A New Medical Device to Provide Independent Ventilation to Two Subjects Using a Single Ventilator: Evaluation in Lung-Healthy Pigs. (30th December 2020)
- Record Type:
- Journal Article
- Title:
- A New Medical Device to Provide Independent Ventilation to Two Subjects Using a Single Ventilator: Evaluation in Lung-Healthy Pigs. (30th December 2020)
- Main Title:
- A New Medical Device to Provide Independent Ventilation to Two Subjects Using a Single Ventilator: Evaluation in Lung-Healthy Pigs
- Authors:
- Lugones, Ignacio
Orofino Giambastiani, Roberto
Robledo, Oscar
Marcos, Martín
Mouly, Javier
Gallo, Agustín
Laulhé, Verónica
Biancolini, María Fernanda - Other Names:
- Goudra Basavana B. Academic Editor.
- Abstract:
- Abstract : Background . The global crisis situation caused by SARS-CoV-2 has created an explosive demand for ventilators, which cannot be met even in developed countries. Designing a simple and inexpensive device with the ability to increase the number of patients that can be connected to existing ventilators would have a major impact on the number of lives that could be saved. We conducted a study to determine whether two pigs with significant differences in size and weight could be ventilated simultaneously using a single ventilator connected to a new medical device called DuplicARⓇ. Methods . Six pigs (median weight 12 kg, range 9–25 kg) were connected in pairs to a single ventilator using the new device for 6 hours. Both the ventilator and the device were manipulated throughout the experiment according to the needs of each animal. Tidal volume and positive end-expiratory pressure were individually controlled with the device. Primary and secondary outcome variables were defined to assess ventilation and hemodynamics in all animals throughout the experiment. Results . Median difference in weight between the animals of each pair was 67% (range: 11–108). All animals could be successfully oxygenated and ventilated for 6 hours through manipulation of the ventilator and the DuplicARⓇ device, despite significant discrepancies in body size and weight. Mean PaCO2 in arterial blood was 42.1 ± 4.4 mmHg, mean PaO2 was 162.8 ± 46.8 mmHg, and mean oxygen saturation was 98 ± 1.3%.Abstract : Background . The global crisis situation caused by SARS-CoV-2 has created an explosive demand for ventilators, which cannot be met even in developed countries. Designing a simple and inexpensive device with the ability to increase the number of patients that can be connected to existing ventilators would have a major impact on the number of lives that could be saved. We conducted a study to determine whether two pigs with significant differences in size and weight could be ventilated simultaneously using a single ventilator connected to a new medical device called DuplicARⓇ. Methods . Six pigs (median weight 12 kg, range 9–25 kg) were connected in pairs to a single ventilator using the new device for 6 hours. Both the ventilator and the device were manipulated throughout the experiment according to the needs of each animal. Tidal volume and positive end-expiratory pressure were individually controlled with the device. Primary and secondary outcome variables were defined to assess ventilation and hemodynamics in all animals throughout the experiment. Results . Median difference in weight between the animals of each pair was 67% (range: 11–108). All animals could be successfully oxygenated and ventilated for 6 hours through manipulation of the ventilator and the DuplicARⓇ device, despite significant discrepancies in body size and weight. Mean PaCO2 in arterial blood was 42.1 ± 4.4 mmHg, mean PaO2 was 162.8 ± 46.8 mmHg, and mean oxygen saturation was 98 ± 1.3%. End-tidal CO2 values showed no statistically significant difference among subjects of each pair. Mean difference in arterial PaCO2 measured at the same time in both animals of each pair was 4.8 ± 3 mmHg, reflecting the ability of the device to ventilate each animal according to its particular requirements. Independent management of PEEP was achieved by manipulation of the device controllers. Conclusion . It is possible to ventilate two lung-healthy animals with a single ventilator according to each one's needs through manipulation of both the ventilator and the DuplicARⓇ device. This gives this device the potential to expand local ventilators surge capacity during disasters or pandemics until emergency supplies can be delivered from central stockpiles. … (more)
- Is Part Of:
- Anesthesiology research and practice. Volume 2020(2020)
- Journal:
- Anesthesiology research and practice
- Issue:
- Volume 2020(2020)
- Issue Display:
- Volume 2020, Issue 2020 (2020)
- Year:
- 2020
- Volume:
- 2020
- Issue:
- 2020
- Issue Sort Value:
- 2020-2020-2020-0000
- Page Start:
- Page End:
- Publication Date:
- 2020-12-30
- Subjects:
- Anesthesiology -- Periodicals
617.9605 - Journal URLs:
- https://www.hindawi.com/journals/arp/ ↗
- DOI:
- 10.1155/2020/8866806 ↗
- Languages:
- English
- ISSNs:
- 1687-6962
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library HMNTS - ELD Digital store
- Ingest File:
- 15415.xml