.882 - Journal URLs:
- http://onlinelibrary.wiley.com/journal/10.1002/(ISSN)1556-9152 ↗
http://www.ingenta.com/journals/browse/whurr/cii ↗
http://www.ingentaconnect.com/content/maney/cii ↗
http://www.tandfonline.com/loi/ycii20 ↗
http://maneypublishing.com/ ↗ - DOI:
- 10.1080/14670100.2020.1863701 ↗
- Languages:
- English
- ISSNs:
- 1467-0100
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 3292.724200
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 16537.xml