A compact furnace for in situ X‐ray absorption spectroscopy: design, fabrication and study of cationic oxidation states in Pr6O11 and NiO. (14th January 2021)
- Record Type:
- Journal Article
- Title:
- A compact furnace for in situ X‐ray absorption spectroscopy: design, fabrication and study of cationic oxidation states in Pr6O11 and NiO. (14th January 2021)
- Main Title:
- A compact furnace for in situ X‐ray absorption spectroscopy: design, fabrication and study of cationic oxidation states in Pr6O11 and NiO
- Authors:
- Sattayaporn, Suchinda
Rodporn, Somboonsup
Kidkhunthod, Pinit
Chanlek, Narong
Yonchai, Chutarat
Rujirawat, Saroj - Abstract:
- Abstract : A well designed compact furnace has been designed for in situ X‐ray absorption spectroscopy at various temperatures up to 1473 K and under controlled atmosphere. Abstract : A well designed compact furnace has been designed for in situ X‐ray absorption spectroscopy (XAS). It enables various heat ramps from 300 K to 1473 K. The furnace consists of heaters, a quartz tube, a circulated refrigerator and a power controller. It can generate ohmic heating via an induction process with tantalum filaments. The maximum heating rate exceeds 20 K min −1 . A quartz tube with gas feedthroughs allows the mixing of gases and adjustment of the flow rate. The use of this compact furnace allows in situ XAS investigations to be carried out in transmission or fluorescence modes under controlled temperature and atmosphere. Moreover, the furnace is compact, light and well compatible to XAS. The furnace was used to study cationic oxidation states in Pr6 O11 and NiO compounds under elevated temperature and reduced atmosphere using the in situ X‐ray absorption near‐edge structure (XANES) technique at beamline 5.2 SUT‐NANOTEC‐SLRI of the Synchrotron Light Research Institute, Thailand. At room temperature, Pr6 O11 contains a mixture of Pr 3+ and Pr 4+ cations, resulting in an average oxidation state of +3.67. In situ XANES spectra of Pr ( L 3 ‐edge) show that the oxidation state of Pr 4+ cations was totally reduced to +3.00 at 1273 K under H2 atmosphere. Considering NiO, Ni 2+ species wereAbstract : A well designed compact furnace has been designed for in situ X‐ray absorption spectroscopy at various temperatures up to 1473 K and under controlled atmosphere. Abstract : A well designed compact furnace has been designed for in situ X‐ray absorption spectroscopy (XAS). It enables various heat ramps from 300 K to 1473 K. The furnace consists of heaters, a quartz tube, a circulated refrigerator and a power controller. It can generate ohmic heating via an induction process with tantalum filaments. The maximum heating rate exceeds 20 K min −1 . A quartz tube with gas feedthroughs allows the mixing of gases and adjustment of the flow rate. The use of this compact furnace allows in situ XAS investigations to be carried out in transmission or fluorescence modes under controlled temperature and atmosphere. Moreover, the furnace is compact, light and well compatible to XAS. The furnace was used to study cationic oxidation states in Pr6 O11 and NiO compounds under elevated temperature and reduced atmosphere using the in situ X‐ray absorption near‐edge structure (XANES) technique at beamline 5.2 SUT‐NANOTEC‐SLRI of the Synchrotron Light Research Institute, Thailand. At room temperature, Pr6 O11 contains a mixture of Pr 3+ and Pr 4+ cations, resulting in an average oxidation state of +3.67. In situ XANES spectra of Pr ( L 3 ‐edge) show that the oxidation state of Pr 4+ cations was totally reduced to +3.00 at 1273 K under H2 atmosphere. Considering NiO, Ni 2+ species were present under ambient conditions. At 573 K, the reduction process of Ni 2+ occurred. The Ni 0 /Ni 2+ ratio increased linearly with respect to the heating temperature. Finally, the reduction process of Ni 2+ was completely finished at 770 K. … (more)
- Is Part Of:
- Journal of synchrotron radiation. Volume 28:Part 2(2021)
- Journal:
- Journal of synchrotron radiation
- Issue:
- Volume 28:Part 2(2021)
- Issue Display:
- Volume 28, Issue 2, Part 2 (2021)
- Year:
- 2021
- Volume:
- 28
- Issue:
- 2
- Part:
- 2
- Issue Sort Value:
- 2021-0028-0002-0002
- Page Start:
- 455
- Page End:
- 460
- Publication Date:
- 2021-01-14
- Subjects:
- X‐ray absorption spectroscopy -- in situ cells -- transmittance and fluorescence -- instrumentations -- in situ XAFS
Synchrotron radiation -- Periodicals
Free electron lasers -- Periodicals
539.73505 - Journal URLs:
- http://onlinelibrary.wiley.com/journal/10.1107/S16005775 ↗
http://journals.iucr.org/s/journalhomepage.html ↗
http://www.blackwell-synergy.com/openurl?genre=journal&issn=0909-0495 ↗
http://onlinelibrary.wiley.com/ ↗
http://firstsearch.oclc.org ↗ - DOI:
- 10.1107/S1600577520015696 ↗
- Languages:
- English
- ISSNs:
- 0909-0495
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 5068.035000
British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 15963.xml