Valve Failure in the Gas Monitor: An Unusual Cause of Falsely Low Capnography Values Under General Anesthesia. Issue 6 (April 2020)
- Record Type:
- Journal Article
- Title:
- Valve Failure in the Gas Monitor: An Unusual Cause of Falsely Low Capnography Values Under General Anesthesia. Issue 6 (April 2020)
- Main Title:
- Valve Failure in the Gas Monitor
- Authors:
- Neira, Victor M.
Fagan, Brian
Milne, Andrew D. - Abstract:
- Abstract : We present a case in which the Dräger Primus (Dräger Medical AG&Co KG, Lüberck, Germany) anesthesia monitor displayed false readings of low end-tidal carbon dioxide (EtCO2 ) immediately after intubation. The patient's physical examination, vital signs, and arterial blood gases were normal. The ventilator, circuit, gas sample line, and water trap appeared intact and functional. A second gas analyzer revealed normal EtCO2 readings, correlating with blood gas values. The defective monitor gas flow rates and gas calibration values were inappropriately low. Partial opening of the solenoid zero valve allowed entrainment of room air, which caused artifactual dilution of the gas sample.
- Is Part Of:
- A & A practice. Volume 14:Issue 6(2020)
- Journal:
- A & A practice
- Issue:
- Volume 14:Issue 6(2020)
- Issue Display:
- Volume 14, Issue 6 (2020)
- Year:
- 2020
- Volume:
- 14
- Issue:
- 6
- Issue Sort Value:
- 2020-0014-0006-0000
- Page Start:
- Page End:
- Publication Date:
- 2020-04
- Subjects:
- Analgesia -- Periodicals
Analgesia -- Case studies -- Periodicals
Anesthesiology -- Periodicals
Anesthesiology -- Case studies -- Periodicals
617.9605 - Journal URLs:
- http://journals.lww.com/pages/default.aspx ↗
- DOI:
- 10.1213/XAA.0000000000001173 ↗
- Languages:
- English
- ISSNs:
- 2575-3126
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 0537.002900
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 16043.xml