Clinical outcomes and programming strategies of implantable cardioverter-defibrillator devices in paediatric hypertrophic cardiomyopathy: a UK National Cohort Study. Issue 3 (22nd November 2020)
- Record Type:
- Journal Article
- Title:
- Clinical outcomes and programming strategies of implantable cardioverter-defibrillator devices in paediatric hypertrophic cardiomyopathy: a UK National Cohort Study. Issue 3 (22nd November 2020)
- Main Title:
- Clinical outcomes and programming strategies of implantable cardioverter-defibrillator devices in paediatric hypertrophic cardiomyopathy: a UK National Cohort Study
- Authors:
- Norrish, Gabrielle
Chubb, Henry
Field, Ella
McLeod, Karen
Ilina, Maria
Spentzou, Georgia
Till, Jan
Daubeney, Piers E F
Stuart, Alan Graham
Matthews, Jane
Hares, Dominic
Brown, Elspeth
Linter, Katie
Bhole, Vinay
Pillai, Krishnakumar
Bowes, Michael
Jones, Caroline B
Uzun, Orhan
Wong, Amos
Yue, Arthur
Sadagopan, Shankar
Bharucha, Tara
Yap, Norah
Rosenthal, Eric
Mathur, Sujeev
Adwani, Satish
Reinhardt, Zdenka
Mangat, Jasveer
Kaski, Juan Pablo - Abstract:
- Abstract: Aims: Sudden cardiac death (SCD) is the most common mode of death in paediatric hypertrophic cardiomyopathy (HCM). This study describes the implant and programming strategies with clinical outcomes following implantable cardioverter-defibrillator (ICD) insertion in a well-characterized national paediatric HCM cohort. Methods and results: Data from 90 patients undergoing ICD insertion at a median age 13 (±3.5) for primary ( n = 67, 74%) or secondary prevention ( n = 23, 26%) were collected from a retrospective, longitudinal multi-centre cohort of children (<16 years) with HCM from the UK. Seventy-six (84%) had an endovascular system [14 (18%) dual coil], 3 (3%) epicardial, and 11 (12%) subcutaneous system. Defibrillation threshold (DFT) testing was performed at implant in 68 (76%). Inadequate DFT in four led to implant adjustment in three patients. Over a median follow-up of 54 months (interquartile range 28–111), 25 (28%) patients had 53 appropriate therapies [ICD shock n = 45, anti-tachycardia pacing (ATP) n = 8], incidence rate 4.7 per 100 patient years (95% CI 2.9–7.6). Eight inappropriate therapies occurred in 7 (8%) patients (ICD shock n = 4, ATP n = 4), incidence rate 1.1/100 patient years (95% CI 0.4–2.5). Three patients (3%) died following arrhythmic events, despite a functioning device. Other device complications were seen in 28 patients (31%), including lead-related complications ( n = 15) and infection ( n = 10). No clinical, device, orAbstract: Aims: Sudden cardiac death (SCD) is the most common mode of death in paediatric hypertrophic cardiomyopathy (HCM). This study describes the implant and programming strategies with clinical outcomes following implantable cardioverter-defibrillator (ICD) insertion in a well-characterized national paediatric HCM cohort. Methods and results: Data from 90 patients undergoing ICD insertion at a median age 13 (±3.5) for primary ( n = 67, 74%) or secondary prevention ( n = 23, 26%) were collected from a retrospective, longitudinal multi-centre cohort of children (<16 years) with HCM from the UK. Seventy-six (84%) had an endovascular system [14 (18%) dual coil], 3 (3%) epicardial, and 11 (12%) subcutaneous system. Defibrillation threshold (DFT) testing was performed at implant in 68 (76%). Inadequate DFT in four led to implant adjustment in three patients. Over a median follow-up of 54 months (interquartile range 28–111), 25 (28%) patients had 53 appropriate therapies [ICD shock n = 45, anti-tachycardia pacing (ATP) n = 8], incidence rate 4.7 per 100 patient years (95% CI 2.9–7.6). Eight inappropriate therapies occurred in 7 (8%) patients (ICD shock n = 4, ATP n = 4), incidence rate 1.1/100 patient years (95% CI 0.4–2.5). Three patients (3%) died following arrhythmic events, despite a functioning device. Other device complications were seen in 28 patients (31%), including lead-related complications ( n = 15) and infection ( n = 10). No clinical, device, or programming characteristics predicted time to inappropriate therapy or lead complication. Conclusion: In a large national cohort of paediatric HCM patients with an ICD, device and programming strategies varied widely. No particular strategy was associated with inappropriate therapies, missed/delayed therapies, or lead complications. … (more)
- Is Part Of:
- Europace. Volume 23:Issue 3(2021)
- Journal:
- Europace
- Issue:
- Volume 23:Issue 3(2021)
- Issue Display:
- Volume 23, Issue 3 (2021)
- Year:
- 2021
- Volume:
- 23
- Issue:
- 3
- Issue Sort Value:
- 2021-0023-0003-0000
- Page Start:
- 400
- Page End:
- 408
- Publication Date:
- 2020-11-22
- Subjects:
- Hypertrophic cardiomyopathy -- Implantable cardioverter-defibrillator -- Programming -- Childhood
Arrhythmia -- Treatment -- Periodicals
Cardiac pacing -- Periodicals
Catheter ablation -- Periodicals
Heart -- Physiology -- Periodicals
Electrophysiology -- Periodicals
617.4120645 - Journal URLs:
- http://europace.oxfordjournals.org/ ↗
http://ukcatalogue.oup.com/ ↗ - DOI:
- 10.1093/europace/euaa307 ↗
- Languages:
- English
- ISSNs:
- 1099-5129
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 3829.340450
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 15978.xml