Frameless Stereotactic Brain Biopsies: Comparison of Minimally Invasive Robot-Guided and Manual Arm-Based Technique. Issue 3 (15th May 2020)
- Record Type:
- Journal Article
- Title:
- Frameless Stereotactic Brain Biopsies: Comparison of Minimally Invasive Robot-Guided and Manual Arm-Based Technique. Issue 3 (15th May 2020)
- Main Title:
- Frameless Stereotactic Brain Biopsies: Comparison of Minimally Invasive Robot-Guided and Manual Arm-Based Technique
- Authors:
- Minchev, Georgi
Kronreif, Gernot
Ptacek, Wolfgang
Kettenbach, Joachim
Micko, Alexander
Wurzer, Ayguel
Maschke, Svenja
Wolfsberger, Stefan - Abstract:
- Abstract: BACKGROUND: Most brain biopsies are still performed with the aid of a navigation-guided mechanical arm. Due to the manual trajectory alignment without rigid skull contact, frameless aiming devices are prone to considerably lower accuracy. OBJECTIVE: To compare a novel minimally invasive robot-guided biopsy technique with rigid skull fixation to a standard frameless manual arm biopsy procedure. METHODS: Accuracy, procedural duration, diagnostic yield, complication rate, and cosmetic result were retrospectively assessed in 40 consecutive cases of frameless stereotactic biopsies and compared between a minimally invasive robotic technique using the iSYS1 guidance device (iSYS Medizintechnik GmbH) (robot-guided group [ROB], n = 20) and a manual arm-based technique (group MAN, n = 20). RESULTS: Application of the robotic technique resulted in significantly higher accuracy at entry point (group ROB median 1.5 mm [0.4-3.2 mm] vs manual arm-based group (MAN) 2.2 mm [0.2-5.2 mm], P = .019) and at target point (group ROB 1.5 mm [0.4-2.8 mm] vs group MAN 2.8 mm [1.4-4.9 mm], P = .001), without increasing incision to suture time (group ROB 30.0 min [20-45 min vs group MAN 32.5 min [range 20-60 min], P = .09) and significantly shorter skin incision length (group ROB 16.3 mm [12.7-23.4 mm] vs group MAN 24.2 mm [18.0-37.0 mm], P = .008). CONCLUSION: According to our data, the proposed technique of minimally invasive robot-guided brain biopsies can improve accuracy withoutAbstract: BACKGROUND: Most brain biopsies are still performed with the aid of a navigation-guided mechanical arm. Due to the manual trajectory alignment without rigid skull contact, frameless aiming devices are prone to considerably lower accuracy. OBJECTIVE: To compare a novel minimally invasive robot-guided biopsy technique with rigid skull fixation to a standard frameless manual arm biopsy procedure. METHODS: Accuracy, procedural duration, diagnostic yield, complication rate, and cosmetic result were retrospectively assessed in 40 consecutive cases of frameless stereotactic biopsies and compared between a minimally invasive robotic technique using the iSYS1 guidance device (iSYS Medizintechnik GmbH) (robot-guided group [ROB], n = 20) and a manual arm-based technique (group MAN, n = 20). RESULTS: Application of the robotic technique resulted in significantly higher accuracy at entry point (group ROB median 1.5 mm [0.4-3.2 mm] vs manual arm-based group (MAN) 2.2 mm [0.2-5.2 mm], P = .019) and at target point (group ROB 1.5 mm [0.4-2.8 mm] vs group MAN 2.8 mm [1.4-4.9 mm], P = .001), without increasing incision to suture time (group ROB 30.0 min [20-45 min vs group MAN 32.5 min [range 20-60 min], P = .09) and significantly shorter skin incision length (group ROB 16.3 mm [12.7-23.4 mm] vs group MAN 24.2 mm [18.0-37.0 mm], P = .008). CONCLUSION: According to our data, the proposed technique of minimally invasive robot-guided brain biopsies can improve accuracy without increasing operating time while being equally safe and effective compared to a standard frameless arm-based manual biopsy technique. … (more)
- Is Part Of:
- Operative neurosurgery. Volume 19:Issue 3(2020)
- Journal:
- Operative neurosurgery
- Issue:
- Volume 19:Issue 3(2020)
- Issue Display:
- Volume 19, Issue 3 (2020)
- Year:
- 2020
- Volume:
- 19
- Issue:
- 3
- Issue Sort Value:
- 2020-0019-0003-0000
- Page Start:
- 292
- Page End:
- 301
- Publication Date:
- 2020-05-15
- Subjects:
- Accuracy -- Biopsy -- Robotic device -- Stereotaxy
Nervous system -- Surgery -- Periodicals
617.480590 - Journal URLs:
- https://academic.oup.com/ons/issue ↗
http://journals.lww.com/onsonline/pages/default.aspx ↗
http://journals.lww.com/pages/default.aspx ↗ - DOI:
- 10.1093/ons/opaa123 ↗
- Languages:
- English
- ISSNs:
- 2332-4252
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 6269.380200
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 15467.xml