100% Nitrous Oxide in the Oxygen Line: How Could This Happen in a Modern Anesthesia Machine?. Issue 3 (November 2020)
- Record Type:
- Journal Article
- Title:
- 100% Nitrous Oxide in the Oxygen Line: How Could This Happen in a Modern Anesthesia Machine?. Issue 3 (November 2020)
- Main Title:
- 100% Nitrous Oxide in the Oxygen Line
- Authors:
- Joo, Sarah S.
Bechtold, Hannah
Jaffe, Richard A.
Brock-Utne, John G. - Abstract:
- Abstract : The inadvertent crossover between O2 and N2 O pipelines has become extremely rare in practice. We describe a case where it was possible to ventilate with 100% N2 O instead of the intended 100% O2 on a modern anesthesia delivery system (Dräger Apollo; Drägerwerk AG & Co KgaA, Lübeck, Germany). This was the result of the incorrect assembly of diameter index safety system (DISS) components during preventative maintenance that defeated the DISS failsafe system. To make incorrect assembly easier to avoid, DISS component labeling could be more prominent and color-coded, or the internal construction of the gas manifold could incorporate DISS.
- Is Part Of:
- A & A practice. Volume 14:Issue 3(2020)
- Journal:
- A & A practice
- Issue:
- Volume 14:Issue 3(2020)
- Issue Display:
- Volume 14, Issue 3 (2020)
- Year:
- 2020
- Volume:
- 14
- Issue:
- 3
- Issue Sort Value:
- 2020-0014-0003-0000
- Page Start:
- Page End:
- Publication Date:
- 2020-11
- Subjects:
- Analgesia -- Periodicals
Analgesia -- Case studies -- Periodicals
Anesthesiology -- Periodicals
Anesthesiology -- Case studies -- Periodicals
617.9605 - Journal URLs:
- http://journals.lww.com/pages/default.aspx ↗
- DOI:
- 10.1213/XAA.0000000000001337 ↗
- Languages:
- English
- ISSNs:
- 2575-3126
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 0537.002900
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 15355.xml