Development of Enhanced, Permanently-Installed, Neutron Activation Diagnostic Hardware for NIF. (May 2016)
- Record Type:
- Journal Article
- Title:
- Development of Enhanced, Permanently-Installed, Neutron Activation Diagnostic Hardware for NIF. (May 2016)
- Main Title:
- Development of Enhanced, Permanently-Installed, Neutron Activation Diagnostic Hardware for NIF
- Authors:
- Edwards, E R
Jedlovec, D R
Carrera, J A
Yeamans, C B - Abstract:
- Abstract: Neutron activation diagnostics are baseline neutron yield and flux measurement instruments at the National Ignition Facility. Up to 19 activation samples are distributed around the target chamber. Currently the samples must be removed to be counted, creating a 1-2 week data turn-around time and considerable labor costs. An improved system consisting of a commercially available LaBr3 (Ce) scintillator and Power over Ethernet electronics is under development. A machined zirconium-702 cap over the detector is the activation medium to measure the 90 Zr(n, 2n) 89 Zr reaction. The detectors are located at the current neutron activation diagnostic sites and monitored remotely. Because they collect data in real time yield values are returned within a few hours after a NIF shot.
- Is Part Of:
- Journal of physics. Volume 717(2016)
- Journal:
- Journal of physics
- Issue:
- Volume 717(2016)
- Issue Display:
- Volume 717, Issue 1 (2016)
- Year:
- 2016
- Volume:
- 717
- Issue:
- 1
- Issue Sort Value:
- 2016-0717-0001-0000
- Page Start:
- Page End:
- Publication Date:
- 2016-05
- Subjects:
- Physics -- Congresses
530.5 - Journal URLs:
- http://www.iop.org/EJ/journal/1742-6596 ↗
http://ioppublishing.org/ ↗ - DOI:
- 10.1088/1742-6596/717/1/012091 ↗
- Languages:
- English
- ISSNs:
- 1742-6588
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 5036.223000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 16275.xml