Field evaluation of a rapid antigen test (Panbio™ COVID-19 Ag Rapid Test Device) for COVID-19 diagnosis in primary healthcare centres. (March 2021)
- Record Type:
- Journal Article
- Title:
- Field evaluation of a rapid antigen test (Panbio™ COVID-19 Ag Rapid Test Device) for COVID-19 diagnosis in primary healthcare centres. (March 2021)
- Main Title:
- Field evaluation of a rapid antigen test (Panbio™ COVID-19 Ag Rapid Test Device) for COVID-19 diagnosis in primary healthcare centres
- Authors:
- Albert, Eliseo
Torres, Ignacio
Bueno, Felipe
Huntley, Dixie
Molla, Estefanía
Fernández-Fuentes, Miguel Ángel
Martínez, Mireia
Poujois, Sandrine
Forqué, Lorena
Valdivia, Arantxa
Solano de la Asunción, Carlos
Ferrer, Josep
Colomina, Javier
Navarro, David - Abstract:
- Abstract: Objectives: To our knowledge no previous study has assessed the performance of a rapid antigen diagnostic immunoassay (RAD) conducted at the point of care (POC). We evaluated the Panbio™ COVID-19 Ag Rapid Test Device for diagnosis of coronavirus 2019 disease (COVID-19) in symptomatic patients ( n = 412) attending primary healthcare centres. Methods: RAD was performed immediately after sampling following the manufacturer's instructions (reading at 15 min). RT-PCRs were carried out within 24 h of specimen collection. Samples displaying discordant results were processed for culture in Vero E6 cells. Presence of severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) in cell cultures was confirmed by RT-PCR. Results: Out of 412 patients, 43 (10.4%) tested positive by RT-PCR and RAD, and 358 (86.9%) tested negative by both methods; discordant results (RT-PCR+/RAD–) were obtained in 11 patients (2.7%). Overall specificity and sensitivity of rapid antigen detection (RAD) was 100% (95%CI 98.7–100%) and 79.6% (95%CI 67.0–88.8%), respectively, taking RT-PCR as the reference. Overall RAD negative predictive value for an estimated prevalence of 5% and 10% was 99% (95%CI 97.4–99.6%) and 97.9% (95%CI 95.9–98.9), respectively. SARS-CoV-2 could not be cultured from specimens yielding RT-PCR+/RAD– results ( n = 11). Conclusion: The Panbio™ COVID-19 Ag Rapid Test Device performed well as a POC test for early diagnosis of COVID-19 in primary healthcare centres. More crucially,Abstract: Objectives: To our knowledge no previous study has assessed the performance of a rapid antigen diagnostic immunoassay (RAD) conducted at the point of care (POC). We evaluated the Panbio™ COVID-19 Ag Rapid Test Device for diagnosis of coronavirus 2019 disease (COVID-19) in symptomatic patients ( n = 412) attending primary healthcare centres. Methods: RAD was performed immediately after sampling following the manufacturer's instructions (reading at 15 min). RT-PCRs were carried out within 24 h of specimen collection. Samples displaying discordant results were processed for culture in Vero E6 cells. Presence of severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) in cell cultures was confirmed by RT-PCR. Results: Out of 412 patients, 43 (10.4%) tested positive by RT-PCR and RAD, and 358 (86.9%) tested negative by both methods; discordant results (RT-PCR+/RAD–) were obtained in 11 patients (2.7%). Overall specificity and sensitivity of rapid antigen detection (RAD) was 100% (95%CI 98.7–100%) and 79.6% (95%CI 67.0–88.8%), respectively, taking RT-PCR as the reference. Overall RAD negative predictive value for an estimated prevalence of 5% and 10% was 99% (95%CI 97.4–99.6%) and 97.9% (95%CI 95.9–98.9), respectively. SARS-CoV-2 could not be cultured from specimens yielding RT-PCR+/RAD– results ( n = 11). Conclusion: The Panbio™ COVID-19 Ag Rapid Test Device performed well as a POC test for early diagnosis of COVID-19 in primary healthcare centres. More crucially, the data suggested that patients with RT-PCR-proven COVID-19 testing negative by RAD are unlikely to be infectious. … (more)
- Is Part Of:
- Clinical microbiology and infection. Volume 27:Number 3(2021)
- Journal:
- Clinical microbiology and infection
- Issue:
- Volume 27:Number 3(2021)
- Issue Display:
- Volume 27, Issue 3 (2021)
- Year:
- 2021
- Volume:
- 27
- Issue:
- 3
- Issue Sort Value:
- 2021-0027-0003-0000
- Page Start:
- 472.e7
- Page End:
- 472.e10
- Publication Date:
- 2021-03
- Subjects:
- COVID-19 -- Early diagnosis -- Primary healthcare centre -- Rapid antigen detection test (RAD) -- SARS-CoV-2
Medical microbiology -- Periodicals
Diagnostic microbiology -- Periodicals
Communicable diseases -- Periodicals
Infection -- Periodicals
616.01 - Journal URLs:
- http://onlinelibrary.wiley.com/journal/10.1111/(ISSN)1469-0691 ↗
http://onlinelibrary.wiley.com/ ↗ - DOI:
- 10.1016/j.cmi.2020.11.004 ↗
- Languages:
- English
- ISSNs:
- 1198-743X
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 3286.305520
British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 15945.xml