Percutaneous pulmonary artery debanding. Issue 3 (March 2021)
- Record Type:
- Journal Article
- Title:
- Percutaneous pulmonary artery debanding. Issue 3 (March 2021)
- Main Title:
- Percutaneous pulmonary artery debanding
- Authors:
- Tomita, Hideshi
Fujii, Takanari
Kise, Hiroaki
Oyama, Nobuo
Shimizu, Takeshi
Nagaoka, Kota
Tarui, Suguru
Miyahara, Yoshinori
Ishino, Kozo - Abstract:
- Highlights: Palliative or total debanding was performed using balloon and/or stenting. In palliative debanding, the balloon diameter was 35–50% to the band circumference. In total debanding, the balloon diameter was >50% to the band circumference. The balloon diameter should be less than 1.5 times to the reference vessel diameter. Transcatheter palliative or total debanding is generally feasible and effective. Abstract: Background: There is a paucity of data on palliative or total percutaneous pulmonary artery debanding (p-debanding), particularly with use of a stent. Methods: Twelve p-debandings in eight patients were included in this study. Age at pulmonary artery banding (PAB) ranged from 3 days to 1 year (median, 13 days), while p-debanding was performed at 2–157 (7) months. The body weight at the p-debanding ranged from 3.2 to 22.2 (7.3) kg. We chose the balloon diameter of 30–50% to the circumference of the band for palliative, and larger than 50% for total p-debanding, respectively. In either way, the balloon diameter did not exceed 1.5 times the reference vessel diameter. Stent was implanted for palliative p-debanding in 2 patients. Results: 1. The circumference of the band ranged from 16 to 23 (20) mm, while the balloon diameter ranged from 20–60 (40)% to that, where larger than 50% was used for 2 procedures intended total p-debanding. 2. PAB diameter increased from 2.5–4.7 (3.0) mm to 2.8–9.5 (4.5) mm ( p < 0.01), however, there was no significant change in theHighlights: Palliative or total debanding was performed using balloon and/or stenting. In palliative debanding, the balloon diameter was 35–50% to the band circumference. In total debanding, the balloon diameter was >50% to the band circumference. The balloon diameter should be less than 1.5 times to the reference vessel diameter. Transcatheter palliative or total debanding is generally feasible and effective. Abstract: Background: There is a paucity of data on palliative or total percutaneous pulmonary artery debanding (p-debanding), particularly with use of a stent. Methods: Twelve p-debandings in eight patients were included in this study. Age at pulmonary artery banding (PAB) ranged from 3 days to 1 year (median, 13 days), while p-debanding was performed at 2–157 (7) months. The body weight at the p-debanding ranged from 3.2 to 22.2 (7.3) kg. We chose the balloon diameter of 30–50% to the circumference of the band for palliative, and larger than 50% for total p-debanding, respectively. In either way, the balloon diameter did not exceed 1.5 times the reference vessel diameter. Stent was implanted for palliative p-debanding in 2 patients. Results: 1. The circumference of the band ranged from 16 to 23 (20) mm, while the balloon diameter ranged from 20–60 (40)% to that, where larger than 50% was used for 2 procedures intended total p-debanding. 2. PAB diameter increased from 2.5–4.7 (3.0) mm to 2.8–9.5 (4.5) mm ( p < 0.01), however, there was no significant change in the diameter in 2 procedures. In one patient, p-debanding was the definitive treatment associated with spontaneous near closure of muscular ventricular septal defect, in another patient of congenitally corrected transposition of the great arteries, severely depressed left ventricular ejection fraction was recovered following p-debanding. 3. Arterial oxygen saturation (SaO2 ) increased from 64–97 (80)% to 66–95 (90)% ( p < 0.01), while in 10 procedures of 6 patients where the indication of p-debanding was hypoxia, SaO2 increased in 8 procedures. There was no significant pulmonary hypertension following p-debanding. Conclusion: Palliative or total p-debanding using balloon and/or stenting is generally feasible and effective. A balloon diameter 35–50% to the band circumference in palliative, and more than 50% in total p-debanding, while in either way less than 1.5 times the reference vessel diameter, is safe. … (more)
- Is Part Of:
- Journal of cardiology. Volume 77:Issue 3(2021)
- Journal:
- Journal of cardiology
- Issue:
- Volume 77:Issue 3(2021)
- Issue Display:
- Volume 77, Issue 3 (2021)
- Year:
- 2021
- Volume:
- 77
- Issue:
- 3
- Issue Sort Value:
- 2021-0077-0003-0000
- Page Start:
- 307
- Page End:
- 312
- Publication Date:
- 2021-03
- Subjects:
- Pulmonary artery debanding -- Balloon dilatation -- Stent -- Congenital heart disease
Cardiology -- Periodicals
616.12 - Journal URLs:
- http://www.clinicalkey.com/dura/browse/journalIssue/09145087 ↗
http://www.sciencedirect.com/science/journal/09145087 ↗
http://www.elsevier.com/journals ↗ - DOI:
- 10.1016/j.jjcc.2020.10.021 ↗
- Languages:
- English
- ISSNs:
- 0914-5087
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 4954.864200
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 15854.xml