- Abstract: Background: Pediatric and congenital heart disease (CHD) patients have a high rate of transvenous (TV) lead failure. Objective: To determine whether TV lead age can aid risk assessment for lead failure to guide the decision of whether a lead should be replaced or reused at the time of a generator change. Methods: Retrospective cohort study of patients <21 years old undergoing TV device implant from 2000 to 2014 at our institution. Patient, device, and lead variables were collected. Leads were compared in groups based on how many generator changes were completed. Results: A total of 393 leads in 257 patients met inclusion criteria, 60 leads failed (15%). Failed leads were more likely to have not yet undergone generator change ( p = .048). CHD ( p = .045), Tendril lead type ( p = .02) and silicone insulation ( p = .02) were associated with failure. In multivariate analysis, younger leads ( p = .022), number of generator changes ( p = .003), CHD ( p = .005) and silicone insulation ( p = .004) remained significant while Tendril lead type did not ( p = .052). Survival curves show an early decline around 4 years. Conclusions: Lead failure rate in pediatric and CHD patients is high. Leads that have not yet undergone a generator change were more likely to fail in this cohort. The strategy of serial replacement based on lead age needs further research to justify in this population.
- Is Part Of:
- Pacing and clinical electrophysiology. Volume 44:Issue 4(2021)
- Journal:
- Pacing and clinical electrophysiology
- Issue:
- Volume 44:Issue 4(2021)
- Issue Display:
- Volume 44, Issue 4 (2021)
- Year:
- 2021
- Volume:
- 44
- Issue:
- 4
- Issue Sort Value:
- 2021-0044-0004-0000
- Page Start:
- 586
- Page End:
- 594
- Publication Date:
- 2021-02-24
- Subjects:
- adult congenital heart disease -- pediatric cardiology -- transvenous cardiac devices
Cardiac pacing -- Periodicals
Electrophysiology -- Periodicals
617.4120645 - Journal URLs:
- http://onlinelibrary.wiley.com/journal/10.1111/(ISSN)1540-8159 ↗
http://www.blackwell-synergy.com/rd.asp?goto=journal&code=pace ↗
http://www.futuraco.com/journalsf.htm ↗
http://onlinelibrary.wiley.com/ ↗
http://firstsearch.oclc.org ↗
http://firstsearch.oclc.org/journal=0147-8389;screen=info;ECOIP ↗ - DOI:
- 10.1111/pace.14166 ↗
- Languages:
- English
- ISSNs:
- 0147-8389
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 6328.210000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 16354.xml