Analytical performance evaluation of five RT‐PCR kits for severe acute respiratory syndrome coronavirus 2. Issue 1 (27th October 2020)
- Record Type:
- Journal Article
- Title:
- Analytical performance evaluation of five RT‐PCR kits for severe acute respiratory syndrome coronavirus 2. Issue 1 (27th October 2020)
- Main Title:
- Analytical performance evaluation of five RT‐PCR kits for severe acute respiratory syndrome coronavirus 2
- Authors:
- Wang, Mengmeng
Chen, Dayang
Wu, Wei
Tang, Huamei
Kan, Lijuan
Zong, Zengyan
Dou, Xiaowen
Ji, Xiang
Xiong, Dan
Zhang, Xiuming - Abstract:
- Abstract: Background: We aimed to evaluate the analytical performance of five commercial RT‐PCR kits (Genekey, Daan, BioGerm, Liferiver, and Yaneng) commonly used in China, since such comparison data are lacking. Methods: A total of 20 COVID‐19 confirmed patients and 30 negative nasopharyngeal swab specimens were analyzed by five kits. The detection ability of five RT‐PCR kits was evaluated with 5 concentration gradients diluted by a single positive sample. The limit of detection was evaluated by N gene fragment solid standard. Two positive clinical specimens were used to evaluate the repeatability and imprecision. Finally, we used six human coronaviruses plasmid and four respiratory pathogens plasmid to check for cross‐reactivity. Results: The positive detection rate was 100% for Genekey, Daan, and BioGerm, and 90% for Liferiver and Yaneng in 20 clinical SARS‐CoV‐2 infection. The coincidence rate of five kits in 10 negative samples was 100%. The detection rate of target genes for Daan, BioGerm, Liferiver, and Yaneng was 100% from Level 1 to Level 3. In Level 4, only Daan detection rate was 100%. In Level 5, five kits presented poor positive rate. The limit of detection declared by each manufacturer was verified. The repeatability for target genes was less than 5% and so did the total imprecision. There is no cross‐reactivity of five kits with six human coronaviruses and four respiratory pathogens for ORF1ab and N gene. Conclusions: Five RT‐PCR kits assessed in this studyAbstract: Background: We aimed to evaluate the analytical performance of five commercial RT‐PCR kits (Genekey, Daan, BioGerm, Liferiver, and Yaneng) commonly used in China, since such comparison data are lacking. Methods: A total of 20 COVID‐19 confirmed patients and 30 negative nasopharyngeal swab specimens were analyzed by five kits. The detection ability of five RT‐PCR kits was evaluated with 5 concentration gradients diluted by a single positive sample. The limit of detection was evaluated by N gene fragment solid standard. Two positive clinical specimens were used to evaluate the repeatability and imprecision. Finally, we used six human coronaviruses plasmid and four respiratory pathogens plasmid to check for cross‐reactivity. Results: The positive detection rate was 100% for Genekey, Daan, and BioGerm, and 90% for Liferiver and Yaneng in 20 clinical SARS‐CoV‐2 infection. The coincidence rate of five kits in 10 negative samples was 100%. The detection rate of target genes for Daan, BioGerm, Liferiver, and Yaneng was 100% from Level 1 to Level 3. In Level 4, only Daan detection rate was 100%. In Level 5, five kits presented poor positive rate. The limit of detection declared by each manufacturer was verified. The repeatability for target genes was less than 5% and so did the total imprecision. There is no cross‐reactivity of five kits with six human coronaviruses and four respiratory pathogens for ORF1ab and N gene. Conclusions: Five RT‐PCR kits assessed in this study showed acceptable analytical performance characteristics and are useful tools for the routine diagnosis of SARS‐CoV‐2. Abstract : In the work, we presented the analytical performance evaluations of five RT‐PCR kits using nasopharyngeal swabs samples from patients with confirmed SARS‐CoV‐2 infection, and negative nasopharyngeal swabs samples (Figure 1). A total of 20 COVID‐19 confirmed patients and 30 negative nasopharyngeal swab specimens were analyzed by five kits. The detection ability of five RT‐PCR kits was evaluated with 5 concentration gradients diluted by a single positive sample. The limit of detection was evaluated by N gene fragment solid standard. Two positive clinical specimens were used to evaluate the repeatability and imprecision. Finally, we used six human coronaviruses plasmid and four respiratory pathogens plasmid to check for cross‐reactivity … (more)
- Is Part Of:
- Journal of clinical laboratory analysis. Volume 35:Issue 1(2021)
- Journal:
- Journal of clinical laboratory analysis
- Issue:
- Volume 35:Issue 1(2021)
- Issue Display:
- Volume 35, Issue 1 (2021)
- Year:
- 2021
- Volume:
- 35
- Issue:
- 1
- Issue Sort Value:
- 2021-0035-0001-0000
- Page Start:
- n/a
- Page End:
- n/a
- Publication Date:
- 2020-10-27
- Subjects:
- analytical performance -- detection ability -- limit of detection -- repeatability and imprecision -- RT‐PCR -- SARS‐CoV‐2
Diagnosis, Laboratory -- Periodicals
Medical laboratory technology -- Periodicals
616 - Journal URLs:
- http://onlinelibrary.wiley.com/ ↗
- DOI:
- 10.1002/jcla.23643 ↗
- Languages:
- English
- ISSNs:
- 0887-8013
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 4958.520000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 15566.xml