The effect of haemolysis on the direct and indirect ion selective electrode measurement of sodium. (May 2021)
- Record Type:
- Journal Article
- Title:
- The effect of haemolysis on the direct and indirect ion selective electrode measurement of sodium. (May 2021)
- Main Title:
- The effect of haemolysis on the direct and indirect ion selective electrode measurement of sodium
- Authors:
- Kalaria, Tejas
Gill, Henry
Harris, Steve
Ford, Clare
Gama, Rousseau - Abstract:
- Background: We compared the effect of haemolysis in sodium measurement using indirect and direct ion-selective electrodes to test the hypothesis that haemolytic effect on sodium would be greater with indirect ion-selective electrode due to electrolyte exclusion effect from released intracellular proteins. Methods: Plasma lithium heparin samples ( n = 36) from four volunteers were prepared to give a range of haemolytic indices (H-indices). Samples were analysed for sodium by indirect ion-selective electrode, H-index and total protein on an Abbott Architect c16000 and sodium by direct ion-selective electrode on a Siemens RAPIDPoint 500. Percentage changes in sodium in paired direct and indirect ion-selective electrode values were compared. Results: Abbott H-index, which represents haemoglobin concentration in g/L, correlated with percentage negative change in sodium by direct ion-selective electrode (ρ 0.995, P < 0.001) and indirect ion-selective electrode (ρ 0.991, P < 0.001). Percentage negative change was less when sodium was measured by direct ion-selective electrode compared to indirect ion-selective electrode (Wilcoxon signed-rank Z = 3.46, P = 0.01). The difference in percentage change in sodium between direct ion-selective electrode and indirect ion-selective electrode correlated with total protein (ρ 0.751, P < 0.001). The negative bias in sodium results exceeded the reference change value of 2.2% at an H-index of 8.31 for indirect ion-selective electrode andBackground: We compared the effect of haemolysis in sodium measurement using indirect and direct ion-selective electrodes to test the hypothesis that haemolytic effect on sodium would be greater with indirect ion-selective electrode due to electrolyte exclusion effect from released intracellular proteins. Methods: Plasma lithium heparin samples ( n = 36) from four volunteers were prepared to give a range of haemolytic indices (H-indices). Samples were analysed for sodium by indirect ion-selective electrode, H-index and total protein on an Abbott Architect c16000 and sodium by direct ion-selective electrode on a Siemens RAPIDPoint 500. Percentage changes in sodium in paired direct and indirect ion-selective electrode values were compared. Results: Abbott H-index, which represents haemoglobin concentration in g/L, correlated with percentage negative change in sodium by direct ion-selective electrode (ρ 0.995, P < 0.001) and indirect ion-selective electrode (ρ 0.991, P < 0.001). Percentage negative change was less when sodium was measured by direct ion-selective electrode compared to indirect ion-selective electrode (Wilcoxon signed-rank Z = 3.46, P = 0.01). The difference in percentage change in sodium between direct ion-selective electrode and indirect ion-selective electrode correlated with total protein (ρ 0.751, P < 0.001). The negative bias in sodium results exceeded the reference change value of 2.2% at an H-index of 8.31 for indirect ion-selective electrode and 9.26 for direct ion-selective electrode. Conclusion: Haemolysis causes negative influence with sodium measured by both indirect and direct ion-selective electrode due to a dilutional hyponatremia. The additional interference in indirect ion-selective electrode is due to the electrolyte exclusion effect but this is unlikely to be clinically significant as it is small in magnitude. … (more)
- Is Part Of:
- Annals of clinical biochemistry. Volume 58:Number 3(2021)
- Journal:
- Annals of clinical biochemistry
- Issue:
- Volume 58:Number 3(2021)
- Issue Display:
- Volume 58, Issue 3 (2021)
- Year:
- 2021
- Volume:
- 58
- Issue:
- 3
- Issue Sort Value:
- 2021-0058-0003-0000
- Page Start:
- 190
- Page End:
- 195
- Publication Date:
- 2021-05
- Subjects:
- Sodium -- haemolysis -- direct ion-selective electrode -- indirect ion-selective electrode
Clinical chemistry -- Periodicals
Clinical biochemistry -- Periodicals
616.075 - Journal URLs:
- http://web.ebscohost.com/ehost/detail?sid=810a7788-77dd-439f-9630-ad7f5b199fd3%40sessionmgr4&vid=1&hid=14&bdata=JnNpdGU9ZWhvc3QtbGl2ZSZzY29wZT1zaXRl#db=mnh&jid=0324055 ↗
http://acb.rsmjournals.com ↗
http://www.usc.edu/hsc/nml/e-resources/info/annclib.html ↗
http://www.uk.sagepub.com/home.nav ↗
http://www.ingentaconnect.com/content/rsm/acb ↗
http://firstsearch.oclc.org ↗ - DOI:
- 10.1177/0004563220987593 ↗
- Languages:
- English
- ISSNs:
- 0004-5632
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 15493.xml