Monte Carlo dosimetric characterization of a new high dose rate 169 Yb brachytherapy source and independent verification using a multipoint plastic scintillator detector. Issue 9 (19th July 2020)
- Record Type:
- Journal Article
- Title:
- Monte Carlo dosimetric characterization of a new high dose rate 169 Yb brachytherapy source and independent verification using a multipoint plastic scintillator detector. Issue 9 (19th July 2020)
- Main Title:
- Monte Carlo dosimetric characterization of a new high dose rate 169 Yb brachytherapy source and independent verification using a multipoint plastic scintillator detector
- Authors:
- Famulari, Gabriel
Linares Rosales, Haydee M.
Dupere, Justine
Medich, David C.
Beaulieu, Luc
Enger, Shirin A. - Abstract:
- Abstract : Purpose: A prototype 169 Yb source was developed in combination with a dynamic rotating platinum shield system (AIM‐Brachy) to deliver intensity modulated brachytherapy (IMBT). The purpose of this study was to evaluate the dosimetric characteristics of the bare/shielded 169 Yb source using Monte Carlo (MC) simulations and perform an independent dose verification using a dosimetry platform based on a multipoint plastic scintillator detector (mPSD). Methods: The TG‐43U1 dosimetric parameters were calculated for the source model using RapidBrachyMCTPS. Real‐time dose rate measurements were performed in a water tank for both the bare/shielded source using a custom remote afterloader. For each dwell position, the dose rate was independently measured by the three scintillators (BCF‐10, BCF‐12, and BCF‐60). For the bare source, dose rate was measured at distances up to 3 cm away from the source over a range of 7 cm along the catheter. For the shielded source, measurements were performed with the mPSD placed at 1 cm from the source at four different azimuthal angles ( 0 ∘, 9 0 ∘, 18 0 ∘, and 27 0 ∘ ). Results: The dosimetric parameters were tabulated for the source model. For the bare source, differences between measured and calculated along‐away dose rates were generally below 5–10%. Along the transverse axis, deviations were, on average (range), 3.3% (0.6–6.2%) for BCF‐10, 1.7% (0.9–2.9%) for BCF‐12, and 2.2% (0.3–4.4%) for BCF‐60. The maximum dose rate reduction due toAbstract : Purpose: A prototype 169 Yb source was developed in combination with a dynamic rotating platinum shield system (AIM‐Brachy) to deliver intensity modulated brachytherapy (IMBT). The purpose of this study was to evaluate the dosimetric characteristics of the bare/shielded 169 Yb source using Monte Carlo (MC) simulations and perform an independent dose verification using a dosimetry platform based on a multipoint plastic scintillator detector (mPSD). Methods: The TG‐43U1 dosimetric parameters were calculated for the source model using RapidBrachyMCTPS. Real‐time dose rate measurements were performed in a water tank for both the bare/shielded source using a custom remote afterloader. For each dwell position, the dose rate was independently measured by the three scintillators (BCF‐10, BCF‐12, and BCF‐60). For the bare source, dose rate was measured at distances up to 3 cm away from the source over a range of 7 cm along the catheter. For the shielded source, measurements were performed with the mPSD placed at 1 cm from the source at four different azimuthal angles ( 0 ∘, 9 0 ∘, 18 0 ∘, and 27 0 ∘ ). Results: The dosimetric parameters were tabulated for the source model. For the bare source, differences between measured and calculated along‐away dose rates were generally below 5–10%. Along the transverse axis, deviations were, on average (range), 3.3% (0.6–6.2%) for BCF‐10, 1.7% (0.9–2.9%) for BCF‐12, and 2.2% (0.3–4.4%) for BCF‐60. The maximum dose rate reduction due to shielding at a radial distance of 1 cm was 88.8 ± 1.2%, compared to 83.5 ± 0.5% as calculated by MC. Conclusions: The dose distribution for the bare/shielded 169 Yb source was independently verified using mPSD with good agreement in regions close to the source. The 169 Yb source coupled with the partial‐shielding system is an effective technique to deliver IMBT. … (more)
- Is Part Of:
- Medical physics. Volume 47:Issue 9(2020)
- Journal:
- Medical physics
- Issue:
- Volume 47:Issue 9(2020)
- Issue Display:
- Volume 47, Issue 9 (2020)
- Year:
- 2020
- Volume:
- 47
- Issue:
- 9
- Issue Sort Value:
- 2020-0047-0009-0000
- Page Start:
- 4563
- Page End:
- 4573
- Publication Date:
- 2020-07-19
- Subjects:
- Geant4 -- IMBT -- mPSD -- shielded applicator -- TG‐43 -- TG‐186 -- Yb‐169
Medical physics -- Periodicals
Medical physics
Geneeskunde
Natuurkunde
Toepassingen
Biophysics
Periodicals
Periodicals
Electronic journals
610.153 - Journal URLs:
- http://scitation.aip.org/content/aapm/journal/medphys ↗
https://aapm.onlinelibrary.wiley.com/journal/24734209 ↗
http://www.aip.org/ ↗ - DOI:
- 10.1002/mp.14336 ↗
- Languages:
- English
- ISSNs:
- 0094-2405
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 5531.130000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 15331.xml