First implantation of retinal prosthesis in a patient with high myopia after surgery and rehabilitation program in Taiwan. Issue 7 (July 2019)
- Record Type:
- Journal Article
- Title:
- First implantation of retinal prosthesis in a patient with high myopia after surgery and rehabilitation program in Taiwan. Issue 7 (July 2019)
- Main Title:
- First implantation of retinal prosthesis in a patient with high myopia after surgery and rehabilitation program in Taiwan
- Authors:
- Hsu, Po-Cheng
Chen, Po-Yin
Chung, Yi-Shun
Lin, Tai-Chi
Hwang, De-Kuang
Chen, Shih-Jen
Kao, Chung-Lan - Abstract:
- Abstract : The implantation of the Argus II retinal prosthesis system in a 54-year-old woman with retinitis pigmentosa who presented with an eye of long axial length at 26.82 mm was successful. Postoperative examination revealed a gap of 700 µm between the electrode array and the retina, which caused decreased visual perception. A modified strategy with quad and quinary electrode stimulation was introduced to generate higher perceptual thresholds. The patient experienced visual functional changes during the first half of the year after surgery, although no remarkable difference was observed in terms of implant–retina distance. Fibrosis around the tack developed and extended between the gap with the retina elevated from the tack toward the center array, 8 months after the surgery. Schisis of the retina developed and filled the gap, resulting in decreased threshold, and the strategy was then shifted back to single electrode stimulation mode. Rehabilitation program is an evolving process that depends on the distance between the array and the retina in the eye with staphyloma. This study first showed the implantation in a patient with high myopia who presented with long axial length after surgery and rehabilitation program in Taiwan.
- Is Part Of:
- Journal of the Chinese Medical Association. Volume 82:Issue 7(2019)
- Journal:
- Journal of the Chinese Medical Association
- Issue:
- Volume 82:Issue 7(2019)
- Issue Display:
- Volume 82, Issue 7 (2019)
- Year:
- 2019
- Volume:
- 82
- Issue:
- 7
- Issue Sort Value:
- 2019-0082-0007-0000
- Page Start:
- Page End:
- Publication Date:
- 2019-07
- Subjects:
- Case Report -- Long ocular axial length -- Retina prosthesis -- Retinitis pigmentosa -- Visual rehabilitation
Medicine -- Periodicals
610.5 - Journal URLs:
- https://journals.lww.com/jcma/pages/default.aspx ↗
http://www.elsevier.com/journals ↗ - DOI:
- 10.1097/JCMA.0000000000000091 ↗
- Languages:
- English
- ISSNs:
- 1726-4901
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 4729.330050
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 15933.xml