The influence of Acuros XB on dose volume histogram metrics and tumour control probability modelling in locally advanced non-small cell lung cancer. (January 2021)
- Record Type:
- Journal Article
- Title:
- The influence of Acuros XB on dose volume histogram metrics and tumour control probability modelling in locally advanced non-small cell lung cancer. (January 2021)
- Main Title:
- The influence of Acuros XB on dose volume histogram metrics and tumour control probability modelling in locally advanced non-small cell lung cancer
- Authors:
- Fleming, Cathy
O'Keeffe, Serena
McDermott, Ronan
Dunne, Mary
McClean, Brendan
León Vintró, Luis - Abstract:
- Highlights: Classic metrics for evaluation of non-small cell treatment plans derived older algorithms. Treatment planning with more accurate algorithms such as Acuros now routinely possible. How does the use of older evaluation metrics affect modern plans? Should we be revising our clinical evaluation metrics in order to continue to align with the outcome data? Abstract: Purpose: Radiation therapy plans are assessed using dose volume metrics derived from clinical toxicity and outcome data. In this study, plans for patients with locally advanced non-small cell lung cancer (LA-NSCLC) are examined in the context of the implementation of the Acuros XB (AXB) dose calculation algorithm focussing on the impact on common metrics. Methods: Volumetric modulated arc therapy (VMAT) plans were generated for twenty patients, using the Analytical Anisotropic Algorithm (AAA) and recalculated with AXB for both dose to water (Dw ) and dose to medium (Dm ). Standard dose volume histogram (DVH) metrics for both targets and organs-at-risk (OARs) were extracted, in addition to tumour control probability (TCP) for targets. Results: Mean dose to the planning target volume (PTV) was not clinically different between the algorithms (within ±1.1 Gy) but differences were seen in the minimum dose, D99% and D98% as well as for conformity and homogeneity metrics. A difference in TCP was seen for AXBDm plans versus both AXBDw and AAA plans. No clinically relevant differences were seen in the lung metrics.Highlights: Classic metrics for evaluation of non-small cell treatment plans derived older algorithms. Treatment planning with more accurate algorithms such as Acuros now routinely possible. How does the use of older evaluation metrics affect modern plans? Should we be revising our clinical evaluation metrics in order to continue to align with the outcome data? Abstract: Purpose: Radiation therapy plans are assessed using dose volume metrics derived from clinical toxicity and outcome data. In this study, plans for patients with locally advanced non-small cell lung cancer (LA-NSCLC) are examined in the context of the implementation of the Acuros XB (AXB) dose calculation algorithm focussing on the impact on common metrics. Methods: Volumetric modulated arc therapy (VMAT) plans were generated for twenty patients, using the Analytical Anisotropic Algorithm (AAA) and recalculated with AXB for both dose to water (Dw ) and dose to medium (Dm ). Standard dose volume histogram (DVH) metrics for both targets and organs-at-risk (OARs) were extracted, in addition to tumour control probability (TCP) for targets. Results: Mean dose to the planning target volume (PTV) was not clinically different between the algorithms (within ±1.1 Gy) but differences were seen in the minimum dose, D99% and D98% as well as for conformity and homogeneity metrics. A difference in TCP was seen for AXBDm plans versus both AXBDw and AAA plans. No clinically relevant differences were seen in the lung metrics. For point doses to spinal cord and oesophagus, the AXBDm values were lower than AXBDw, by up to 1.0 Gy. Conclusion: Normalisation of plans to the mean/median dose to the target does not need to be adjusted when moving from AAA to AXB. OAR point doses may decrease by up to 1 Gy with AXBDm, which can be accounted for in clinical planning. Other OAR metrics do not need to be adjusted. … (more)
- Is Part Of:
- Physica medica. Volume 81(2021)
- Journal:
- Physica medica
- Issue:
- Volume 81(2021)
- Issue Display:
- Volume 81, Issue 2021 (2021)
- Year:
- 2021
- Volume:
- 81
- Issue:
- 2021
- Issue Sort Value:
- 2021-0081-2021-0000
- Page Start:
- 295
- Page End:
- 301
- Publication Date:
- 2021-01
- Subjects:
- VMAT -- NSCLC -- AAA -- Acuros -- DVCs -- TCP
Medical physics -- Periodicals
Biophysics -- Periodicals
Biophysics -- Periodicals
Imagerie médicale -- Périodiques
Radiothérapie -- Périodiques
Rayons X -- Sécurité -- Mesures -- Périodiques
Physique -- Périodiques
Médecine -- Périodiques
610.153 - Journal URLs:
- http://www.sciencedirect.com/science/journal/11201797 ↗
http://www.clinicalkey.com/dura/browse/journalIssue/11201797 ↗
http://www.clinicalkey.com.au/dura/browse/journalIssue/11201797 ↗
http://www.elsevier.com/journals ↗
http://www.physicamedica.com ↗ - DOI:
- 10.1016/j.ejmp.2020.11.030 ↗
- Languages:
- English
- ISSNs:
- 1120-1797
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 6475.070000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 15857.xml