Confirmation of no impact from different anticoagulant counter ions on bioanalytical method. (April 2012)
- Record Type:
- Journal Article
- Title:
- Confirmation of no impact from different anticoagulant counter ions on bioanalytical method. (April 2012)
- Main Title:
- Confirmation of no impact from different anticoagulant counter ions on bioanalytical method
- Authors:
- Bergeron, Annik
Bergeron, Mélanie
Garofolo, Fabio - Abstract:
- Background: In the past several years, the impact of changing counter ions while keeping the same anticoagulant in bioanalytical LC–MS/MS methods has become a highly discussed topic. In order to confirm that there is no impact from counter ions, matrix effect and stability evaluations were performed on bicalutamide LC–MS/MS bioanalytical methods.Results: Independently from the anticoagulant counter ion used, the matrix effect evaluation met acceptance criteria, even when using conditions expected to increase matrix effect, such as protein precipitation with an analog internal standard. Freeze–thaw along with storage stabilities, namely short- and long-term, demonstrated less than 8% deviation regardless of the counter ion used.Conclusion: Differences in the anticoagulant counter ion used has no impact on the bicalutamide bioanalytical LC–MS/MS method.
- Is Part Of:
- Bioanalysis. Volume 4:Number 8(2012)
- Journal:
- Bioanalysis
- Issue:
- Volume 4:Number 8(2012)
- Issue Display:
- Volume 4, Issue 8 (2012)
- Year:
- 2012
- Volume:
- 4
- Issue:
- 8
- Issue Sort Value:
- 2012-0004-0008-0000
- Page Start:
- 897
- Page End:
- 908
- Publication Date:
- 2012-04
- Subjects:
- Drugs -- Analysis -- Periodicals
615.19005 - Journal URLs:
- http://www.future-science.com/loi/bio ↗
http://www.future-science.com/ ↗ - DOI:
- 10.4155/bio.12.54 ↗
- Languages:
- English
- ISSNs:
- 1757-6180
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 16274.xml