A Comparison of International Normalized Ratio Results by Point-of-Care Device and Clinical Laboratory Analyzers in a Vascular Surgery Department. Issue 4 (December 2020)
- Record Type:
- Journal Article
- Title:
- A Comparison of International Normalized Ratio Results by Point-of-Care Device and Clinical Laboratory Analyzers in a Vascular Surgery Department. Issue 4 (December 2020)
- Main Title:
- A Comparison of International Normalized Ratio Results by Point-of-Care Device and Clinical Laboratory Analyzers in a Vascular Surgery Department
- Authors:
- Alserr, Ayman H. K.
Menshawey, Rahma
Kotb, Ahmed
Hussein, Nourhan
Kotp, Noha
Ashraf-Taha, Mahmoud
Anwar, Nouran
Abdalla, Asmaa
Abdullah, Menna
Ela, Shaimaa Abol
Khairy, Hussein - Abstract:
- Abstract : Purpose: This prospective comparative study aims to evaluate the accuracy and validity of CoaguChek XS point-of-care system (Roche Diagnostics GmbH, Mannheim, Germany), compared with the standard central laboratory international normalized ratio (INR) results, in a vascular surgery ward. Methods: Data based on 54 INR pairs, measured by the central laboratory analyzer and CoaguChek XS, were compared using the intraclass correlation coefficient. Other outcomes included the time and cost of the service from the patient's perspectives. Correlation was assessed using Pearson correlation test. Bland-Altman analysis was done to determine mean difference and 95% limits of agreement. Cohen κ test was done to assess agreement between the 2 methods. Results: The mean INR values for CoaguChek XS and standard laboratory (Sysmex Ca-1500 system; Siemens Healthcare Diagnostics, Erlangen, Germany) were 1.21 ± 0.28 and 1.24 ± 0.43, respectively. The Pearson correlation coefficient in all 108 measurements was ( r ) = 0.591 (95% confidence interval, 0.38–0.74; P < 0.0001), indicating a moderately positive correlation. Overall intraclass correlation coefficient was 0.803 (95% confidence interval, 0.324–0.707) indicating good reliability. Bland-Altman analysis showed a bias of 0.035 and a 95% limits of agreement ranging from −0.63 to 0.70. There was substantial agreement between the 2 methods ( κ = 0.694; SE, 0.108; 95% confidence interval, 0.481–0.907). CoaguChek XS results wereAbstract : Purpose: This prospective comparative study aims to evaluate the accuracy and validity of CoaguChek XS point-of-care system (Roche Diagnostics GmbH, Mannheim, Germany), compared with the standard central laboratory international normalized ratio (INR) results, in a vascular surgery ward. Methods: Data based on 54 INR pairs, measured by the central laboratory analyzer and CoaguChek XS, were compared using the intraclass correlation coefficient. Other outcomes included the time and cost of the service from the patient's perspectives. Correlation was assessed using Pearson correlation test. Bland-Altman analysis was done to determine mean difference and 95% limits of agreement. Cohen κ test was done to assess agreement between the 2 methods. Results: The mean INR values for CoaguChek XS and standard laboratory (Sysmex Ca-1500 system; Siemens Healthcare Diagnostics, Erlangen, Germany) were 1.21 ± 0.28 and 1.24 ± 0.43, respectively. The Pearson correlation coefficient in all 108 measurements was ( r ) = 0.591 (95% confidence interval, 0.38–0.74; P < 0.0001), indicating a moderately positive correlation. Overall intraclass correlation coefficient was 0.803 (95% confidence interval, 0.324–0.707) indicating good reliability. Bland-Altman analysis showed a bias of 0.035 and a 95% limits of agreement ranging from −0.63 to 0.70. There was substantial agreement between the 2 methods ( κ = 0.694; SE, 0.108; 95% confidence interval, 0.481–0.907). CoaguChek XS results were obtained, on average, 213 minutes faster than the standard laboratory. Conclusions: CoaguChek XS appears to have positive correlation, substantial agreement, and good reliability when compared with the standard laboratory. Thus, it is recommended to be used in hospital settings. … (more)
- Is Part Of:
- Point of care. Volume 19:Issue 4(2020)
- Journal:
- Point of care
- Issue:
- Volume 19:Issue 4(2020)
- Issue Display:
- Volume 19, Issue 4 (2020)
- Year:
- 2020
- Volume:
- 19
- Issue:
- 4
- Issue Sort Value:
- 2020-0019-0004-0000
- Page Start:
- Page End:
- Publication Date:
- 2020-12
- Subjects:
- point-of-care device -- CoaguChek XS -- international normalized ratio -- laboratory analyzers
Point-of-care testing -- Periodicals
Point-of-Care Systems -- standards -- Periodicals
616.075 - Journal URLs:
- http://www.poctjournal.com ↗
http://ovidsp.ovid.com/ovidweb.cgi?T=JS&NEWS=n&CSC=Y&PAGE=toc&D=yrovft&AN=00134384-000000000-00000 ↗
http://journals.lww.com ↗ - DOI:
- 10.1097/POC.0000000000000213 ↗
- Languages:
- English
- ISSNs:
- 1533-029X
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 6541.842900
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 15463.xml