Reference dosimetry in MRI-linacs: evaluation of available protocols and data to establish a Code of Practice. (26th February 2021)
- Record Type:
- Journal Article
- Title:
- Reference dosimetry in MRI-linacs: evaluation of available protocols and data to establish a Code of Practice. (26th February 2021)
- Main Title:
- Reference dosimetry in MRI-linacs: evaluation of available protocols and data to establish a Code of Practice
- Authors:
- de Pooter, Jacco
Billas, Ilias
de Prez, Leon
Duane, Simon
Kapsch, Ralf-Peter
Karger, Christian P
van Asselen, Bram
Wolthaus, Jochem - Abstract:
- Abstract: With the rapid increase in clinical treatments with MRI-linacs, a consistent, harmonized and sustainable ground for reference dosimetry in MRI-linacs is needed. Specific for reference dosimetry in MRI-linacs is the presence of a strong magnetic field. Therefore, existing Code of Practices (CoPs) are inadequate. In recent years, a vast amount of papers have been published in relation to this topic. The purpose of this review paper is twofold: to give an overview and evaluate the existing literature for reference dosimetry in MRI-linacs and to discuss whether the literature and datasets are adequate and complete to serve as a basis for the development of a new or to extend existing CoPs. This review is prefaced with an overview of existing MRI-linac facilities. Then an introduction on the physics of radiation transport in magnetic fields is given. The main part of the review is devoted to the evaluation of the literature with respect to the following subjects: beam characteristics of MRI-linac facilities; formalisms for reference dosimetry in MRI-linacs; characteristics of ionization chambers in the presence of magnetic fields; ionization chamber beam quality correction factors; and ionization chamber magnetic field correction factors. The review is completed with a discussion as to whether the existing literature is adequate to serve as basis for a CoP. In addition, it highlights subjects for future research on this topic.
- Is Part Of:
- Physics in medicine & biology. Volume 66:Number 5(2021)
- Journal:
- Physics in medicine & biology
- Issue:
- Volume 66:Number 5(2021)
- Issue Display:
- Volume 66, Issue 5 (2021)
- Year:
- 2021
- Volume:
- 66
- Issue:
- 5
- Issue Sort Value:
- 2021-0066-0005-0000
- Page Start:
- Page End:
- Publication Date:
- 2021-02-26
- Subjects:
- reference dosimetry -- MRI-linacs -- MR-guided radiotherapy -- Monte Carlo -- ionization chambers -- Code of Practice -- primary standards
Biophysics -- Periodicals
Medical physics -- Periodicals
610.153 - Journal URLs:
- http://ioppublishing.org/ ↗
http://iopscience.iop.org/0031-9155 ↗ - DOI:
- 10.1088/1361-6560/ab9efe ↗
- Languages:
- English
- ISSNs:
- 0031-9155
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 15864.xml