Effect of an integral quality monitor on 4‐, 6‐, 10‐MV, and 6‐MV flattening filter‐free photon beams. Issue 1 (3rd December 2020)
- Record Type:
- Journal Article
- Title:
- Effect of an integral quality monitor on 4‐, 6‐, 10‐MV, and 6‐MV flattening filter‐free photon beams. Issue 1 (3rd December 2020)
- Main Title:
- Effect of an integral quality monitor on 4‐, 6‐, 10‐MV, and 6‐MV flattening filter‐free photon beams
- Authors:
- Nguyen, Trang Hong Thi
Yokoyama, Haruna
Kojima, Hironori
Isomura, Naoki
Takemura, Akihiro
Ueda, Shinichi
Noto, Kimiya - Abstract:
- Abstract: Purpose: To investigate the effect of an integral quality monitor (IQM; iRT Systems GmbH, Koblenz, Germany) on 4, 6, 10, and 6‐MV flattening filter‐free (FFF) photon beams. Methods: We assessed surface dose, PDD20, 10, TPR20, 10, PDD curves, inline and crossline profiles, transmission factor, and output factor with and without the IQM. PDD, transmission factor, and output factor were measured for square fields of 3, 5, 10, 15, 20, 25, and 30 cm and profiles were performed for square fields of 3, 5, 10, 20, and 30 cm at 5‐, 10‐, and 30‐cm depth. Results: The differences in surface dose of all energies for square fields of 3, 5, 10, 15, 20, and 25 cm were within 3.7% whereas for a square field of 30 cm, they were 4.6%, 6.8%, 6.7%, and 8.7% for 4‐MV, 6‐MV, 6‐MV‐FFF, and 10‐MV, respectively. Differences in PDD20, 10, TPR20, 10, PDD, profiles, and output factors were within ±1%. Local and global gamma values (2%/2 mm) were below 1 for PDD beyond d max and inline/crossline profiles in the central beam region, respectively. The gamma passing rates (10% threshold) for PDD curves and profiles were above 95% at 2%/2 mm. The transmission factors for 4‐MV, 6‐MV, 6‐MV‐FFF, and 10‐MV for field sizes from 3 × 3 to 30 × 30 cm 2 were 0.926–0.933, 0.937–0.941, 0.937–0.939, and 0.949–0.953, respectively. Conclusions: The influence of the IQM on the beam quality (in particular 4‐MV X‐ray has not verified before) was tested and introduced a slight beam perturbation at the surface andAbstract: Purpose: To investigate the effect of an integral quality monitor (IQM; iRT Systems GmbH, Koblenz, Germany) on 4, 6, 10, and 6‐MV flattening filter‐free (FFF) photon beams. Methods: We assessed surface dose, PDD20, 10, TPR20, 10, PDD curves, inline and crossline profiles, transmission factor, and output factor with and without the IQM. PDD, transmission factor, and output factor were measured for square fields of 3, 5, 10, 15, 20, 25, and 30 cm and profiles were performed for square fields of 3, 5, 10, 20, and 30 cm at 5‐, 10‐, and 30‐cm depth. Results: The differences in surface dose of all energies for square fields of 3, 5, 10, 15, 20, and 25 cm were within 3.7% whereas for a square field of 30 cm, they were 4.6%, 6.8%, 6.7%, and 8.7% for 4‐MV, 6‐MV, 6‐MV‐FFF, and 10‐MV, respectively. Differences in PDD20, 10, TPR20, 10, PDD, profiles, and output factors were within ±1%. Local and global gamma values (2%/2 mm) were below 1 for PDD beyond d max and inline/crossline profiles in the central beam region, respectively. The gamma passing rates (10% threshold) for PDD curves and profiles were above 95% at 2%/2 mm. The transmission factors for 4‐MV, 6‐MV, 6‐MV‐FFF, and 10‐MV for field sizes from 3 × 3 to 30 × 30 cm 2 were 0.926–0.933, 0.937–0.941, 0.937–0.939, and 0.949–0.953, respectively. Conclusions: The influence of the IQM on the beam quality (in particular 4‐MV X‐ray has not verified before) was tested and introduced a slight beam perturbation at the surface and build‐up region and the edge of the crossline/inline profiles. To use IQM in pre‐ and intra‐treatment quality assurance, a tray factor should be put into treatment planning systems for the dose calculation for the 4‐, 6‐, 10‐, and 6‐MV flattening filter‐free photon beams to compensate the beam attenuation of the IQM detector. … (more)
- Is Part Of:
- Journal of applied clinical medical physics. Volume 22:Issue 1(2021)
- Journal:
- Journal of applied clinical medical physics
- Issue:
- Volume 22:Issue 1(2021)
- Issue Display:
- Volume 22, Issue 1 (2021)
- Year:
- 2021
- Volume:
- 22
- Issue:
- 1
- Issue Sort Value:
- 2021-0022-0001-0000
- Page Start:
- 76
- Page End:
- 91
- Publication Date:
- 2020-12-03
- Subjects:
- beam quality -- integral quality monitor (IQM) -- photon energy
Medical physics -- Periodicals
Clinical medicine -- Periodicals
Health Physics
Clinical Medicine
Electronic journals
Periodicals
Periodicals
Fulltext
Internet Resources
610.153 - Journal URLs:
- http://aapm.onlinelibrary.wiley.com/hub/journal/10.1002/(ISSN)1526-9914/ ↗
http://bibpurl.oclc.org/web/7294 ↗
http://www.jacmp.org/ ↗
http://onlinelibrary.wiley.com/ ↗ - DOI:
- 10.1002/acm2.13106 ↗
- Languages:
- English
- ISSNs:
- 1526-9914
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 15783.xml